Loading…
Loading…
Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
CalibrateTarget | 4061 | Target Calib | Captures time-of-flight to target on steam probe types at room ambient. | msrc_UNKNOWN | Details |
abort_from_targ_cal | 4062 | Targ Cal Abort | msrc_UNKNOWN | Details | |
DisplayEventLog | 4060 | Display Event Log | Displays history of significant events such as Power On, Faults, and All Faults Cleared. The time interval displayed with a particular event is the time elapsed since the previous event. The interval associated with a Power On event is always 0. | msrc_UNKNOWN | Details |
dac_trim | 1004 | Trim 4 and 20 | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop Test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
warning_message | 1017 | Set loop to manual | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |