Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
edit_type | 4028 | Sensor type | Select sensor type mv,Ohm,T/C,RTD | msrc_UNKNOWN | Details |
reset_function | 4029 | Reset min/max | Resets min/max value measured (datalogger function) | msrc_UNKNOWN | Details |
sensor_trim_reset_function | 402B | Factory trim | Factory trim resets the sensor trim offset/gain to factory setting | msrc_UNKNOWN | Details |
one_point_sensor_trim | 403D | One point trim | Sensor offset Trim, Calibrate the sensor. The differens between the enteret value and the messuret value, may max differ 10% of USL-LSL | msrc_UNKNOWN | Details |
two_point_sensor_trim | 403E | Two point trim | Sensor offset/gain Trim, Calibrate the sensor. The differens between the enteret value and the messuret value, may max differ 10% USL-LSL. Hi point should be minmum 10 % larger then low Point | msrc_UNKNOWN | Details |
pre_function | 401E | msrc_UNKNOWN | Details | ||
post_function | 401F | msrc_UNKNOWN | Details | ||
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
Name | ID | Label | Description | Type | Source |