Loading…

Variable Name: device_sensor_status

Variable ID: 0x4002

Variable Label: Sensor Status

Variable Help: Sensor Status from Additional Status Command

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 RTD Resistance too high (PT100 > 500R, PT1000 > 5000R)
0x0002 RTD Short (PT100<5R, PT1000 < 50R)
0x0004 Resistance input too high (>500R)
0x0008 Voltage input too high
0x0010 Sensor circuit supply outside normal limits
0x0020 THC burnout or bad sensor connection
0x0040 RTD Excitation current out of regulation
0x0080 RTD terminals high resistance

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status
48 1043 read_additional_device_status

Variable Write Command List

None