Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
Dev_UTCTimeZone_1 | 4091 | Time zone | Select the time zone. Every time the time zone is changed, a logbook entry is created. | R/W | Enum | 2048 |
FLOW_CorrVolumeFlow_1 | 4092 | Corrected volume flow | Displays the currently calculated volume flow compensated for the reference density. Additional information: - The reference density can be calculated or fixed. - The applicable unit of measure is specified in the "System units" submenu. | R | Float | 2 |
FLOW_CorrVolumeFlowOffset_1 | 4093 | Corrected volume flow offset | Enter the offset by which to shift the zero point for the corrected volume flow. Additional information: The applicable unit of measure is Nm3/s. | R/W | Float | 2048 |
FLOW_CorrVolumeFlowScale_1 | 4094 | Corrected volume flow factor | Enter the multiplication factor to apply to the corrected volume flow value. | R/W | Float | 2048 |
FLOW_Density_1 | 4095 | Density | Displays the density currently measured. Additional information: The applicable unit of measure is specified in the "System units" submenu. | R | Float | 2 |
FLOW_DensityDampingTime_1 | 4096 | Density damping | Enter time constant for damping (PT1 element) of the measured value for density: - Value = 0: No damping - Value > 0: Damping increases Additional information: Damping is implemented by means of a proportional transmission behavior with first order delay (PT1 element). | R/W | Float | 2048 |
FLOW_DensityForcing_1 | 4097 | Density calculation | R | Enum | 2048 | |
FLOW_DensityOffset_1 | 44F4 | Density offset | Enter the offset by which to shift the zero point for density. Additional information: The applicable unit of measure is kg/m3. | R/W | Float | 2048 |
FLOW_DensityScale_1 | 44F5 | Density factor | Enter the multiplication factor to apply to the density value. | R/W | Float | 2048 |
FLOW_DensityStatus_1 | 4098 | Density status | R | Unsigned | 2 | |
Name | ID | Label | Description | Type | Access | Class |