Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_TB82_temp_auto_sol | 4024 | Auto TC option | This is valid when temp comp type is Automatic. The Automatic Temperature Compensation option applies an algorithm that adjusts for the temperature effect of a solute. The available options are dependent upon the current SENSOR_GROUP. Sensor Group A: Std KCl, TC Coeff, NaOH, NaCl, HCl, H2SO4, User Defined. Sensor Group B: Std KCl, TC Coeff, User Defined. Sensor Group C: Std KCl, TC Coeff, Pure H2O, User Defined. | R/W | Enum | 8 |
var_TB82_temp_pure_h2o | 4025 | Auto TC pure H2O options | This is valid when temp-comp type is Automatic and Auto TC option is Pure H2O. This parameter provides Pure H2O option selection for temp-comp. This selection is restricted to Sensor Group C. | R/W | Enum | 8 |
var_TB82_tc_coef | 4026 | Temp comp coeff | Temperature Compensation Coefficient: This is valid when tem-comp type is Automatic and Auto TC option is TC-coeff. The Temp Coeff is applied as percent of change of conductivity from a reference temp per deg C. Valid range is 0 to 9.99% per deg C. | R/W | Float | 8 |
var_TB82_var_ref_temp | 4027 | Variable ref temp | Variable Reference Temperature provides the ability to use use any temp as the reference temp used to determine the temp-compensated conductivity value (i.e., Primary value). | R/W | Float | 8 |
var_TB82_user_defined_tc_t1 | 404B | Temp 1 | Temp 1 of user defined temperature coefficient. First end point of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t2 | 404C | Temp 2 | Temp 2 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t3 | 404D | Temp 3 | Temp 3 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t4 | 404E | Temp 4 | Temp 4 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t5 | 404F | Temp 5 | Temp 5 of user defined temperature coefficient. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
var_TB82_user_defined_tc_t6 | 4050 | Temp 6 | Temp 6 of user defined temperature coefficient. Last end point of the curve. Valid only when Auto-TC option is user-defined | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |