Variable Name: epsilonR
Variable ID: 0x4117
Variable Label: Epsilon R
Variable Help: Use this function to enter the relative permittivity. · Allowable input range: 1.1000 - 8.0000 · Default: 2.0000 Enter the figure of 2.0000 if the relative permittivity is not known, as numerous products have "Epsilon R" value approx. 2. However, expect reduced measuring accuracy and/or skipping of the measured value on changeover from tank bottom tracing to the normal measuring mode (= evaluation of reflection from the product surface). (Fct. 3.5.8)
Variable Handling: R/W
Variable Class: Correction
Variable Size: 4
Variable DataType: Float
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: 8
Variable Min: 1.1
Variable Read Command List
None
Variable Write Command List
None