Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
C4W_TemperatureCompensationMode_1 | 404D | Calculation model | Select a formula for the temperature compensation of viscosity | R/W | Enum | 2048 |
C4W_TemperatureMode_1 | 404E | Temperature correction source | Select the temperature correction source. | R/W | Enum | 2048 |
C4W_ViscosityReferenceTemperature_1 | 404F | Reference temperature | Enter reference temperature used to calculate the temperature compensated viscosity | R/W | Float | 2048 |
C4W_X1_1 | 4050 | Compensation coefficient X1 | Enter compensation coefficient used to calculate the temperature compensated viscosity | R/W | Float | 2048 |
C4W_X2_1 | 4051 | Compensation coefficient X2 | Enter compensation coefficient used to calculate the temperature compensated viscosity | R/W | Float | 2048 |
C4W_ZeroPoint_1 | 4052 | Zero point | Displays the zero point correction value for the sensor. NOTE Users logged on in the Service role have write access! | R/W | Float | 2048 |
C4W_ZeroPointAdjustControl_1 | 4053 | Zero point adjustment control | Start or cancel a zero point adjustment. Additional information: The following conditions must be met to perform a zero point adjustment successfully: - The actual flow rate must be 0. - The pressure must be at least 1.034 bar. | R/W | Enum | 2048 |
C4W_ZeroPointAdjustState_1 | 4054 | Zero point adjustment status | Displays the status of the zero point adjustment. | R | Enum | 2048 |
CO_CorrGain_1 | 4055 | CorrGain 1 | R | Float | 2048 | |
CO_CorrOffset_1 | 4056 | CorrOffset 1 | R | Float | 2048 | |
Name | ID | Label | Description | Type | Access | Class |