Loading…
Loading…
Variable Name: diag16_text
Variable ID: 0x4915
Variable Label: EoPunterSondenend
Variable Help: EOP Signal liegt unterhalb der Sondenlänge. Check settings: Probe Length, Dielectric Range. Increase Sensitivity. View Echo Curve.
Variable Handling: R/W
Variable Class: Local
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | EoPunterSondenend |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
None
Variable Write Command List
None