Loading…
Loading…
Variable Name: xmtr_specific_status_3
Variable ID: 0x0FB8
Variable Label: Status group 4
Variable Help: Field Device Specific Status Byte 4- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | SYSTEM ERROR COM MODULE: EEPROM PARAMETER DATA ERROR |
0x0002 | SYSTEM ERROR COM MODULE: EEPROM TOTALIZER DATA ERROR |
0x0004 | SYSTEM ERROR AMPLIFIER: SOFTWARE VERSION INCOMPATIBLE |
0x0008 | SYSTEM ERROR AMPLIFIER: HARDWARE VERSION INCOMPATIBLE |
0x0010 | SYSTEM ERROR AMPLIFIER: SOFTWARE TYPE CHANGED |
0x0020 | SYSTEM ERROR COM MODULE: EEPROM DEFAULT VALUES |
0x0040 | SYSTEM ERROR COM MODULE: HARDWARE TYPE INCOMPATIBLE |
0x0080 | SYSTEM ERROR COM MODULE: HARDWARE VERSION INCOMPATIBLE |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 0FCE | read_additional_transmitter_status |
Variable Write Command List
None