Name ID Label Description Type Access Class
C4W_TemperatureCompensationMode_1 404D Calculation model Select a formula for the temperature compensation of viscosity R/W Enum 2048
C4W_TemperatureMode_1 404E Temperature correction source Select the temperature correction source. R/W Enum 2048
C4W_ViscosityReferenceTemperature_1 404F Reference temperature Enter reference temperature used to calculate the temperature compensated viscosity R/W Float 2048
C4W_X1_1 4050 Compensation coefficient X1 Enter compensation coefficient used to calculate the temperature compensated viscosity R/W Float 2048
C4W_X2_1 4051 Compensation coefficient X2 Enter compensation coefficient used to calculate the temperature compensated viscosity R/W Float 2048
C4W_ZeroPoint_1 4052 Zero point Displays the zero point correction value for the sensor. NOTE Users logged on in the Service role have write access! R/W Float 2048
C4W_ZeroPointAdjustControl_1 4053 Zero point adjustment control Start or cancel a zero point adjustment. Additional information: The following conditions must be met to perform a zero point adjustment successfully: - The actual flow rate must be 0. - The pressure must be at least 1.034 bar. R/W Enum 2048
C4W_ZeroPointAdjustState_1 4054 Zero point adjustment status Displays the status of the zero point adjustment. R Enum 2048
CO_CorrGain_1 4055 CorrGain 1 R Float 2048
CO_CorrOffset_1 4056 CorrOffset 1 R Float 2048