Variable Name: device_specific_status_16
Variable ID: 0x1033
Variable Label: Status Group 16
Variable Help: Field Device Specific Status Byte 17- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0002 | F062 Sensor connection faulty -> Remedy information: 1. Check or replace sensor electronic module (ISEM) 2. If available: Check connection cable between sensor and transmitter 3. Replace sensor |
0x0010 | F222 Voltage drift detected -> Remedy information: Change main electronic module |
0x0020 | 144 Measurement error too high -> Remedy information: 1. Check or change sensor 2. Check process conditions |
0x0040 | 944 Monitoring failed -> Remedy information: Check process conditions for Heartbeat Monitoring |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None