Loading…

Variable Name: Word_421

Variable ID: 0x433F

Variable Label: Status Word 421

Variable Help: Reads Modbus register no. 421

Variable Handling: R

Variable Class: Dynamic | Correction

Variable Size: 2

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Burst Mode Enabled
0x0002 Power Reset Occurred
0x0004 Transmitter Initializing/Warming Up (Low Power Fault)
0x0008 Sensor/Xmtr Communication Failure (A28)
0x0010 Paper Out
0x0020 Event #2 Triggered
0x0040 Event #1 Triggered
0x0080 Sensor/Xmtr Communication Failure (A26)
0x0100 Calibration Failure (Autozero, Density, Temperature)
0x0200 Excess Calibration Correction, Autozero too Low
0x0400 Excess Calibration Correction, Autozero too High
0x0800 Process too Noisy to Perform Auto Zero
0x1000 Transmitter Failed
0x2000 Data Loss Possible (Totals+Inventories Questionable)
0x4000 Calibration-In-Progress (Autozero, Density, Temperature)
0x8000 Slug Flow

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
215 41AC read_additional_diags

Variable Write Command List

None