Loading…

Variable Name: device_status

Variable ID: 0x0097

Variable Label:

Variable Help: Field Device Status- Data indicating that the Field Device has detected condition(s) in the Field Device relating to: its hardware / the validity of a variable / its operating status / an internal process

Variable Handling: R/W

Variable Class: Dynamic | Diagnostic | HART

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Process applied to the primary variable is outside the operating limits of the field device
0x0002 Process applied to the non-primary variable is outside the operating limits of the field device
0x0004 Analog output 1 and its digital representation are outside the operating range limits, and not responding to input
0x0008 Analog output 1 and its digital representation are in fixed mode, and not responsive to input changes
0x0010 Field device has more status available
0x0020 A reset or self test of the field device has occurred, or power has been removed and reapplied
0x0040 A modification has been made to the configuration of the field device
0x0080 Field device has malfunctioned due to a hardware error or failure

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
216 4213 read_api_configuration
218 4215 read_external_temperature_configuration
128 41B0 read_static_data_material_and_sensor
130 41B2 read_static_data_special_mass_units
131 41B3 read_static_data_setup_parameters
132 41B4 read_flocal_number
132 41B4 read_flocal_number
136 41B8 read_frequency_output_factors
138 41B9 read_flow_switch_setpoint
140 41BB read_static_data_special_volume_units
164 41C1 read_tseries_parameters
167 41C3 read_zeroing_time
170 41C6 read_max_pulse_width
173 41C8 read_discrete_output_assignment
175 41C9 read_temperature_cal_factor
179 41CC read_zeroing_info
181 41CE read_fault_output_params
188 41D2 read_flowing_density_parameters
190 41D4 read_additional_density_parameters
193 41D7 read_polling_setup_params
195 41D8 read_specified_cutoff
197 41DA read_present_freq
198 41DB read_density_parameters
198 41DB read_density_parameters
200 41DD read_alarm_parameters
203 41E0 read_alarm_variable_assignment
205 41E2 read_pressure_compensation_factors
206 41E3 read_update_rate_settings
209 41E6 read_meter_factors
210 41E7 read_output_channel_setup
214 41E9 read_enhanced_diags
215 41EA read_additional_diags
220 41EB read_ldo_setup
0 00AE read_unique_identifier
1 00AF read_pv
2 00B0 read_pv_current_and_percent_range
3 00B1 read_dynamic_variables_and_pv_current
12 00B4 read_message
13 00B5 read_tag_descriptor_date
14 00B6 read_pv_sensor_info
15 00B7 read_pv_output_info
16 00B8 read_final_assembly_number
33 0FA5 read_transmitter_variables
48 0FCE read_additional_transmitter_status
50 0FD4 read_transmitter_variable_assignments
54 0FDA read_transmitter_variable_information
60 0FE5 read_analog_output_and_percent_of_range
61 0FE7 read_dynamic_variables_and_pv_analog_output
63 0FED read_analog_output_information
70 0FF5 read_analog_output_endpoint_values

Variable Write Command List