Loading…

Variable Name: var_ds_fCurrentUpperLimit

Variable ID: 0x4376

Variable Label: Upper Limit

Variable Help: Enter upper limit for current output. According NE43 it is recommended to set Upper Limit value to 20.5mA

Variable Handling: R/W

Variable Class: AnalogOut

Variable Size: 4

Variable DataType: Float

Variable Display Format: |en|.2f

Variable Edit Format: |en|.2f

Variable Unit:

Variable Max:

Variable Min:

Variable Read Command List

Command Number Command ID Command Name
201 43F7 cmd_rd_loop_limits

Variable Write Command List

Command Number Command ID Command Name
202 43F8 cmd_wr_loop_limits