0 |
2700Analog Mass Flow |
1 |
.1f |
2 |
No Additional Information for Above |
3 |
Additional Information for Above |
4 |
.2f |
5 |
.3f |
6 |
.4f |
7 |
.5f |
8 |
4.2f |
9 |
8d |
10 |
8.2f |
11 |
8.3f |
12 |
8.6f |
13 |
.6f |
14 |
10.4f |
15 |
6.2f |
16 |
5.4f |
17 |
5.2f |
18 |
5d |
19 |
5.3f |
20 |
1d |
21 |
2d |
22 |
6.5f |
23 |
7.2f |
24 |
4d |
25 |
4.1f |
26 |
9.8f |
27 |
10.7f |
28 |
3d |
29 |
FMU860 |
30 |
DT |
31 |
DVC |
32 |
FCX |
33 |
AS800 |
34 |
340T |
35 |
340C |
36 |
344 |
37 |
Siemens Micro K |
38 |
SITRANS L |
39 |
SIPAN PH |
40 |
SITRANS P |
41 |
75SMT |
42 |
UNIDELTA |
43 |
8320 |
44 |
SX5100 |
45 |
TSV175 |
46 |
DMU |
47 |
3780 |
48 |
BM70 |
49 |
8u |
50 |
low |
51 |
high |
52 |
tmp_av |
53 |
number |
54 |
uValue |
55 |
fValue |
56 |
alarmNum |
57 |
select |
58 |
floatFreqVal |
59 |
floatRateVal |
60 |
floatPulsesPerUnit |
61 |
floatUnitsPerPulse |
62 |
floatFaultVal |
63 |
intLMVFaultTimeout |
64 |
Current Selection;Mass Flow Rate;Temperature;Density;Volume Flow Rate;Density At Reference Temperature;Volume Flow At Reference Temperature;Average Observed Density;Average Observed Temperature;Density at Reference (Concentration);Specific Gravity (Concentration);Temperature-Corrected Volume Flow Rate (Concentration);Net Mass Flow Rate (Concentration);Net Volume Flow Rate (Concentration);Concentration;Drive Gain;Pressure;External Temperature;Baume (Concentration);Gas Standard Volume Flow |
65 |
Priority: Failed - Fix Now
Continuity failure of drive circuit, LPO, or RPO, or LPO-RPO mismatch when driving.
Recommended Actions:
1. Check for slug flow.
2. Check the test points.
3. Check the sensor circuitry.
4. Check sensor tubes for plugging.
5. If the problem persists, contact Micro Motion. |
66 |
No Additional Information Available |
67 |
Priority: Failed - Fix Now
The value computed for the resistance of the line RTD or the Meter/Case RTD is outside limits.
Recommended Actions:
1. Check the sensor RTD circuitry.
2. Verify that process temperature is within range of sensor and transmitter.
3. If the problem persists, contact Micro Motion |
68 |
Priority: Failed - Fix Now
The measured flow has exceeded the maximum flow rate of the sensor.
Recommended Actions:
1. If other alerts are present (typically, A003, A006, A008, A102, or A105), resolve those alerts first. If the A005 alert persists, continue with the suggestions below.
2. Verify process and check for slug flow.
3. Check the test points
4. Check the sensor circuitry.
5. Check the sensor tubes for erosion.
6. If the problem persists, contact Micro Motion. |
69 |
Priority: Failed - Fix Now
One or more required characterization parameters have not been entered.
Recommended Actions:
1. Check the characterization. Specifically, verify the FCF and K1 values.
2. If the problem persists, contact Micro Motion |
70 |
Flow Cal:
Value %0
K1:
Value %1 uSec
K2:
Value %2 uSec |
71 |
Flow FCF:
Value %0
K1:
Value %1 uSec
K2:
Value %2 uSec |
72 |
Priority: Failed - Fix Now
The measured density has exceeded 0-10 g/cm3.
Recommended Actions:
1. If other alerts are present (typically, A003, A006, A102, or A105), resolve those alerts first. If the A008 alert persists, continue with the suggestions below.
2. Verify process. Check for air in the flow tubes, tubes not filled, foreign material in tubes, or coating in tubes.
3. Check for slug flow.
4. Check the sensor circuitry.
5. Verify calibration factors in transmitter configuration.
6. Check the test points.
7. If the problem persists, contact Micro Motion |
73 |
Priority: Failed - Fix Now
Transmitter in power-up mode.
Recommended Actions:
XX |
74 |
Priority: Failed - Fix Now
Mechanical zero: The resulting zero was greater than 3 microseconds
Temperature or Density Calibrations: many possible causes
Recommended Actions:
1. If alert appears during a transmitter zero, ensure that there is no flow through the sensor, then retry.
2. Cycle power to the flowmeter, then retry.
3. If appropriate, restore the factory zero to return the flowmeter to operation. |
75 |
Priority: Failed - Fix Now
Mechanical zero: Calibration procedure failed due to reverse flow through the sensor.
Recommended Actions:
1. Ensure that there is no flow through the sensor, then retry.
2. Cycle power to the flowmeter, then retry.
3. If appropriate, restore the factory zero to return the flowmeter to operation. |
76 |
Priority: Failed - Fix Now
Mechanical zero: Calibration procedure failed due to flow through the sensor.The resulting zero was greater than 3 microseconds.
Recommended Actions:
1. Ensure that there is no flow through the sensor, then retry.
2. Cycle power to the flowmeter, then retry.
3. If appropriate, restore the factory zero to return the flowmeter to operation. |
77 |
Priority: Failed - Fix Now
Mechanical zero: The resulting zero was greater than 3 microseconds.
Recommended Actions:
1. Remove or reduce sources of electromechanical noise, then retry. Sources of noise include mechanical pumps, pipe stress at sensor, electrical interference, or vibration effects from nearby machinery.
2. Cycle power to the flowmeter, then retry.
3. If appropriate, restore the factory zero to return the flowmeter to operation. |
78 |
Priority: Failed - Fix Now
Many possible causes.
Recommended Actions:
1. Cycle power to the flowmeter.
2. The transmitter might need service. Contact Micro Motion. |
79 |
Priority: Failed - Fix Now
The value computed for the resistance of the Line RTD is outside limits.
Recommended Actions:
1. Check the sensor RTD circuitry.
2. Verify that process temperature is within range of sensor and transmitter.
3. Check the characterization. Specifically, verify the FCF and K1 values
4. If the problem persists, contact Micro Motion. |
80 |
Priority: Failed - Fix Now
The value computed for the resistance of the Meter/Case RTD is outside limits.
Recommended Actions:
1. Check the sensor RTD circuitry.
2. Verify that process temperature is within range of sensor and transmitter.
3. Check the characterization. Specifically, verify the FCF and K1 values
4. If the problem persists, contact Micro Motion. |
81 |
Priority: Failed - Fix Now
The transmitter has experienced an EEPROM checksum failure.
Recommended Actions:
1.Cycle power to the transmitter.
2.If problem persists, call Micro Motion. |
82 |
Priority: Failed - Fix Now
A RAM location in the transmitter cannot be written to.
Recommended Actions:
1.Cycle power to the transmitter.
2.If problem persists, call Micro Motion. |
83 |
Priority: Failed - Fix Now
The flow calibration factor and/or K1 has not been entered since the last master rese
Recommended Actions:
1. Check the characterization. Specifically, verify the FCF and K1 values.
2. If the problem persists, contact Micro Motion. |
84 |
Priority: Failed - Fix Now
The sensor is recognized as a straight tube but the K1 value indicates a curved tube, or vice versa.
Recommended Actions:
1. Check the characterization. Specifically, verify the FCF and K1 values.
2. Check the sensor RTD circuitry.
3. If the problem persists, contact Micro Motion. |
85 |
Priority: Failed - Fix Now
Uncorrectable checksum mismatch.
Recommended Actions:
1. Cycle power to the flowmeter.
2. The transmitter might need service. Contact Micro Motion. |
86 |
Priority: Failed - Fix Now
Communication errors between transmitter and core processor.
Recommended Actions:
1. Verify the wiring between the Transmitter and Core Processor.
2. Cycle power to the transmitter.
3. If problem persists, call Micro Motion. |
87 |
Priority: Failed - Fix Now
Weights and Measures security seal has been broken.
Recommended Actions:
1. Alarm can be cleared by user, but authorized procedure is required to reestablish security. |
88 |
Priority: Failed - Fix Now
An attempt to write data to the core processor has failed.
Recommended Actions:
1. Cycle power to the flowmeter.
2. The transmitter might need service or upgrading. Contact Micro Motion. |
89 |
Priority: Failed - Fix Now
The core processor is not receiving enough power.
Recommended Actions:
1. Check the power supply and all power wiring. |
90 |
Priority: Failed - Fix Now
Meter verification is in progress, with outputs set to fault.
Recommended Actions:
1. Allow the procedure to complete.
2. If desired, abort the procedure and restart with outputs set to last measured value. |
91 |
Priority: Failed - Fix Now
No signal from Left Pickoff or Right Pickoff, suggesting that sensor tubes are not vibrating.
Recommended Actions:
1. Verify process. Check for air in the flow tubes, tubes not filled, foreign material in tubes, or coating in tubes. |
92 |
Priority: Advisory
The calculated amount of current output is outside of the linear range.
Recommended Actions:
1. Verify that mA output range values are set to appropriate values. |
93 |
Primary Variable:
%0
%0:
Value %1 %2 |
94 |
PV LRV:
Value %0 %1
PV URV:
Value %2 %3 |
95 |
Priority: Advisory
Non-zero HART address configured, or user has set the mA output to a fixed value.
Recommended Actions:
1. Check the HART address. If non-zero, set mA Output Action parameter to enable Live mA Output.
2. Exit mA output trim.
3. Exit mA output loop test.
4. Check to see if the output has been fixed via digital communications. |
96 |
HART Address:
%0
mAO Fault Action:
%1 |
97 |
Priority: Maintenance - Fix Soon
The drive power (current/voltage) is at its maximum.
Recommended Actions:
1. Check the sensor circuitry.
2. If this is the only active alarm, it can be ignored. If desired, reconfigure the alarm severity to Ignore. |
98 |
Priority: Maintenance - Fix Soon
The core processor was unable to successfully store the totalizers and inventories on the last power down and must rely on the previously saved values. These values may be up to two hours old.
Recommended Actions:
1. Contact Micro Motion to upgrade the transmitter software.
Note: The transmitter is still functional. |
99 |
Priority: Advisory
A calibration procedure is in progress.
Recommended Actions:
1. Allow the flowmeter to complete calibration.
2. For zero calibration procedures, you may abort the calibration, set the zero time parameter to a lower value, and restart the calibration. |
100 |
Priority: Maintenance - Fix Soon
The density has exceeded the user-defined slug (density) limits.
Recommended Actions:
1. Check that sensor is completely full.
2. Check that slug limits are set to appropriate values. |
101 |
Slug High Limit:
Value %0 g/Cucm
Slug Low Limit:
Value %1 g/Cucm
Slug Duration:
Value %2 sec |
102 |
Priority: Advisory
Device is in HART burst mode.
Recommended Actions:
1. No action required.
2. If desired, reconfigure the alarm severity to Ignore. |
103 |
Priority: Maintenance - Fix Soon
Transmitter has been restarted.
Recommended Actions:
1. No action required. |
104 |
Priority: Maintenance - Fix Soon
Event 1 has been triggered.
Recommended Actions:
1. No action required.
2. If desired, reconfigure the alarm severity to Ignore. |
105 |
Event 1 Setpoint:
Value %0
Event 1 Variable:
%1
Event 1 Type:
%2 |
106 |
Priority: Maintenance - Fix Soon
Event 2 has been triggered.
Recommended Actions:
1. No action required.
2. If desired, reconfigure the alarm severity to Ignore. |
107 |
Event 2 Setpoint:
Value %0
Event 2 Variable:
%1
Event 2 Type:
%2 |
108 |
Priority: Advisory
Process variable assigned to frequency output is outside configured scale limits.
Recommended Actions:
1. Change frequency output scaling factors to more appropriate settings. |
109 |
Third Variable:
%0
%0:
Value %1 %2 |
110 |
Rate Factor:
Value %0 Hz
Pulses/Unit:
Value %1
Units/Pulse:
Value %2 |
111 |
Priority: Advisory
User has set the frequency output to a fixed value.
Recommended Actions:
1. Exit frequency output loop test. |
112 |
Tube Frequency:
Value %0 Hz |
113 |
Priority: Maintenance - Fix Soon
No response received from polled device.
Recommended Actions:
1. HART polling connection to external device has failed. Ensure that external device is available:
2. Verify device operation.
3. Verify wiring.
4. Verify polling configuration. |
114 |
Poll Control:
%0
Ext Device Tag 1:
%1
Polled Variable 1:
%2 |
115 |
Ext Device Tag 2:
%0
Polled Variable 2:
%1 |
116 |
Priority: Maintenance - Fix Soon
The process temperature is outside Petroleum Measurement defined extrapolation limits.
Recommended Actions:
1. Verify process.
2. Verify configuration for Petroleum Measurement reference table and temperature. |
117 |
Priority: Maintenance - Fix Soon
The process density is outside Petroleum Measurement defined extrapolation limits.
Recommended Actions:
1. Verify process.
2. Verify configuration for Petroleum Measurement reference table and density. |
118 |
Priority: Advisory
The user has set the discrete output to a fixed state.
Recommended Actions:
1. No action required.
2. If desired, reconfigure the alarm severity to Ignore. |
119 |
Priority: Advisory
The configured density/temperature/concentration values do not result in a proper Concentration Measurement (CM) curve.
Recommended Actions:
1. Verify Concentration Measurement (CM) configuration. |
120 |
Priority: Maintenance - Fix Soon
The process temperature or density is outside the user-defined extrapolation limits.
Recommended Actions:
1. Verify the process temperature and density.
2. Verify Concentration Measurement (CM) configuration. |
121 |
Priority: Maintenance - Fix Soon
Meter verification in progress, with outputs set to last measured value.
Recommended Actions:
1. Allow the procedure to complete.
2. If desired, abort the procedure and restart with outputs set to fault. |
122 |
Priority: Advisory
Simulation mode is enabled.
Recommended Actions:
1. Disable output simulation. |
123 |
Priority: Advisory
Discrete Event 1 has been triggered.
Recommended Actions:
1. No action required. |
124 |
Discrete Event Var:
%0
Discrete Event Type:
%1 |
125 |
Setpoint A:
Value %0
Setpoint B:
Value %1 |
126 |
Priority: Advisory
Discrete Event 2 has been triggered.
Recommended Actions:
1. No action required. |
127 |
Priority: Advisory
Discrete Event 3 has been triggered.
Recommended Actions:
1. No action required. |
128 |
Priority: Advisory
Discrete Event 4 has been triggered.
Recommended Actions:
1. No action required. |
129 |
Priority: Advisory
Discrete Event 5 has been triggered.
Recommended Actions:
1. No action required. |
130 |
Priority: Advisory
Discrete Input is active.
Recommended Actions:
1. No action required. |
131 |
Priority: Maintenance - Fix Soon
Primary variable has exceeded sensor limits.
Recommended Actions:
1. Verify process and check for slug flow.
2. Check the test points.
3. Check the sensor circuitry
4. Check the sensor tubes for erosion.
5. If the problem persists, contact Micro Motion. |
132 |
Priority: Maintenance - Fix Soon
Non-primary variable has exceeded sensor limits.
Recommended Actions:
1. Verify process and check for slug flow.
2. Check the test points.
3. Check the sensor circuitry
4. Check the sensor tubes for erosion.
5. If the problem persists, contact Micro Motion. |
133 |
Priority: Maintenance - Fix Soon
The calculated amount of current output is outside of the linear range.
Recommended Actions:
1. Verify that mA output range values are set to appropriate values. |
134 |
Priority: Maintenance - Fix Soon
Power has been removed and reapplied resulting in the reinstallations of the setup information.
Recommended Actions:
1. The first command to recognize this condition will automatically reset this flag.
2. This flag may also be set following a Master Reset or a Self Test. |
135 |
Priority: Maintenance - Fix Soon
A write or set command has been executed.
Recommended Actions:
1. No action required. |
136 |
Priority: Maintenance - Fix Soon
A hardware error or failure has been detected by the device.
Recommended Actions:
1. Check the test points.
2. Check the sensor circuitry.
3. If the problem persists, contact Micro Motion. |