Loading…

变量名称: FLOW_MinMeasVarValue_1

变量ID: 0x40AD

变量标签: Low value partial filled pipe detection

变量描述: Enter the lower limit value for the selected process variable. If the measured value drops below the limit value, diagnostic message "862 Partly filled pipe" is generated. Additional information: - This setting applies only if the "Density unit" parameter is not set to °API. - The lower limit value must be lower than the upper limit value ("High value partial filled pipe detection" parameter).

变量处理: 读/写

变量分类: 设备

变量长度: 4

变量数据类型: 浮点

变量显示格式: |en|.4f

变量编辑格式: |en|.4f

变量单位:

变量最大值:

变量最小值:

变量读取命令列表

命令号 命令ID 命令名称
173 42E8 read_index_parameter_173

变量写入命令列表

命令号 命令ID 命令名称
178 42E6 write_index_parameter_178