Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
tv_Units | 40D5 | Ref.Impd.(TV) unit | Tertiary variable reference impedance engineering units | R | Enum | 8 |
qv_Value | 4031 | Sensor Input (QV) | Quaternary variable sensor input | R | Float | 10 |
qv_Units | 40D6 | Sensor Input (QV) unit | Quaternary variable sensor input engineering units | R | Enum | 8 |
var_analyzerType | 400F | Analyzer type | Primary Value type selection | R/W | Enum | 8 |
var_phSensor | 4016 | pH sensor type | Type of pH sensor. Valid for primary value type = pH | R/W | Enum | 8 |
var_refImpedanceLimit | 401E | Ref. Impedance Limit | The Reference Electrode Impedance that triggers a diagnostic condition. | R/W | Float | 8 |
var_isopotentialPt | 4017 | Isopotential point | The Isopotential Point is the potential of the sensor that is not affected by sample fluid temperature changes. When the PH_SENSOR_TYPE is Glass or Antimony, the Isopotential Point is READ_ONLY. Isopotential Point can be modified only for a Custom pH Sensor Type. Valid for primary value type = pH | R/W | Float | 8 |
var_asymmetricPot | 4018 | Asymmetric Poten. | The Asymmetric Potential value is the electrical potential across the measuring and reference half-cell of the sensor at the isopotential point. When PH_SENSOR_TYPE is Glass or Antimony, Asymmetric Potential is READ_ONLY. Asymmetric Potential can be modified only for Custom pH sensor type. Valid for primary value type = pH | R/W | Float | 8 |
var_sensorDiag | 401D | Sensor diag. | Sensor Diagnostics selection determines whether certain sensor diagnostics are enabled or disabled | R/W | Enum | 8 |
var_valence | 4019 | Valence | Ion valence ranging from -3 to +3. The valence determines the millivolt change per decade of concentration. Valid for primary value type = Ion Concentration | R/W | Enum | 8 |
Name | ID | Label | Description | Type | Access | Class |