Loading…
Loading…
Variable Name: input_almout
Variable ID: 0x40B1
Variable Label: Alarm output
Variable Help: The 4-20mA alarm output setting is the function to fix the current output to a selected fixed value if an alarm occurs when self-diagnostic function is performed.
Variable Handling: R/W
Variable Class: Device
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | UNDER 3.0mA |
0x0002 | 4.0mA |
0x0004 | HOLD |
0x0008 | OVER 24.0mA |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
253 | 40CA | read_almout |
Variable Write Command List
Command Number | Command ID | Command Name |
---|---|---|
252 | 40CB | write_almout |