Loading…

Method Name: two_point_sensor_trim

Method ID: 0x403E

Method Label: Two point trim

Method Help: Sensor offset/gain Trim, Calibrate the sensor. The differens between the enteret pressure and the messuret pressure, may max differ 10% USL-LSL. Hi point should be minmum 10 % larger then low Point


  • {
  • long ids[1];
  • char status[3];
  • float trim_point_hi_true;
  • float trim_point_hi_mes;
  • float trim_point_low_true;
  • float trim_point_low_mes;
  • float x;
  • int rc;
  • _fassign(16441,0.0);
  • _fassign(16442,1.0);
  • send(141,status);
  • if (status[0])
  • {
  • display_response_status(141,status[0]);
  • DELAY(2,literal_string(148));
  • process_abort();
  • }
  • DELAY(4,literal_string(150));