Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
meth_var_alarmdelay_1 | 4057 | Alarm Delay | Alarm delay time of alarm signal | msrc_UNKNOWN | Details |
meth_var_filter_1 | 4058 | Output Filter | Filtertime of output current | msrc_UNKNOWN | Details |
meth_sel_alarmLED | 4078 | Alarm LED Hold | Alarm LED Hold: Para. Alarm Hold on : on off : off | msrc_UNKNOWN | Details |
meth_fixed_output_current | 401A | Loop Test | Fix Analog Output- Allows the User to manually manipulate a selected Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
meth_sel_unit | 401C | PV Range/Unit | Select the PV Unit | msrc_UNKNOWN | Details |
meth_sel_cell | 402B | Sensor Selection | Select 2-electrode cell or 4-electrode cell | msrc_UNKNOWN | Details |
meth_sel_temp_unit | 4024 | Temp Unit | Select the temperature unit for HART and local display | msrc_UNKNOWN | Details |
meth_sel_temp_sensor | 4025 | Temp Probe | Select the temperature sensor probe | msrc_UNKNOWN | Details |
meth_sel_TC | 4026 | Temp Compensation | The displayed conductivity value is calculated from measured data and TC (not with SAL) | msrc_UNKNOWN | Details |
meth_lin_log | 4036 | Characteristic | Select linear or logarithmic meas. value / output current relation | msrc_UNKNOWN | Details |
meth_log_lr | 4037 | Log: Lower Range | Select log lower range value (only with log) | msrc_UNKNOWN | Details |
meth_log_ur | 4038 | Log: Upper Range | Select log upper range value (only with log) | msrc_UNKNOWN | Details |
meth_sel_sensocheck | 402C | Sensocheck | Sensocheck continuously monitors the sensor | msrc_UNKNOWN | Details |
select_22mA_on_error | 402D | 22mA on Error | Switch 22mA signal for error message on or off | msrc_UNKNOWN | Details |
select_hold_last_fix | 402E | HOLD Last/Fix | The output current behavior in the HOLD mode can be specified. Last- The output current is frozen at its last value, Fix- A constant value from 3.8mA to 22mA | msrc_UNKNOWN | Details |
meth_self_test | 4019 | Self Test | Performs transmitter self test | msrc_UNKNOWN | Details |
meth_master_reset | 401B | Master Reset | Performs master reset in the Field Device | msrc_UNKNOWN | Details |
meth_config_changed_reset | 4079 | Reset Configuration Changed Sign | Performs reset of configuration changed sign in the Field Device | msrc_UNKNOWN | Details |
meth_take_sample_calibration | 4063 | Take Sample | Sample Calibration Sequence Take Sample | msrc_UNKNOWN | Details |
meth_lab_sample_calibration | 4064 | Input Lab Sample | Sample Calibration Sequence Input Lab Sample | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |