Loading…

Variable Name: device_specific_status_1

Variable ID: 0x102C

Variable Label: Status Group 1

Variable Help: Field Device Specific Status Byte 2- Data that has Field Device status information relating to: failures / warnings / status of process's.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0002 C485 Process variable simulation active -> Remedy information: Deactivate simulation
0x0004 F022 Temperature sensor defective -> Remedy information: 1. Check or replace sensor electronic module (ISEM) 2. If available: Check connection cable between sensor and transmitter 3. Replace sensor
0x0008 046 Sensor limit exceeded -> Remedy information: 1. Check sensor 2. Check process conditions
0x0010 F910 Tubes not oscillating -> Remedy information: 1. Check electronics 2. Inspect sensor
0x0020 C453 Flow override active -> Remedy information: Deactivate flow override
0x0080 842 Process value above limit -> Remedy information: 1. Decrease process value 2. Check application 3. Check sensor

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None