Loading…

Variable Name: DEVICE_STATUS_VALUE01

Variable ID: 0x4033

Variable Label: Device Status Value01

Variable Help: The parameter contains detailed information on the current status of the gas detector as a warning matrix.

Variable Handling: R

Variable Class: Dynamic | Correction

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 zero calibration expired
0x0002 span calibration expired
0x0004 measured value fallen below the warning threshold
0x0008 Life time expired, sensor almost used up
0x0010 Sensor warming up, no measured values
0x0020 Temperature too low
0x0040 Alarms disabled

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
128 4035 read_measurement_value_alarm_condition

Variable Write Command List

None