Loading…

Variable Name: device_specific_status_16

Variable ID: 0x1033

Variable Label: Status Group 16

Variable Help: Field Device Specific Status Byte 17- Data that has Field Device status information relating to: failures / warnings / status of process's.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0002 F062 Sensor connection faulty -> Remedy information: 1. Check or replace sensor electronic module (ISEM) 2. If available: Check connection cable between sensor and transmitter 3. Replace sensor
0x0010 F222 Voltage drift detected -> Remedy information: Change main electronic module
0x0020 144 Measurement error too high -> Remedy information: 1. Check or change sensor 2. Check process conditions
0x0040 944 Monitoring failed -> Remedy information: Check process conditions for Heartbeat Monitoring

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None