Variable Name: device_specific_status_1

Variable ID: 0x102C

Variable Label: Sensor And Output

Variable Help: Specific Statuses Related to Analog Output Current, Sensor and Measurement.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Output Current not stable
0x0002 ADC failure
0x0004 Sensor open circuit
0x0008 Pressure range diapason overrange
0x0010 Pressure sensor limits overrange
0x0020 Not ready
0x0040 Sensor Temperature limits overrange
0x0080 Device Temperature limits overrange

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None