Loading…

Variable Name: statusmap_variable_154

Variable ID: 0x0A2C

Variable Label: Module 2 Failure (AL-05)

Variable Help: Module-2: Start-up sequence, power supply failed.(AL-05) [Action]: 1.Check the cable and connection. 2.Carry out reduction measures of the communication noise. 3.Replace the amplifier. Or contact sales office or service center.

Variable Handling: R/W

Variable Class:

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0003 Failure
0x0005 Function Check
0x0004 Out of Specification
0x0001 Maintenance Required
0x0000 No Effect

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
523 1303 read_condensed_status_map

Variable Write Command List

Command Number Command ID Command Name
524 1304 write_condensed_status_map