Loading…
Loading…
Variable Name: device_specific_status_14
Variable ID: 0x1031
Variable Label: Status Group 14
Variable Help: Field Device Specific Status Byte 15- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0002 | 830 Sensor temperature too high Remedy information Reduce ambient temp. around the sensor housing |
0x0004 | 831 Sensor temperature too low Remedy information Increase ambient temp. around the sensor housing |
0x0008 | 832 Electronic temperature too high Remedy information Reduce ambient temperature |
0x0010 | 833 Electronic temperature too low Remedy information Increase ambient temperature |
0x0020 | 834 Process temperature too high Remedy information Reduce process temperature |
0x0040 | 835 Process temperature too low Remedy information Increase process temperature |
0x0080 | 912 Medium inhomogeneous Remedy information 1. Check process cond. 2. Increase system pressure |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None