Loading…

Variable Name: enable_extern_temp_modulus

Variable ID: 0x40BA

Variable Label: External Temperature Modulus

Variable Help: Enabled: Process measurement will be adjusted for the effect of temperature on the vibrating element. Disabled: Process measurement will not be adjusted for the effect of temperature.

Variable Handling: R/W

Variable Class: Correction

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 Disable
0x0001 Enable

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
218 40B7 read_external_temperature_configuration

Variable Write Command List

Command Number Command ID Command Name
219 40BD write_external_temperature_configuration