Loading…
Loading…
Variable Name: device_specific_status_15
Variable ID: 0x1032
Variable Label: Status Group 15
Variable Help: Field Device Specific Status Byte 16- Data that has Field Device status information relating to: failures / warnings / status of process's.
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0002 | F081 Sensor initialization faulty -> Remedy information: 1. Restart device 2. Contact service |
0x0004 | F062 Sensor connection faulty -> Remedy information: Check sensor connection |
0x0008 | F252 Module incompatible -> Remedy information: 1. Check if correct electronic modul is plugged 2. Replace electronic module |
0x0010 | F270 Main electronics defective -> Remedy information: Replace main electronics |
0x0040 | F272 Main electronics faulty -> Remedy information: 1. Restart device 2. Contact service |
0x0080 | M263 Incompatibility detected -> Remedy information: Check electronic module type |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None