Name | ID | Label | Description | Type | Source |
---|---|---|---|---|---|
method_master_reset | 4003 | Master reset | msrc_UNKNOWN | Details | |
method_TB82_reset_cal | 4088 | Reset Cal | Reset the process and temperature sensor calibration data to the values set at the factory | msrc_UNKNOWN | Details |
method_TB82_temp_cal | 408D | Temp Cal | Temperature calibration | msrc_UNKNOWN | Details |
method_TB82_one_point_cal | 4089 | One point calib | One Point process calibration | msrc_UNKNOWN | Details |
method_two_point_cal | 408A | Two point calib | Two Point Calibration - Calibration procedure where two different pH values are used, typically at or near the Lower Range Value and Upper Range Value of the transmitter. | msrc_UNKNOWN | Details |
method_release_output_hold | 408F | Release output hold | msrc_UNKNOWN | Details | |
method_exit_from_2pt_cal | 4090 | Exit from 2 point cal | msrc_UNKNOWN | Details | |
method_dummy_abort | 408E | Dummy abort method | msrc_UNKNOWN | Details | |
method_exit_remote_2_point_cal | 408C | Exit Remote Two-Pt Cal | The method exits from an on-going remote 2 point cal procedure that was started remotely from HART commands | msrc_UNKNOWN | Details |
method_read_2pt_cal_activity | 408B | Read Two-Point PV cal activity | Reads whether a REMOTE or a LOCAL Two-Point PV calibration is currently in progress | msrc_UNKNOWN | Details |
method_refresh_variables | 4000 | Refresh variables | Read all transmitter data to refresh their values. This will update any variable that was modified via the device keypad. | msrc_UNKNOWN | Details |
applied_rerange | 1007 | Rerange-Apply values | Applied values- Perform a device rerange by applying the desired Upper Range Value and Lower Range Values to the device input | msrc_UNKNOWN | Details |
loop_test | 1002 | Loop test | Loop test- Allows the User to manually manipulate the Analog Output to a selected constant output value | msrc_UNKNOWN | Details |
scaled_dac_trim | 1005 | Scaled D/A trim | Digital to Analog Scaled Trim- Allows the calibration of the Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
dac_trim | 1004 | D/A trim | Digital to Analog Trim- Allows the calibration of a selected Analog Output with an external reference at the operating endpoints of the Analog Output | msrc_UNKNOWN | Details |
return_to_normal | 1016 | Aborting return to normal | msrc_UNKNOWN | Details | |
leave_fixed_current_mode | 1003 | Leaving fixed crnt mode | Leaving Fixed Current Mode- Takes the Analog Output out of a mode where the output was a constant value, and returns the output to a mode where the output tracks the input | msrc_UNKNOWN | Details |
Name | ID | Label | Description | Type | Source |