Loading…
Loading…
Variable Name: vSnsrErrE8SincePowerOn
Variable ID: 0x4303
Variable Label: Sensor Error E8 since Power On
Variable Help:
Variable Handling: R
Variable Class: Dynamic | Diagnostic | Device
Variable Size: 1
Variable DataType: BitEnum
Variable Enum Value | Variable Enum Text |
---|---|
0x0001 | EC: No extension block available in sensor; PIR7x00: Unknown Error; OTHER: Sensor sends his error signal |
0x0002 | EC: Sensor test failed; PIR7x00: Basic Initialization required; OTHER: Inconsistency at analog signal |
0x0004 | EC: Potentiostat voltage cannot be read; PIR7x00: Unknown Error; OTHER: Supply current is not ok |
0x0008 | EC: Internal potentiostat voltage cannot be read; PIR7x00: Clean mirror; OTHER: Signal out of permitted range |
0x0010 | EC: Too many lookup values in sensor; PIR7x00: Unknown Error; OTHER: Analog pellister voltage out of permitted range |
0x0020 | EC: Sensor EEPROM is not supported; PIR7x00: Configure gas again; OTHER: Supply current is not ok |
0x0040 | EC: SIOS cannot process sensor test parameter; PIR7x00: Terminal voltage is not ok; OTHER: Unknown Error |
0x0080 | EC: Sensor test failed, sensor can be refilled; PIR7x00: Difference beteween ICOM and 4-20mA interface at the PIR7X00 sensor; OTHER: Unknown Error |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
205 | 42F9 | read_error_history |
Variable Write Command List
None