Loading…

Variable Name: error_status

Variable ID: 0x4004

Variable Label: Error Status

Variable Help: SELF DIAGNOSIS MESSAGE, ERROR CODE DESCRIPTION

Variable Handling: R

Variable Class: Dynamic | Diagnostic

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 No Errors
0x0001 E1 Power supply not in order
0x0002 E2 Temp.sensor not in order
0x0003 E3 Angle transducer not in order
0x0004 E4 AD convertor not in order (Problem in AMP board)
0x0005 E5 CPU not in order (Problem in CPU/Indication board)
0x0006 E6 EEPROM not in order (Problem in AMP board)

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
146 4042 read_error_status

Variable Write Command List

None