Loading…

Variable Name: Word_422

Variable ID: 0x44DC

Variable Label: Status Word 422

Variable Help: Reads Modbus register no. 422

Variable Handling: R

Variable Class: Dynamic | Correction

Variable Size: 2

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 API.VCF: Temperature Outside Standard Range
0x0002 API.VCF:Line Density Outside Standard Range
0x0004 Line RTD Temperature Out-Of-Range
0x0008 Meter RTD Temperature Out-Of-Range
0x0010 Flow Direction (0=Forward/Zero, 1=Reverse)
0x0020 Factory Configuration Data Is Invalid
0x0040 Enhanced Density: Unable to fit curve data
0x0080 Last Measured Value Override Active
0x0100 Enhanced Density Extrapolation Alarm
0x0200 Cal Factors Unentered (Flocal Mandatory)
0x0400 1000/2000/3000 EEPROM Checksum Error
0x0800 1000/2000/3000 RAM Test Error
0x1000 Unrecognized/Unentered Sensor Type (K1 Manditory)
0x2000 Core configuration database corrupt
0x4000 Core powerdown totals corrupt
0x8000 Core program corrupt

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
215 434E read_additional_diags

Variable Write Command List

None