Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
mlfb_meth | 2B28 | Device order num | Device order num- Device Order number. Respects to the current configuration | msrc_UNKNOWN | Details |
siepres_self_test | 2B2D | Self test | Selftest- Commands the SITRANS P electronics to perform a selftest. Reports any errors that occured. | msrc_UNKNOWN | Details |
siepres_master_reset | 2B25 | Master reset | Master reset- Commands the SITRANS P electronics to reset. | msrc_UNKNOWN | Details |
init_setting | 2B24 | Restore mfgr trims | Restore mfgr trims- User trims (Sensor, DAC) will be erased. Factory trims will be restored. | msrc_UNKNOWN | Details |
applied_rerange | 1007 | Apply values | Applied values- Perform a device rerange by applying the desired Upper Range Value and Lower Range Values to the device input. | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
scaled_dac_trim | 1005 | Scaled D/A trim | Digital to Analog Scaled Trim- Allows the calibration of the Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
zero_trim | 1006 | Zero trim | Zero trim- Make the sensor input the new zero input reference. The resulting offset must be within the defined limits. | msrc_UNKNOWN | Details |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
send_warning | 1399 | Send warning | Send Warning- Yields a units code notice to the User. | msrc_UNKNOWN | Details |
sensor_a_lower_sensor_trim | 1787 | Lower sensor trim | Pressure Sensor Lower Trim- Calibrate the pressure sensor. | msrc_UNKNOWN | Details |
sensor_a_upper_sensor_trim | 1789 | Upper sensor trim | Pressure Sensor Upper Trim- Calibrate the pressure sensor. | msrc_UNKNOWN | Details |
leave_trim_mode | 1788 | Exit trim mode | Exit Trim Mode- Notifies the User that the trim is finished. | msrc_UNKNOWN | Details |