Variable Name: var_TB82_temp_auto_sol
Variable ID: 0x4024
Variable Label: Auto TC option
Variable Help: This is valid when temp comp type is Automatic. The Automatic Temperature Compensation option applies an algorithm that adjusts for the temperature effect of a solute. The available options are dependent upon the current SENSOR_GROUP. Sensor Group A: Std KCl, TC Coeff, NaOH, NaCl, HCl, H2SO4, User Defined. Sensor Group B: Std KCl, TC Coeff, User Defined. Sensor Group C: Std KCl, TC Coeff, Pure H2O, User Defined.
Variable Handling: R/W
Variable Class: Correction
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | Standard KCl |
0x0001 | TC Coefficient |
0x0002 | NaOH |
0x0003 | NaCl |
0x0004 | HCl |
0x0005 | H2SO4 |
0x0006 | Pure H2O |
0x0007 | User defined |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
140 | 4071 | read_TB82_temp_sensor_info |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
144 | 4075 | write_TB82_temp_auto_sol |