Loading…

Variable Name: ao_1_fault_level

Variable ID: 0x40D5

Variable Label: Fault Level (Direct)

Variable Help: This parameter allows you to select the lower or upper mA output level that will indicate a fault condition. For a 12-20 mA output (Direct), the range is 11.5 to 11.8 mA for downscale or 20.5 to 22.0 for upscale. For a 4-20 mA output (Level-Shifted), the range is 3.0 to 3.6 mA for downscale or 21.0 to 24.0 for upscale.

Variable Handling: R/W

Variable Class: Dynamic | AnalogOut

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|4.2f

Variable Edit Format: |en|4.2f

Variable Unit:

Variable Max:

Variable Min:

Variable Read Command List

None

Variable Write Command List

None