Variable Name: transfer_function

Variable ID: 0x1F84

Variable Label: Transfer Function

Variable Help: Transfer Function Here you can specify how the current output should behave in case of a change in level: LINEAR A continuous 4-20 mA signal is output which is proportional to the count rate. DISCRETE SWITCH Depending on the set switching threshold (alarm threshold for the level) a current of 4 mA or 20 mA is output. Values between 4 and 20 mA will not be output.

Variable Handling: R/W

Variable Class: InputBlock

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 LINEAR
0x00E6 DISCRETE SWITCH

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
15 00B7 read_pv_output_info

Variable Write Command List

Command Number Command ID Command Name
47 0FB4 write_pv_transfer_function