Loading…

Variable Name: vSnsrErrE8SincePowerOn

Variable ID: 0x4303

Variable Label: Sensor Error E8 since Power On

Variable Help:

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 EC: No extension block available in sensor; PIR7x00: Unknown Error; OTHER: Sensor sends his error signal
0x0002 EC: Sensor test failed; PIR7x00: Basic Initialization required; OTHER: Inconsistency at analog signal
0x0004 EC: Potentiostat voltage cannot be read; PIR7x00: Unknown Error; OTHER: Supply current is not ok
0x0008 EC: Internal potentiostat voltage cannot be read; PIR7x00: Clean mirror; OTHER: Signal out of permitted range
0x0010 EC: Too many lookup values in sensor; PIR7x00: Unknown Error; OTHER: Analog pellister voltage out of permitted range
0x0020 EC: Sensor EEPROM is not supported; PIR7x00: Configure gas again; OTHER: Supply current is not ok
0x0040 EC: SIOS cannot process sensor test parameter; PIR7x00: Terminal voltage is not ok; OTHER: Unknown Error
0x0080 EC: Sensor test failed, sensor can be refilled; PIR7x00: Difference beteween ICOM and 4-20mA interface at the PIR7X00 sensor; OTHER: Unknown Error

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
205 42F9 read_error_history

Variable Write Command List

None