Variable Name: dig_comm_fault_code
Variable ID: 0x40A7
Variable Label: Comm Fault ind.
Variable Help: This fault setting applies to Modbus on RS485 and the display. Downscale: Hold process variables at values that are less than the Lower Sensor Limits and stop totalizing. Upscale: Hold process variables at values that are greater than the Upper Sensor Limits and stop totalizing. Not-a-Number: Set all process variables to IEEE NaN. IntZero-All 0: Set all process variables to 0. IntZero-Flow 0: Set flow rate to 0. None: Process variables displayed as measured (NOTE: Use alternative method to detect fault conditions when NONE is selected)
Variable Handling: R/W
Variable Class: Service
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | Upscale |
0x0001 | Downscale |
0x0002 | IntZero-All 0 |
0x0003 | Not-a-Number |
0x0004 | IntZero-Flow 0 |
0x0005 | Ninguno |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
None
Variable Write Command List
None