Loading…
Name ID Label Description Type Access Class
var_TB82_concentration_range 4014 Concen range The concentration range can be set when the solution type is set to user defined. For all other solution types it is read-only R/W Enum 8
var_TB82_temp_sensor_type 4022 Temp snsr type Temperature sensor type R/W Enum 8
var_TB82_temp_comp_type 4023 Temp comp type Determines the type of Temperature Compensation algorithm applied to the Primary Value R/W Enum 8
var_TB82_manual_temp_setpoint_unit 4043 Manual temp unit R/W Enum 8
var_TB82_manual_temp_setpoint 4028 Manual Temp. Manual Temperature value is used in the temp comp algorithm when the temp comp type is MANUAL. R/W Float 8
var_TB82_temp_auto_sol 4024 Auto TC option This is valid when temp comp type is Automatic. The Automatic Temperature Compensation option applies an algorithm that adjusts for the temperature effect of a solute. The available options are dependent upon the current SENSOR_GROUP. Sensor Group A: Std KCl, TC Coeff, NaOH, NaCl, HCl, H2SO4, User Defined. Sensor Group B: Std KCl, TC Coeff, User Defined. Sensor Group C: Std KCl, TC Coeff, Pure H2O, User Defined. R/W Enum 8
var_TB82_temp_pure_h2o 4025 Auto TC pure H2O options This is valid when temp-comp type is Automatic and Auto TC option is Pure H2O. This parameter provides Pure H2O option selection for temp-comp. This selection is restricted to Sensor Group C. R/W Enum 8
var_TB82_tc_coef 4026 Temp comp coeff Temperature Compensation Coefficient: This is valid when tem-comp type is Automatic and Auto TC option is TC-coeff. The Temp Coeff is applied as percent of change of conductivity from a reference temp per deg C. Valid range is 0 to 9.99% per deg C. R/W Float 8
var_TB82_var_ref_temp 4027 Variable ref temp Variable Reference Temperature provides the ability to use use any temp as the reference temp used to determine the temp-compensated conductivity value (i.e., Primary value). R/W Float 8
var_TB82_user_defined_tc_t1 404B Temp 1 Temp 1 of user defined temperature coefficient. First end point of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t2 404C Temp 2 Temp 2 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t3 404D Temp 3 Temp 3 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t4 404E Temp 4 Temp 4 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t5 404F Temp 5 Temp 5 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_t6 4050 Temp 6 Temp 6 of user defined temperature coefficient. Last end point of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k1 4051 K1/Kref Ratio of uncompensated conductivity at Temp 1 to conductivity at reference temperature (K1/Kref). First endpoint of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k2 4052 K2/Kref Ratio of uncompensated conductivity at Temp 2 to conductivity at reference temperature (K2/Kref). Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k3 4053 K3/Kref Ratio of uncompensated conductivity at Temp 3 to conductivity at reference temperature (K3/Kref). Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k4 4054 K4/Kref Ratio of uncompensated conductivity at Temp 4 to conductivity at reference temperature (K4/Kref). Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k5 4055 K5/Kref Ratio of uncompensated conductivity at Temp 5 to conductivity at reference temperature (K5/Kref). Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_tc_k6 4056 K6/Kref Ratio of uncompensated conductivity at Temp 6 to conductivity at reference temperature (K6/Kref). Last endpoint of the curve. Valid only when Auto-TC option is user-defined R/W Float 8
var_TB82_user_defined_conc_x1 4015 x1 Cond x1 Conductivity point of the user defined concentration curve. R/W Float 8
var_TB82_user_defined_conc_x2 4017 x2 Cond x2 Conductivity point of the user defined concentration curve. R/W Float 8
var_TB82_user_defined_conc_x3 4019 x3 Cond x3 Conductivity point of the user defined concentration curve. R/W Float 8
var_TB82_user_defined_conc_x4 401B x4 Cond x4 Conductivity point of the user defined concentration curve. R/W Float 8
Name ID Label Description Type Access Class