Loading…
Loading…
Variable Name: static_pressure_units
Variable ID: 0x4048
Variable Label: SP Unit
Variable Help: Static Pressure Variable Unit- Engineering unit to be displayed with the Digital Value representation, Minimum Span, Upper Range Value, Lower Range Value, Lower Sensor Limit, and Upper Sensor Limit for the Static Pressure.
Variable Handling: R/W
Variable Class: Correction
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0007 | бар |
0x0008 | мбар |
0x000B | Па |
0x000C | кПа |
0x000E | atm |
0x00ED | МПа |
0x00AE | hPa |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
148 | 405E | read_pres_n_sp_hl_alarm_configuration |
169 | 409C | read_sp_trim_points_and_devs |
135 | 4164 | read_static_press_configuration |
173 | 4168 | read_static_press_range_limits |
209 | 428D | read_command_209 |
3 | 00B1 | read_dynamic_variables_and_pv_current |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
137 | 404B | write_static_press_unit |