Loading…

Variable Name: device_specific_status_0

Variable ID: 0x102B

Variable Label: System Error Summary

Variable Help: Estado específico del dispositivo de campo Byte 1: Datos que tiene el estado del dispositivo de campo información relativa a: fallos/advertencias/estado de los procesos.

Variable Handling: R

Variable Class: Dynamic | Diagnostic | Device

Variable Size: 1

Variable DataType: BitEnum

Variable Enum Value Variable Enum Text
0x0001 Low value alarm, Frequency or pulse out error
0x0002 High value alarm
0x0004 Too high or low baud rate
0x0010 Current or frequency outputs exceeds the software limits
0x0020 Current or frequency outputs exceeds the hardware limits
0x0040 Parameters restored or saved to flash, flash backup cleared, slug in flow error, password protected item was written
0x0080 No sensor connected, resonant frequency, temperature error, low sensor warning or driver overflow

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None