Loading…
Loading…
Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
self_test | 402B | セルフ テスト | Self-test - Transmitter self-test in progress. The analog output is frozen. Recognised failures are displayed. | msrc_UNKNOWN | Details |
master_test | 402C | Transmitter reset | Transmitter reset - The transmitter is restarted. Non-permanent stored data are lost. Self-test is subsequently conducted. | msrc_UNKNOWN | Details |
wire_trim | 402D | Lead balancing | Lead balancing - The resistance of the connection lines to sensor can be compensated for. This step is only necessary for measurement resistors in 2- or 3-wire circuitry. | msrc_UNKNOWN | Details |
reset_parameter | 402E | Data reset | Data reset - All data, which were not made permanent are reset to their original values. This allows the old data to be restored when a parameter setting is interrupted. | msrc_UNKNOWN | Details |
configure_sensor | 4033 | Sensor config. | Input parameter - the parameters for wiring to the input (sensor type, etc.) can be fixed via the function. | msrc_UNKNOWN | Details |
select_unit | 4034 | Process unit | フィールド機器バリアブル単位 - 単位はデジタル値, 最小スパン, Upper レンジ値, Lower レンジ値, Lower センサー リミット, Upper センサー リミットの表示に使用されます。 | msrc_UNKNOWN | Details |
test_range | 4035 | ノーマル モードに戻します | msrc_UNKNOWN | Details | |
configure_output | 4038 | Output config. | Configuration of analog output - The setting determines the behaviour of the output in case of failure. | msrc_UNKNOWN | Details |
send_data | 4039 | Send data | msrc_UNKNOWN | Details | |
return_to_normal | 1016 | ノーマル モードに戻します | msrc_UNKNOWN | Details | |
burn_eeprom | 100C | EEPROM の焼付け | 電気的に消去可能 Read Only Memory の焼付け - フィールド機器が Random Access Memory に存在するデータを電気的に消去可能 Read Only Memory にコピーする場合の操作で, ホストによる変更を永続的なものにします。 | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |