Loading…
Loading…
Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
KDGM_add_100 | 4044 | Add 100 | msrc_UNKNOWN | Details | |
KDGM_subtract_100 | 4045 | Subtract 100 | msrc_UNKNOWN | Details | |
load_defaults | 4074 | Load Defaults | Load Defaults - restores the transmitter user parameters to default settings in the selected Base Units. | msrc_UNKNOWN | Details |
SM_base_units_method | 4075 | Set Base Units | Base Units - The transmitter can operate using Base Units of metres, feet or inches. Changing the Base Units will RESET the transmitter and restore the transmitter user parameters to default settings in the selected Base Units. | msrc_UNKNOWN | Details |
KDGM_curve_profile_method | 407E | Set Non-Linear Profile | Non-Linear Profile - select a non-linear profile for the appropriate Flow or Contents application. This will automatically set up the required parameters such as P013, P014 and P030 to P039. A selection of SPECIAL allows individual point values to be changed. | msrc_UNKNOWN | Details |
automatic_method_p025 | 406A | Set Temperature | Temperature - if set to Automatic (NaN), the integral temperature sensor is used to determine the temperature in the ullage space. If set to a value then the temperature will be fixed at that value. The temperature is used for calculating the speed of sound. | msrc_UNKNOWN | Details |
automatic_method_p026 | 406B | Set Threshold | Threshold - if set to Automatic (NaN), the threshold is proportional to the echo size. If fixed it sets the minimum detected echo size. | msrc_UNKNOWN | Details |
auto_cycle_method | 4077 | Change Mode | Simulation - the transmitter can simulate the level ramping between minimum and maximum. A complete cycle takes approximately 100 seconds. Simulation may be paused. | msrc_UNKNOWN | Details |
learn_false_echo_method | 40E2 | Learn False Echo | Learn False Echo - store the present target echo as a false echo in the transmitter's database. | msrc_UNKNOWN | Details |
clear_false_echoes_method | 40E3 | Clear False Echoes | Clear False Echo - remove the selected false echo from the transmitter's database. | msrc_UNKNOWN | Details |
auto_map_method | 40E4 | Auto Tank Map | Auto Tank Map - scan the tank for false echoes and store them in the transmitter's database. | msrc_UNKNOWN | Details |
restart_method | 40E6 | Restart Device | Restart Device - the transmitter will be restarted. | msrc_UNKNOWN | Details |
setup_wizard_method | 414D | Basic Setup | Basic Setup - Set up the main parameters. | msrc_UNKNOWN | Details |
present_depth_method | 414E | Present Depth | Set the Bottom Reference automatically by entering the present depth of liquid. Present Depth - enter the present depth of liquid and the transmitter will automatically set the Bottom Reference. Set as Empty - the transmitter will automatically set the Bottom Reference for zero depth. Exit - will exit method. | msrc_UNKNOWN | Details |
calibrate_temperature_method | 414F | Calibrate Temperature | Calibrate temperature sensor in use. | msrc_UNKNOWN | Details |
scale_method | 41F3 | msrc_UNKNOWN | Details | ||
alert_status_method | 418D | Active Alerts | msrc_UNKNOWN | Details | |
loop_test | 1002 | Loop Test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A Trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
applied_rerange | 1007 | Apply Values | Applied values- Perform a device rerange by applying the desired Upper Range Value and Lower Range Values to the device input | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |