0 |
|
1 |
|
2 |
Serial number |
3 |
Figure |
4 |
Figure: Display |
5 |
Figure: Application |
6 |
Figure: Adjustment |
7 |
Figure: Linearization |
8 |
Figure: Current output |
9 |
DD version |
10 |
1.05.05-00 |
11 |
Measurement loop name |
12 |
Name of the measurement loop |
13 |
Sensor |
14 |
Sensor acc. to WHG |
15 |
Use as a part of an overfill protection system according to WHG |
16 |
No |
17 |
Yes |
18 |
SIL qualified sensor |
19 |
Max. adjustment in percent |
20 |
Level in percent for which the distance 'Distance A' must be entered |
21 |
.2f |
22 |
% |
23 |
Min. adjustment in percent |
24 |
Level in percent for which the distance 'Distance B' must be entered |
25 |
Distance A (max. adjustment) |
26 |
Distance from sensor reference plane to product surface at max. level |
27 |
.3f |
28 |
Distance B (min. adjustment) |
29 |
Distance from sensor reference plane to product surface at min. level |
30 |
Integration time (t) |
31 |
Damping of measured value peaks |
32 |
.0f |
33 |
s |
34 |
Type of linearization |
35 |
Selection of the linearization curve |
36 |
Linear |
37 |
Horizontal cylindrical vessel |
38 |
Spherical vessel |
39 |
Venturi, trapezoidal weir, rectangle weir |
40 |
Palmer-Bowlus Flume |
41 |
V-Notch, triangle weir |
42 |
User programmable |
43 |
Menu language |
44 |
Selection of the language for the indicating and adjustment module |
45 |
Deutsch |
46 |
English |
47 |
Français |
48 |
Español |
49 |
Pycckuu |
50 |
Italiano |
51 |
Nederlands |
52 |
Chinese |
53 |
Japanese |
54 |
Lighting |
55 |
Activation of the background lighting of the indicating and adjustment module |
56 |
Off |
57 |
On |
58 |
Indication value |
59 |
Determines which measured value should be displayed |
60 |
Distance |
61 |
Filling height |
62 |
Percent |
63 |
Lin. percent |
64 |
Scaled |
65 |
Current |
66 |
Graph |
67 |
Selection of the presentation for the measured value image |
68 |
Measured value |
69 |
Measured value with bar graph |
70 |
Failure mode |
71 |
Current consumption of the sensor in case of failure |
72 |
22 mA |
73 |
20.5 mA |
74 |
< 3.6 mA |
75 |
No change |
76 |
Output characteristic |
77 |
Selection of the current characteristics |
78 |
4...20 mA |
79 |
20...4 mA |
80 |
Min. current |
81 |
Min. current consumption of the sensor (exception: value in case of failure) |
82 |
3.8 mA |
83 |
4 mA |
84 |
Max. current |
85 |
Max. current consumption of the sensor (exception: value in case of failure) |
86 |
20 mA |
87 |
Activity to be carried out |
88 |
Selection, which reset should be carried out |
89 |
Restore delivery status |
90 |
Restore basic settings |
91 |
Device name |
92 |
VEGAPULS WL 61 |
93 |
Linearization table percentage values |
94 |
Linearization table lin. percentage values |
95 |
Lin% |
96 |
Actually measured distance value |
97 |
Percentage value |
98 |
Linearized |
99 |
Linearized value (e.g. volume-proportional value) |
100 |
Device status |
101 |
Level offset |
102 |
Expanded Device Type |
103 |
Identification of the instrument type through Fieldbus |
104 |
Device address |
105 |
Filling height of the level |
106 |
.1f |
107 |
.4f |
108 |
Measured value, current output |
109 |
mA |
110 |
HART Device Status |
111 |
Distance unit of the instrument |
112 |
Selection of the physical distance unit for indication and parameter adjustment |
113 |
m |
114 |
ft |
115 |
Temperature unit of the instrument |
116 |
Selection of the physical temperature unit for indication and parameter adjustment |
117 |
°C |
118 |
K |
119 |
°F |
120 |
Measured value offset |
121 |
Manufacturer ID |
122 |
x |
123 |
Bustype ID |
124 |
Scaling format |
125 |
Determination of the presentation of decimal points |
126 |
# |
127 |
#.# |
128 |
#.## |
129 |
#.### |
130 |
#.#### |
131 |
Sensor type |
132 |
Scaling unit |
133 |
Measuring unit for the scaled value |
134 |
mm |
135 |
cm |
136 |
in |
137 |
mbar |
138 |
bar |
139 |
Pa |
140 |
kPa |
141 |
MPa |
142 |
psi |
143 |
mmH2O |
144 |
mmHg |
145 |
inH2O |
146 |
ftH2O |
147 |
inHg |
148 |
torr |
149 |
User defined |
150 |
l |
151 |
hl |
152 |
m³ |
153 |
in³ |
154 |
gal |
155 |
ft³ |
156 |
bbl |
157 |
ImpGal |
158 |
yd³ |
159 |
bbl_liq |
160 |
bush |
161 |
kg |
162 |
t |
163 |
lb |
164 |
g |
165 |
STon |
166 |
LTon |
167 |
oz |
168 |
kg/m³ |
169 |
kg/dm³ |
170 |
g/cm³ |
171 |
g/ml |
172 |
g/l |
173 |
lb/gal |
174 |
lb/in³ |
175 |
STon/yd³ |
176 |
deg Brix |
177 |
deg Baume lt |
178 |
deg Baume hv |
179 |
% sol wt |
180 |
% sol vol |
181 |
API |
182 |
pcf |
183 |
SGU |
184 |
l/s |
185 |
l/min |
186 |
l/h |
187 |
Ml/d |
188 |
m³/s |
189 |
m³/min |
190 |
m³/h |
191 |
m³/d |
192 |
gal/s |
193 |
gal/min |
194 |
gal/h |
195 |
gal/d |
196 |
Mgal/d |
197 |
ft³/s |
198 |
ft³/min |
199 |
ft³/h |
200 |
ft³/d |
201 |
bbl/s |
202 |
bbl/min |
203 |
bbl/h |
204 |
bbl/d |
205 |
ImpGal/s |
206 |
ImpGal/min |
207 |
ImpGal/h |
208 |
ImpGal/d |
209 |
g/s |
210 |
g/min |
211 |
g/h |
212 |
kg/s |
213 |
kg/min |
214 |
kg/h |
215 |
kg/d |
216 |
lb/s |
217 |
lb/min |
218 |
lb/h |
219 |
lb/d |
220 |
STon/min |
221 |
STon/h |
222 |
STon/d |
223 |
LTon/h |
224 |
LTon/d |
225 |
t/min |
226 |
t/h |
227 |
t/d |
228 |
- |
229 |
pF |
230 |
ct/s |
231 |
lb/ft |
232 |
kg/m |
233 |
Hz |
234 |
dB |
235 |
mV |
236 |
unknown |
237 |
m/s |
238 |
m/h |
239 |
ft/s |
240 |
ft/min |
241 |
Hardware version |
242 |
Hardware version of the main electronics |
243 |
Type of medium |
244 |
Selection of the type of medium |
245 |
Liquids |
246 |
Application |
247 |
Special vessel feature |
248 |
Storage tank |
249 |
Storage tank with product agitation |
250 |
Stirrer vessel/Reactor |
251 |
Dosing vessel |
252 |
Plastic vessel (measurement through the vessel top) |
253 |
Transportable plastic tank (measurement through the tank top) |
254 |
Open water (gauge measurement) |
255 |
Open flume (flow measurement) |
256 |
Rain water overflow (weir) |
257 |
Demonstration |
258 |
Storage tank on ships (cargo tank) |
259 |
Stilling tube |
260 |
Bypass |
261 |
Vessel form for bulk solids |
262 |
Silo (tall and narrow) |
263 |
Bunker (large volume) |
264 |
Bunker with quick filling |
265 |
Heap (point measurement/profile detection) |
266 |
Crusher |
267 |
Medium/Dielectric constant |
268 |
Water based solutions/Acids/Alkalis (DK > 10) |
269 |
Medium |
270 |
Powder, dust (granulation size < 1 mm) |
271 |
Granules, pellets (granulation 1 ... 10 mm) |
272 |
Crushed stones, pebbles, boulders (granulation size > 10 mm) |
273 |
Vessel top |
274 |
Form of the vessel ceiling |
275 |
Flat |
276 |
Torispherical |
277 |
Vessel bottom |
278 |
Form of the vessel bottom |
279 |
Conical |
280 |
Inclined |
281 |
Tube inside diameter |
282 |
To compensate for the propagation speed of radar signals within tubes |
283 |
Vessel height/Measuring range |
284 |
Determine the range for the echo evaluation |
285 |
DTM ID |
286 |
Status distance |
287 |
Failure |
288 |
Upper limitation |
289 |
Lower limitation |
290 |
Simulation value |
291 |
Substitute value |
292 |
Constant value |
293 |
Selection of the physical unit for indication and parameter adjustment |
294 |
Status percent |
295 |
Status lin. percent |
296 |
Status scaled |
297 |
Status current output |
298 |
Selection of an activity to process the gating out of false signals |
299 |
Create new |
300 |
Update |
301 |
Delete part |
302 |
Delete all |
303 |
Sounded distance |
304 |
Distance of the sensor reference plane to the product surface |
305 |
from |
306 |
Distance from the sensor reference plane to the range starting point |
307 |
to |
308 |
Distance from the sensor reference plane to the range end point |
309 |
Amplitude certainty of the 0 % curve |
310 |
Safety of the false signal suppression |
311 |
Position certainty of the 0 % curve in the near range |
312 |
Position certainty of the 100 % curve in the near range |
313 |
Position certainty of the 0 % curve in the far range |
314 |
Position certainty of the 100 % curve in the far range |
315 |
Gradient of the manual sectors |
316 |
Position certainty at the end of the false signal memory |
317 |
Position certainty depending on the import range |
318 |
Use default false signal suppression |
319 |
Limitation measuring range begin |
320 |
Limitation measuring range begin active |
321 |
Certainty at vessel bottom or measuring range end |
322 |
Additional factor safety at the measuring range end |
323 |
Certainty at vessel bottom or measuring range end in % of the set measuring range |
324 |
PIN |
325 |
Determine the PIN which should be valid for the function 'Lock adjustment' |
326 |
04d |
327 |
corresponds to 0 % |
328 |
Scaled value for 0 % |
329 |
corresponds to 100 % |
330 |
Scaled value for 100 % |
331 |
Device status acc. to NE 107 |
332 |
OK |
333 |
Function check |
334 |
Maintenance required |
335 |
Out of specification |
336 |
Malfunction |
337 |
Detail status |
338 |
Detailed status value for identification of the exact cause of error |
339 |
Probe length |
340 |
Measuring range |
341 |
Primary Variable (PV) |
342 |
Selection, which measured value should be outputted as Primary Variable (PV) |
343 |
Secondary Variable (SV) |
344 |
Selection, which measured value should be outputted as Secondary Variable (SV) |
345 |
Third Variable (TV) |
346 |
Selection, which measured value should be outputted as Third Variable (TV) |
347 |
Quaternary Variable (QV) |
348 |
Selection, which measured value should be outputted as Quaternary Variable (QV) |
349 |
Delay 'Fault signal with echo loss' |
350 |
Date of the min. distance |
351 |
Distance (min.) |
352 |
Date of the max. distance |
353 |
Distance (max.) |
354 |
Date of the min. electronics temperature |
355 |
Electronics temperature (min.) |
356 |
Date of the max. electronics temperature |
357 |
Electronics temperature (max.) |
358 |
Averaging factor on increasing amplitude |
359 |
Averaging factor on decreasing amplitude |
360 |
'First large echo' active |
361 |
Amplitude threshold 'First large echo' |
362 |
Function measured value filter |
363 |
Automatically |
364 |
Low averaging |
365 |
High averaging |
366 |
'Smooth raw value curve' active |
367 |
Offset threshold value curve |
368 |
Function 'Combine echoes' |
369 |
Deactivated |
370 |
Only bottom section |
371 |
Complete measuring range |
372 |
Amplitude difference 'Combine echoes' |
373 |
Position difference 'Combine echoes' |
374 |
Width focusing range |
375 |
Time for opening of the focus range |
376 |
Minimum measurement reliability |
377 |
Min. measuring reliability outside the focusing range |
378 |
Fault signal on loss of echo |
379 |
Electronics temperature |
380 |
Actual electronics temperature |
381 |
Defines, if the diagnostic message 'Function check' should be outputted or suppressed |
382 |
Defines, if the diagnostic message 'Maintenance required' should be outputted or suppressed |
383 |
Defines, if the diagnostic message 'Out of specification' should be outputted or suppressed |
384 |
Measurement reliability |
385 |
Password |
386 |
Password to release the service level |
387 |
Software version |
388 |
Software version of the main electronics |
389 |
Device-Revision |
390 |
Compatibility information for integration into the device catalogue |
391 |
Loop current mode |
392 |
Fixed current (4 mA) with HART |
393 |
Analogue current output with HART |
394 |
Increase of the system noise |
395 |
0 |
396 |
Electronics version |
397 |
Electronics version of the main electronics |
398 |
--- |
399 |
PS60HC |
400 |
PS60HK |
401 |
PS60HS |
402 |
PS60PAC |
403 |
PS60PAK |
404 |
PS60PAS |
405 |
PS60FFC |
406 |
PS60FFK |
407 |
PS60FFS |
408 |
PLICSZEBVL |
409 |
PLICSZEBVH |
410 |
PS60WLH |
411 |
PS60WLPA |
412 |
PS60WLFF |
413 |
FX80H |
414 |
FX80HSIL |
415 |
FX80PA |
416 |
FX80FF |
417 |
PLICSZEZSA |
418 |
BR80H |
419 |
BR80HSIL |
420 |
BR80PA |
421 |
BR80FF |
422 |
BR80Z |
423 |
BR80HK |
424 |
BR80HKSIL |
425 |
BR80PAK |
426 |
BR80FFK |
427 |
BR80ZK |
428 |
PS60HW |
429 |
PS60PAW |
430 |
PS60FFW |
431 |
FX80MB |
432 |
PS64HW |
433 |
BR18-2L |
434 |
BR18-2LE |
435 |
BR18-2LEB |
436 |
BR18-3L |
437 |
Version of the connected supplementary electronics |
438 |
System noise treatment |
439 |
Application-dependent |
440 |
Overfill safety |
441 |
MESSAGE |
442 |
Message stored in the instrument for free use |
443 |
Fieldbus Profile Rev. |
444 |
Setup |
445 |
Diagnostics |
446 |
Additional settings |
447 |
Info |
448 |
Measured values |
449 |
Figure: Setup |
450 |
Figure: Min./Max. adjustment |
451 |
Figure: Damping |
452 |
Figure: Simulation |
453 |
Figure: Status signals |
454 |
Figure: Further settings |
455 |
Figure: False signal suppression |
456 |
Figure: Scaling |
457 |
Figure: Reset |
458 |
Figure: HART variables |
459 |
Figure: Service login |
460 |
Figure: Info |
461 |
Settings in the special parameters were carried out. A change of application is only possible after resetting the sensor to delivery status. |
462 |
Min./max. adjustment |
463 |
Damping |
464 |
Current output |
465 |
The sensor is in multidrop mode |
466 |
Status signals |
467 |
The instrument has a self-monitoring acc. to NE 107. Define the diagnostic states to be outputted by the instrument. |
468 |
Measured value status |
469 |
Peak indicator, distance |
470 |
Peak indicator, electronics temperature |
471 |
False signal suppression |
472 |
Scaling |
473 |
Reset |
474 |
HART variables |
475 |
Service login |
476 |
Special parameters |
477 |
Level |
478 |
Additional measured values |
479 |
Outputs |
480 |
Read values from device |
481 |
Write values to device |
482 |
Linearization table |
483 |
Linearization curve |
484 |
Table |
485 |
Echo curve |
486 |
Echo curve of the setup |
487 |
Adjustment |
488 |
Echo curve (curve) |
489 |
Show echo curve |
490 |
Settings |
491 |
Load actual echo curves |
492 |
Load false signal suppression |
493 |
Load setup curve |
494 |
Load selected curves |
495 |
Store echo curve of the setup |
496 |
Zoom range from |
497 |
Zoom range up to |
498 |
Device characteristics |
499 |
Details |
500 |
Version |
501 |
Update instrument features |
502 |
Simulation |
503 |
Activate simulation |
504 |
Deactivate simulation |
505 |
Accept simulation value |
506 |
Identification |
507 |
HART Primary Value (PV) |
508 |
Trend display |
509 |
Service login required |
510 |
Parameter adjustment |
511 |
Maintenance |
512 |
Service logout |
513 |
Reset? |
514 |
Reset peak indicator |
515 |
Reset 'Configuration changed' |
516 |
VEGAPULS 61 |
517 |
VEGAPULS 62 |
518 |
VEGAPULS 63 |
519 |
VEGAPULS 64 |
520 |
VEGAPULS 65 |
521 |
VEGAPULS 66 |
522 |
VEGAPULS 67 |
523 |
VEGAPULS 68 |
524 |
VEGAPULS WL61 |
525 |
VEGAPULS SR68 |
526 |
VEGAPULS 69 |
527 |
VEGAFLEX 80 series |
528 |
VEGAFLEX 80 series SIL |
529 |
VEGABAR 80 series |
530 |
VEGABAR 80 series SIL |
531 |
VEGADIF 80 series |
532 |
VEGADIF 80 series SIL |
533 |
VEGA Grieshaber KG |
534 |
Siemens |
535 |
KROHNE Messtechnik GmbH |
536 |
WIKA |
537 |
Balluff GmbH |
538 |
Error in the calibration |
539 |
Exchange electronics or return device for repair |
540 |
Error in the device setting |
541 |
Repeat setup or reset |
542 |
Installation/setup error |
543 |
Check installation and/or parameter adjustment |
544 |
Measurement function disturbed |
545 |
Carry out a reset or briefly separate operating voltage |
546 |
Communication error |
547 |
Briefly separate operating voltage, exchange device |
548 |
No measured value available |
549 |
Adjusted span too small |
550 |
Change adjustment according to the limit values |
551 |
Error in the linearization table |
552 |
Check linearization table |
553 |
No executable software |
554 |
Repeat update, exchange electronics or device |
555 |
Error in the electronics |
556 |
General software error |
557 |
Briefly separate operating voltage |
558 |
Measured value is determined |
559 |
Wait for switch-on phase |
560 |
Impermissible electronics temperature |
561 |
Check ambient temperature, isolate electronics |
562 |
Error in the delivery status |
563 |
Repeat reset or load XML file with sensor data into the sensor |
564 |
Error in the non-active linearization table |
565 |
Error in the event memory |
566 |
Reliability too low |
567 |
Check the installation and process conditions |
568 |
Error on an device interface |
569 |
Check the connections, exchange the electronics or return the device for repair |
570 |
Check ambient temperature, isolate electronics or use a device with higher temperature range |
571 |
Overfill protection for liquids |
572 |
Make sure that no more filling is made and check the level in the vessel |
573 |
Optimize sensor position and orientation |
574 |
Open calculation assistant |
575 |
Common Tables |
576 |
Interoperable * |
577 |
Universal |
578 |
Common Practice |
579 |
Pressure Universal |
580 |
Pressure Common |
581 |
Menu |
582 |
Rosemount Magflow |
583 |
Rosemount Pressure |
584 |
Moore Products |
585 |
Siemens Pressure |
586 |
P2 Common |
587 |
PV |
588 |
Wireless |
589 |
Munge |
590 |
Analytical Oxygen/Combustion/Methane/CO |
591 |
Flow Totalizer - Rate Indicator |
592 |
Flow Variable Area |
593 |
Flow Coriolis Mass |
594 |
Flow Ultrasonic - Gas |
595 |
Flow Ultrasonic - Liquid |
596 |
Level Ultrasonic |
597 |
Level Radar - Microwave |
598 |
Pressure Absolute |
599 |
Level Capacitance |
600 |
Pressure Differential |
601 |
Level Hydrostatic |
602 |
Level |
603 |
Pressure Absolute/Relative |
604 |
Level Microwave |
605 |
Level Radar |
606 |
Level Density - Nuclear |
607 |
Flow Electromagnetic |
608 |
Flow Vortex |
609 |
Flow Magmeter - Rack Version |
610 |
Flow High Performance Magmeter |
611 |
Flow Mass |
612 |
Flow Ultrasonic |
613 |
Flow Magnetic |
614 |
Flow Magnetic - Custody Transfer |
615 |
Flow 2-wire Magnetic |
616 |
Flow Coriolis Mass - Custody Transfer |
617 |
Flow 2-wire Vortex |
618 |
Flow Thermal |
619 |
Temperature Head Mounted |
620 |
Temperature Compact |
621 |
Temperature DIN Rail Mounted |
622 |
Analytical pH |
623 |
Analytical Conductivity - Conductive |
624 |
Analytical Conductivity - Inductive |
625 |
Temperature Sensor |
626 |
Level Servo |
627 |
Temperature DIN Rail |
628 |
Temperature |
629 |
Analytical ORP |
630 |
Analytical pION |
631 |
Analytical IConc |
632 |
Analytical Electrode Conductivity |
633 |
Analytical Toroidal Conductivity |
634 |
Valve Positioner |
635 |
Valve Controller |
636 |
Pressure |
637 |
Pressure Multivariable |
638 |
Analytical pH/ORP |
639 |
Flow Coriolis |
640 |
Analytical Conductivity |
641 |
Valve Positioner/Pneumatic |
642 |
Valve Actuator - Electrical |
643 |
Temperature Dual Input |
644 |
Valve Actuator |
645 |
Analytical Corrosion |
646 |
Analytical Density - Nuclear |
647 |
Level Continuous Gamma |
648 |
Level Laser Ranging |
649 |
Pressure Safety |
650 |
Flow Vortex/Swirl |
651 |
Flow |
652 |
Analytical 4 Electrode Conductivity |
653 |
Analytical 2 Electrode Conductivity |
654 |
Analytical Troidal Conductivity |
655 |
Analytical pH/ORP/pIon/Ion Conc |
656 |
Analytical Oxygen |
657 |
Pressure P, dP + Multivariable |
658 |
Temperature 2-wire |
659 |
Pressure w/ PID |
660 |
Pressure Saturated Steam |
661 |
Analytical Density - Gamma Ray |
662 |
Level Gamma Ray |
663 |
Analytical Weight - Gamma Ray |
664 |
Level w/ Density - Gamma Ray |
665 |
Analytical Motorized Density |
666 |
Flow Multivariable Mass |
667 |
Temperature Multivariable |
668 |
Specialty HART-to-Analog Signal Conv |
669 |
Flow dP |
670 |
Pressure Low Power |
671 |
Pressure Certified SIL2 |
672 |
Interoperable |
673 |
Analytical Density |
674 |
Valve Positioner/Electropneumatic |
675 |
Flow Magnetic Inductive |
676 |
Analytical |
677 |
Analytical Opacity Monitor |
678 |
Analytical Oxygen/Combustion |
679 |
Analytical 2-wire Conductivity |
680 |
Analytical 2-wire Conductivity w/ pure H20 |
681 |
Analytical 2-wire Conductivity w/ % conc |
682 |
Analytical 2-wire Oxygen/Chlorine/Ozone |
683 |
Analytical 2-wire pH/ORP |
684 |
Analytical 2-wire Contacting/Toroidal |
685 |
Analytical Oxygen - High Temperature |
686 |
Analytical Conductivity w/ % conc |
687 |
Analytical pH/ORP/ISE |
688 |
Analytical Amperometric Analyzer |
689 |
Analytical Dual Input Analyzer |
690 |
Pressure Pressure/Level |
691 |
Analytical Consistency |
692 |
Analytical Microwave-Pulp Consistency |
693 |
Analytical Microwave Consistency |
694 |
Valve Digital Positioner |
695 |
Valve Smart Valve Positioner |
696 |
Temperature Differential |
697 |
Analytical Dissolved Oxygen |
698 |
Analytical pH Controller |
699 |
Analytical Conductivity/Resistivity Controller |
700 |
Analytical Conductivity/Resistivity |
701 |
Level Displacement |
702 |
Temperature w/ PID |
703 |
Valve Universal Positioner |
704 |
Level Radar - Reflex |
705 |
Pressure Gauge |
706 |
Pressure Sealed Gauge |
707 |
Valve Digital Positioner w/ PID |
708 |
Valve Intelligent Positioner |
709 |
Analytical Gas Detector |
710 |
Analytical Infrared Gas Detector |
711 |
Analytical Electrochemical Gas Detector |
712 |
Analytical Measuring Instruments |
713 |
Level Magnostrictive |
714 |
Valve On/Off Valve Controller |
715 |
Level Radar - Pulse |
716 |
Level Continuous Capactive |
717 |
Valve Damper Actuator |
718 |
Valve Electric Damper Actuator |
719 |
Flow Electronic |
720 |
Flow Vortex - Multiparameter Mass |
721 |
Analytical UV Flame Detector |
722 |
Analytical UVIR Flame Detector |
723 |
Analytical IR Flame Detector |
724 |
Analytical Fire Detector |
725 |
Analytical Gas Detector Display Interface |
726 |
Analytical Toxic Gas Detector |
727 |
Analytical Flame Detector |
728 |
Level Nuclear |
729 |
Analytical Microwave |
730 |
Analytical Density - Radiometric |
731 |
Level Radiometric |
732 |
Analytical Gas/Steam |
733 |
Level Nucleonic |
734 |
Analytical Blade Consistency |
735 |
Analytical Rotary Consistency |
736 |
Pressure Absolute/Differential |
737 |
Positioner |
738 |
Analytical Combustible Gas Detector |
739 |
Analytical Point Infrared Gas Detector |
740 |
Analytical Multi-Detector Monitor System |
741 |
Valve Positioner/Electro-hydraulic |
742 |
Valve Monitor |
743 |
Enable on token-passing data link layer only |
744 |
Enable on TDMA data link layer only |
745 |
Enable on both TDMA and token-passing data link layer |
746 |
WirelessHART Device |
747 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
748 |
Primary |
749 |
Primary Analog Channel |
750 |
Secondary |
751 |
Secondary Analog Channel |
752 |
Tertiary |
753 |
Tertiary Analog Channel |
754 |
Quaternary |
755 |
Quaternary Analog Channel |
756 |
Quinary |
757 |
Quinary Analog Channel |
758 |
Critical Power Failure |
759 |
Lock all |
760 |
Trim not supported |
761 |
Configuration cannot be changed |
762 |
Locked by gateway |
763 |
Analog Input Channel |
764 |
Analog Output Channel |
765 |
Secondary Analog Channel Saturated |
766 |
Tertiary Analog Channel Saturated |
767 |
Quaternary Analog Channel Saturated |
768 |
Quinary Analog Channel Saturated |
769 |
Secondary Analog Channel Fixed |
770 |
Tertiary Analog Channel Fixed |
771 |
Quaternary Analog Channel Fixed |
772 |
Quinary Analog Channel Fixed |
773 |
Simulation active |
774 |
The device is in simulation mode and the level output is not representative of the process. |
775 |
Non-Volatile memory failure |
776 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
777 |
Volatile memory error |
778 |
The RAM memory check is invalid or maybe corrupt |
779 |
Watchdog reset executed |
780 |
A watchdog reset has been performed |
781 |
Voltage conditions out of range |
782 |
A voltage condition is outside its allowable range |
783 |
Environmental conditions out of range |
784 |
An internal or environmental condition is beyond acceptable limits. |
785 |
Electronic failure |
786 |
A hardware problem not related to the sensor has been detected. |
787 |
Subdevice list changed |
788 |
Duplicate master detected |
789 |
Capacity Denied |
790 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
791 |
Bandwidth allocation pending |
792 |
Block transfer pending |
793 |
Continuous |
794 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
795 |
Window |
796 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
797 |
Rising |
798 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
799 |
Falling |
800 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
801 |
On-change |
802 |
The Burst Message is triggered when any value in the message changes. |
803 |
Battery life |
804 |
Percent range |
805 |
Loop current |
806 |
Primary variable |
807 |
Secondary variable |
808 |
Tertiary variable |
809 |
Quaternary variable |
810 |
Off |
811 |
Enable event notification on token-passing data link layer |
812 |
Enable event notification on TDMA data link layer |
813 |
Enable event notification on TDMA and token-passing data link layer |
814 |
Configuration changed event pending |
815 |
Device status event pending |
816 |
More status available event pending |
817 |
Disable |
818 |
Enable single data point trending |
819 |
Enable filtered trending |
820 |
The sample is filtered using a time constant equal to one-third the trend sample period |
821 |
Enable average trending |
822 |
The value obtained by dividing the sum of all samples in the period by the total number of samples |
823 |
Read receive time |
824 |
Write date and time |
825 |
Non-Volatile Clock |
826 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
827 |
Clock Uninitialized |
828 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
829 |
Line Power |
830 |
Device is powered via a hardwired connection to a plant power source. |
831 |
Battery Power |
832 |
Device operates solely of its (internal) battery. |
833 |
Recharg.B./Power Sc. |
834 |
Rechargeable battery power or power scavenging. The device has a short-term supply of energy (must be at least one hour) that is continuously being replenished. The device's power is being replenished by harvesting and converting energy from the environment surrounding the device (e.g. solar, vibration, heat). |
835 |
Idle |
836 |
Active search |
837 |
Looking for neighbors. |
838 |
Negotiating |
839 |
Quarantined |
840 |
Operational |
841 |
Suspended |
842 |
Deep sleep |
843 |
Ultra-low power |
844 |
Wireless Signal Found |
845 |
Network packets have been recognized |
846 |
Wireless Signal Identified |
847 |
Wireless Time Synchronized |
848 |
WirelessHart Signal Identified |
849 |
Network Admission Requested |
850 |
Set on transmission of first join request |
851 |
Join Retrying |
852 |
Set after the first join request retry (i.e. when number of join requests > 2). Cleared when device is authenticated or when Active Search started. |
853 |
Join Failed |
854 |
Set on transition from Active Search mode to Deep Sleep mode. Cleared on transition from Deep Sleep to Active Search. |
855 |
Network Security Clearance Granted |
856 |
Network Manager Session Established |
857 |
Network Joined |
858 |
Network Joined. Normal superframe and links obtained. |
859 |
Network Bandwidth Requested |
860 |
Negotiating Network Properties. Gateway session obtained. |
861 |
Join Complete |
862 |
Normal operating mode. |
863 |
None |
864 |
No restrictions |
865 |
SI Only |
866 |
Restricted to SI Unit codes only |
867 |
Open - accept all devices |
868 |
Use Whitelist for access |
869 |
Use Blacklist |
870 |
Use Whitelist and Blacklist |
871 |
Lockdown - disable join links |
872 |
Process automation device |
873 |
Discrete device |
874 |
Hybrid:Process automation+discrete |
875 |
I/O System |
876 |
WirelessHART Process automation device |
877 |
WirelessHART Discrete device |
878 |
WirelessHART Process automation+discrete |
879 |
WirelessHART Gateway |
880 |
WirelessHART Access point |
881 |
WirelessHART Process adapter |
882 |
WirelessHART Discrete adapter |
883 |
WirelessHART enabled handheld/portable maintenance tool |
884 |
Don't attempt to join |
885 |
Join now |
886 |
Attempt to join immediately on powerup or reset |
887 |
US |
888 |
United States of America |
889 |
JP |
890 |
Japan |
891 |
DE |
892 |
Germany |
893 |
FR |
894 |
France |
895 |
ES |
896 |
Spain |
897 |
RU |
898 |
Russian Federation |
899 |
CN |
900 |
People's Republic of China |
901 |
Nominal |
902 |
Device Power Status is nominal |
903 |
Low |
904 |
Device Power Status is low |
905 |
Critically Low |
906 |
Device Power Status is critically low |
907 |
Recharging - Low |
908 |
Recharging - High |
909 |
Secondary Master |
910 |
Primary Master |
911 |
PV Analog Channel Saturated |
912 |
PV Analog Channel Fixed |
913 |
Device Profile |
914 |
Device Diagnostic Status 0 |
915 |
Device Diagnostic Status 1 |
916 |
I/O and Subdevice Status |
917 |
WirelessHART Status |
918 |
Time Set Option |
919 |
Real Time Clock Flags |
920 |
Country |
921 |
Country of intended device installation |
922 |
SI Unit Control |
923 |
Indicates if device should only display SI unit codes |
924 |
Burst Trigger Units |
925 |
Burst Trigger Class |
926 |
Wired HART Enabled |
927 |
Enabled on Wired HART device connection only |
928 |
Wireless HART Enabled |
929 |
Enabled on WirelessHART device connection only |
930 |
Both Wired and WirelessHart Enabled |
931 |
Enabled on both Wired and WirelessHART connections |
932 |
Burst Msg Trigger Mode |
933 |
Event Status |
934 |
Event Notification Control |
935 |
Operate mode Mask |
936 |
Not Defined |
937 |
Power Source |
938 |
Power source of the device |
939 |
Wireless Mode |
940 |
Join Status |
941 |
The current location in the joining sequence of operations. |
942 |
Network Access Mode |
943 |
Join Mode |
944 |
Power Status |
945 |
Power status of the device |
946 |
Master Mode |
947 |
This sets the I/O System's Data Link Layer operation on all channels as either primary or secondary master mode. |
948 |
Trend Units |
949 |
Trend Device Variable Unit Code |
950 |
Trend Classification |
951 |
Trend Device Variable Classification |
952 |
Trend Control |
953 |
Device Status Mask |
954 |
Ext dev status Mask |
955 |
Device Diagnostic Status 0 Mask |
956 |
Device Diagnostic Status 1 Mask |
957 |
Device Diagnostic Status 2 Mask |
958 |
Device Diagnostic Status 3 Mask |
959 |
AO saturated Mask |
960 |
AO fixed Mask |
961 |
Latched Device Status |
962 |
Latched Ext dev status |
963 |
Latched Operate Mode |
964 |
Latched Device Diagnostic Status 0 |
965 |
Latched Device Diagnostic Status 1 |
966 |
Latched Device Diagnostic Status 2 |
967 |
Latched Device Diagnostic Status 3 |
968 |
Latched AO saturated |
969 |
Latched AO fixed |
970 |
Wireless Module Device Type |
971 |
Device type of the Wireless Radio |
972 |
Wireless Module Manufacturer ID |
973 |
Manufacturer of the Wireless Radio contained in this device |
974 |
Subdevice Device Type |
975 |
Device type of the subdevice attached to this system. |
976 |
Subdevice Manufacturer Id |
977 |
Manufacturer of the subdevice attached to this system. |
978 |
Time Stamp |
979 |
Relative time value on network |
980 |
8u |
981 |
Dynamic Variables Returned for Device Variables |
982 |
Command Response Truncated |
983 |
Configuration Change Counter Mismatch |
984 |
Status bytes mismatch |
985 |
8.3f |
986 |
Failure while executing the action |
987 |
No echo curve loaded |
988 |
Execute action |
989 |
Action successfully carried out. |
990 |
The instrument is locked! The action cannot be executed. |
991 |
Features were updated. |
992 |
The password is wrong! |
993 |
Restore delivery status;Restore basic settings;Cancel |
994 |
Do you really want to reset the selected peak indicators? |
995 |
Cancel;Create new;Update;Delete range;Delete all |
996 |
False signal suppression was deleted completely. |
997 |
Range of the false signal suppression was deleted. |