Variable Name: device_status
Variable ID: 0x0097
Variable Label: Device status
Variable Help: Field Device Status- Data indicating that the Field Device has detected condition(s) in the Field Device relating to: its hardware / the validity of a variable / its operating status / an internal process
Variable Handling: R
Variable Class: Dynamic | Diagnostic | HART
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | Process applied to the primary variable is outside the operating limits of the field device |
0x0002 | Process applied to the non-primary variable is outside the operating limits of the field device |
0x0004 | Analog output 1 and its digital representation are outside the operating range limits, and not responding to input |
0x0008 | Analog output 1 and its digital representation are in fixed mode, and not responsive to input changes |
0x0010 | Field device has more status available |
0x0020 | A reset or self test of the field device has occurred, or power has been removed and reapplied |
0x0040 | A modification has been made to the configuration of the field device |
0x0080 | Field device has malfunctioned due to a hardware error or failure |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
163 | 4042 | get_fast_info |
164 | 404A | get_slow_info |
165 | 404C | get_setup_info |
188 | 4061 | read_Solenoid_state |
150 | 4066 | read_clock |
152 | 4068 | read_run_time_seconds |
143 | 4098 | read_logging_event_counter |
155 | 409C | read_event_index |
145 | 409D | read_Wlogging_event_number |
146 | 409F | read_Alogging_event_number |
147 | 40A2 | read_Flogging_event_number |
156 | 40A4 | read_Clogging_event_number |
148 | 40A7 | read_Mlogging_event_number |
0 | 00AE | read_unique_identifier |
1 | 00AF | read_pv |
2 | 00B0 | read_pv_current_and_percent_range |
3 | 00B1 | read_dynamic_variables_and_pv_current |
9 | 00DD | read_device_variables_and_status |
9 | 00DD | read_device_variables_and_status |
7 | 00DB | read_loop_configuration |
8 | 00DC | read_dynamic_variable_classification |
14 | 00B6 | read_pv_sensor_info |
15 | 00B7 | read_pv_output_info |
12 | 00B4 | read_message |
13 | 00B5 | read_tag_descriptor_date |
16 | 00B8 | read_final_assembly_number |
20 | 00DE | read_long_tag |
48 | 1043 | read_additional_device_status |
Variable Write Command List