名称 | ID | 标签 | 描述 | 类型 | 访问 | 分类 |
---|---|---|---|---|---|---|
CO_FailsafeMode_1 | 4057 | Failure behavior current output | Specify how the output should behave in the event of a device alarm. | 读/写 | 枚举 | 2048 |
CO_FailsafeValue_1 | 4058 | Failure current | Enter the value for the "Fixed value" option in the "Failure behavior current output @1" parameter. | 读/写 | 浮点 | 2048 |
Unit_Variable_CO_FailsafeValue_1 | 405A | 读 | 枚举 | 2048 | ||
CO_FixedValue_1 | 405D | Fixed current | Enter the value for the "Fixed value" option. | 读/写 | 浮点 | 2048 |
Unit_Variable_CO_FixedValue_1 | 405E | 读 | 枚举 | 2048 | ||
CO_Input04mA_1 | 4061 | Lower range value output | Enter lower range value for the measured value range. Additional information: - Depending on the setting selected for the "Measuring mode current output @1" parameter, the value specified for this parameter and the "Upper range value output @1" parameter must have the same algebraic sign or not. - As a rule, the lower range value is scaled to be lower than the upper range value. As a result, the behavior of the current output is proportional to the process variable assigned. If the lower range value is scaled to be higher than the upper range value, then the behavior of the current output will be inversely proportional to the process variable assigend. | 读/写 | 浮点 | 2048 |
Unit_Variable_CO_Input04mA_1 | 4064 | 读 | 枚举 | 2048 | ||
CO_Input20mA_1 | 406B | Upper range value output | Enter upper range value for the measured value range. | 读/写 | 浮点 | 2048 |
Unit_Variable_CO_Input20mA_1 | 406C | 读 | 枚举 | 2048 | ||
CO_InputSelector_1 | 4062 | Process variable current output | Select process variable for current output | 读/写 | 枚举 | 2048 |