Loading…

Variable Name: Word_419

Variable ID: 0x43EB

Variable Label: Status Word 419

Variable Help: Reads Modbus register no. 419

Variable Handling: R

Variable Class: Dynamic | Correction

Variable Size: 2

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Core EEPROM Checksum Error (Config, Powerdown, Program)
0x0002 Core RAM Test Error
0x0004 Real-Time Interrupt Failure
0x0008 Sensor Failure
0x0010 Temperature Sensor Out-of-Range
0x0020 Calibration Failure (Autozero, Density, Temperature)
0x0040 Other Failure
0x0080 Transmitter Initializing/Warming Up
0x0100 PV Out-Of-Limits
0x0200 Non-PV Out-Of-Limits
0x0400 Simulation Mode Active (A132)
0x0800 Undefined
0x1000 Watchdog Error
0x2000 Cold Start (HART bit)
0x4000 Transmitter Configuration Changed (HART bit)
0x8000 Fault (Failure has occurred which affects accuracy)

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
215 41EA read_additional_diags

Variable Write Command List

None