Name | ID | Label | Description | Type | Access | Class |
---|---|---|---|---|---|---|
var_refImpedanceLimit | 401E | Ref. Impedance Limit | The Reference Electrode Impedance that triggers a diagnostic condition. | R/W | Float | 8 |
var_isopotentialPt | 4017 | Isopotential point | The Isopotential Point is the potential of the sensor that is not affected by sample fluid temperature changes. When the PH_SENSOR_TYPE is Glass or Antimony, the Isopotential Point is READ_ONLY. Isopotential Point can be modified only for a Custom pH Sensor Type. Valid for primary value type = pH | R/W | Float | 8 |
var_asymmetricPot | 4018 | Asymmetric Poten. | The Asymmetric Potential value is the electrical potential across the measuring and reference half-cell of the sensor at the isopotential point. When PH_SENSOR_TYPE is Glass or Antimony, Asymmetric Potential is READ_ONLY. Asymmetric Potential can be modified only for Custom pH sensor type. Valid for primary value type = pH | R/W | Float | 8 |
var_sensorDiag | 401D | Sensor diag. | Sensor Diagnostics selection determines whether certain sensor diagnostics are enabled or disabled | R/W | Enum | 8 |
var_valence | 4019 | Valence | Ion valence ranging from -3 to +3. The valence determines the millivolt change per decade of concentration. Valid for primary value type = Ion Concentration | R/W | Enum | 8 |
var_magnitudes | 401A | Magnitudes | The number of magnitudes that defines the transmitter output. Valid for primary value type = Ion Concentration | R/W | Enum | 8 |
var_endMag | 401B | End Magnitude | End point magnitude that can be set to 10, 100 or 1000. Valid for primary value type = Ion Concentration | R/W | Enum | 8 |
var_endMv | 401C | End mV | End millivolt value associated with the end magnitude value. Valid for primary value type = Ion Concentration | R/W | Integer | 8 |
var_tempSensorType | 4024 | Temp sensor type | Temperature sensor type | R | Enum | 8 |
var_tempARStatus | 4026 | TC Recognition Status | Temp sensor recognition status. Device scans for TC at power up, when TC type is switched from MANUAL to AUTOMATIC, and when TC recognition is run through the appropriate menu. | R | Enum | 8 |
tempCompType | 4020 | Temp compensation | Determines the type of Temperature Compensation algorithm applied to the Primary Value | R/W | Enum | 8 |
var_manualTemp | 4021 | Manual Temp Setpoint. | Manual Temperature Compensation Value - is used in Nernstian temp. comp. when the Temp Comp selection is Manual | R/W | Float | 8 |
var_pHSolnCoeff | 4022 | pH solution coeff | The solution coefficient is in terms of pH units per 10 deg-C for the pH analyzer state and has a range of +/- 10 pH units per 10 deg-C. The solution coeff is used only when Auto-Solution temp comp is selected. | R/W | Float | 8 |
var_mVSolnCoeff | 4023 | mV solution coeff | The solution coefficient is in terms of millivolts per 10 deg-C for the ORP, pION and ion-conc states and has a range of +/-20 mV per 10 deg-C. The solution coeff is used only when Auto-Solution temp comp is selected. | R/W | Float | 8 |
info_text1 | 4107 | Enter all the transfer function points and then send the curve to the device | Information indicating not to transmit individual curve points one at a time to the device | R | ASCII | 4096 |
var_funcGenTableX1 | 40BE | Function-gen x1 | Function-generator curve x1 value (input value point 1 entered as a percentage of Primary value range). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableX2 | 40BF | Function-gen x2 | Function-generator curve x2 value (input value point 2 entered as a percentage of Primary value range). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableX3 | 40C0 | Function-gen x3 | Function-generator curve x3 value (input value point 3 entered as a percentage of Primary value range). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableX4 | 40C1 | Function-gen x4 | Function-generator curve x4 value (input value point 4 entered as a percentage of Primary value range). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableX5 | 40C2 | Function-gen x5 | Function-generator curve x5 value (input value point 5 entered as a percentage of Primary value range). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableY1 | 40C3 | Function-gen y1 | Function-generator curve y1 value (Analog output value point 1 entered as a percentage of Analog Output). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableY2 | 40C4 | Function-gen y2 | Function-generator curve y2 value (Analog output value point 2 entered as a percentage of Analog Output). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableY3 | 40C5 | Function-gen y3 | Function-generator curve y3 value (Analog output value point 3 entered as a percentage of Analog Output). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableY4 | 40C6 | Function-gen y4 | Function-generator curve y4 value (Analog output value point 4 entered as a percentage of Analog Output). Valid only when Linearity is Function generator | R/W | Float | 8 |
var_funcGenTableY5 | 40C7 | Function-gen y5 | Function-generator curve y5 value (Analog output value point 5 entered as a percentage of Analog Output). Valid only when Linearity is Function generator | R/W | Float | 8 |
Name | ID | Label | Description | Type | Access | Class |