Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_TB82_concentration_range | 4014 | Concen range | The concentration range can be set when the solution type is set to user defined. For all other solution types it is read-only | R/W | Enum | 8 |
var_TB82_temp_sensor_type | 4022 | Temp snsr type | Temperature sensor type | R/W | Enum | 8 |
var_TB82_temp_comp_type | 4023 | Temp comp type | Determines the type of Temperature Compensation algorithm applied to the Primary Value | R/W | Enum | 8 |
var_TB82_manual_temp_setpoint_unit | 4043 | Manual temp unit | R/W | Enum | 8 | |
var_TB82_manual_temp_setpoint | 4028 | Manual Temp. | Manual Temperature value is used in the temp comp algorithm when the temp comp type is MANUAL. | R/W | Float | 8 |
var_TB82_temp_auto_sol | 4024 | Auto TC option | This is valid when temp comp type is Automatic. The Automatic Temperature Compensation option applies an algorithm that adjusts for the temperature effect of a solute. The available options are dependent upon the current SENSOR_GROUP. Sensor Group A: Std KCl, TC Coeff, NaOH, NaCl, HCl, H2SO4, User Defined. Sensor Group B: Std KCl, TC Coeff, User Defined. Sensor Group C: Std KCl, TC Coeff, Pure H2O, User Defined. | R/W | Enum | 8 |
var_TB82_temp_pure_h2o | 4025 | Auto TC pure H2O options | This is valid when temp-comp type is Automatic and Auto TC option is Pure H2O. This parameter provides Pure H2O option selection for temp-comp. This selection is restricted to Sensor Group C. | R/W | Enum | 8 |
var_TB82_tc_coef | 4026 | Temp comp coeff | Temperature Compensation Coefficient: This is valid when tem-comp type is Automatic and Auto TC option is TC-coeff. The Temp Coeff is applied as percent of change of conductivity from a reference temp per deg C. Valid range is 0 to 9.99% per deg C. | R/W | Float | 8 |
var_TB82_var_ref_temp | 4027 | Variable ref temp | Variable Reference Temperature provides the ability to use use any temp as the reference temp used to determine the temp-compensated conductivity value (i.e., Primary value). | R/W | Float | 8 |
var_TB82_user_defined_tc_t1 | 404B | Temp 1 | Temp 1 of user defined temperature coefficient. First end point of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t2 | 404C | Temp 2 | Temp 2 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t3 | 404D | Temp 3 | Temp 3 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t4 | 404E | Temp 4 | Temp 4 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t5 | 404F | Temp 5 | Temp 5 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t6 | 4050 | Temp 6 | Temp 6 of user defined temperature coefficient. Last end point of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k1 | 4051 | K1/Kref | Ratio of uncompensated conductivity at Temp 1 to conductivity at reference temperature (K1/Kref). First endpoint of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k2 | 4052 | K2/Kref | Ratio of uncompensated conductivity at Temp 2 to conductivity at reference temperature (K2/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k3 | 4053 | K3/Kref | Ratio of uncompensated conductivity at Temp 3 to conductivity at reference temperature (K3/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k4 | 4054 | K4/Kref | Ratio of uncompensated conductivity at Temp 4 to conductivity at reference temperature (K4/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k5 | 4055 | K5/Kref | Ratio of uncompensated conductivity at Temp 5 to conductivity at reference temperature (K5/Kref). Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_k6 | 4056 | K6/Kref | Ratio of uncompensated conductivity at Temp 6 to conductivity at reference temperature (K6/Kref). Last endpoint of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_conc_x1 | 4015 | x1 Cond | x1 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x2 | 4017 | x2 Cond | x2 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x3 | 4019 | x3 Cond | x3 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
var_TB82_user_defined_conc_x4 | 401B | x4 Cond | x4 Conductivity point of the user defined concentration curve. | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |