Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
KDGM_add_100 | 4044 | Add 100 | msrc_UNKNOWN | Details | |
KDGM_subtract_100 | 4045 | Subtract 100 | msrc_UNKNOWN | Details | |
KDGM_trim_4mA | 407B | Trim 4mA | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
KDGM_trim_20mA | 407C | Trim 20mA | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
load_defaults | 4074 | LOAD DEFAULTS | Load Defaults - restores the transmitter to its ex-factory settings in the selected Base Units | msrc_UNKNOWN | Details |
SM_base_units_method | 4075 | Set Base Units | Base Units - The transmitter can operate using Base Units of metres, feet or inches. Changing the Base Units will RESET the transmitter and restore it to ex-factory settings in the selected Base Units | msrc_UNKNOWN | Details |
KDGM_curve_profile_method | 407E | Set NLP Profile | Curve profile - Selecting a non-linear profile will setup the required parameters such as P013, P014 and P030 to P039. SPECIAL allows individual point values to be changed | msrc_UNKNOWN | Details |
automatic_method_p025 | 406A | Set Temperature | Temperature - If set to a value the temperature will be fixed, else if set to Automatic (NaN), the integral temperature sensor will be used. The temperature is used for speed of sound calculations. | msrc_UNKNOWN | Details |
automatic_method_p026 | 406B | Set Threshold | Threshold - If set to Automatic (NaN), the threshold is proportional to the echo size. If fixed it sets the minimum valid echo size | msrc_UNKNOWN | Details |
auto_cycle_method | 4077 | Set Simulation | Auto Cycle - The transmitter can be made to simulate the level ramping between minimum and maximum. A complete cycle takes approximately 100 seconds. Auto-cycling may be paused. | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |