0 |
Model 705 3.x |
1 |
MAGNETROL |
2 |
Device ID |
3 |
Poll Address |
4 |
Faults |
5 |
Software Failure |
6 |
Software Failure - An unrecoverable software error has been detected; consult Factory if problem persists. |
7 |
CPU Failure |
8 |
CPU Failure - Microprocessor failure; consult Factory if problem persists. |
9 |
EEPROM Failure |
10 |
Non-Volatile Memory Failure - EEPROM Read or Write failure; consult Factory if problem persists. |
11 |
Default Params |
12 |
Default Parameters - The application configuration information has been reset to default values. Please reconfigure the instrument using the Calibration Menu. |
13 |
No End of Ramp |
14 |
No End of Ramp - The End-of-Ramp signal from the analog board is absent. This may indicate a hardware failure, bad connection, or missing analog board. |
15 |
Loop Failure |
16 |
Loop current differs from commanded value. Check Loop Trim. Consult factory if problem persists. |
17 |
Fiducial Shift |
18 |
Fiducial ticks have shifted from expected value; Consult factory if problem persists. |
19 |
Slope Error |
20 |
Ramp circuit is generating improper voltage; Consult factory if problem persists. |
21 |
No Probe |
22 |
No Probe - No probe is connected to the transmitter. |
23 |
No Fiducial |
24 |
No Fiducial Pulse - Check configuration, especially Fiducial Type and Probe Model. Check connection between circuit boards in module. |
25 |
Safety Zone Alrm |
26 |
The measured level has entered the safety zone with the SZ Fault enabled. The fault is cleared when the level falls below the safety zone hysteresis zone unless it is latched. A latched safety zone fault must be manually reset. |
27 |
No Signal |
28 |
Missing Level Signal - Check probe connection and application parameters, especially Probe Model and Probe Length. |
29 |
EOP High |
30 |
End-of-Probe signal is higher than expected. Check configuration, especially Probe Model and Probe Length. |
31 |
Hi Volume Alrm |
32 |
The measured level is more than 5% above the highest point in the strapping table. |
33 |
Lvl < Probe Length |
34 |
The apparent position of the upper level pulse is beyond the end of the probe, i.e. the Sensor Level Distance > Probe Length. Check configuration, especially probe length and threshold. May indicate unexpected water bottom situation. |
35 |
EOP < Probe Length |
36 |
The apparent position of the end-of-probe pulse is beyond the end of the probe, i.e. the Sensor Level Distance > Probe Length. Check configuration, especially probe length and threshold. May indicate process condition where upper level was not detected causing EOP to be displaced. |
37 |
Warnings |
38 |
Seal Leak |
39 |
May indicate failing seal in the 7xS probe where steam is condensing in the vicinity of high-frequency connector. Consult factory if problem persists. |
40 |
Fiducial Spread |
41 |
Fiducial ticks variation is excessive suggesting that the fiducial signal is unstable. Consult factory if problem persists. |
42 |
Hi Temperature |
43 |
Temperature exceeding +80 C has been measured in the electronics compartment. |
44 |
Lo Temperature |
45 |
Temperature below -40 C has been measured in the electronics compartment. |
46 |
Calib Required |
47 |
Default calibration parameters are in use; level measurement may be inaccurate. The module requires re-calibration by the factory. |
48 |
EOP Too Low |
49 |
The End-of-Probe signal is lower than expected. Check configuration, especially Probe Model, Probe Length, Compensation, and Upper Dielectric. |
50 |
Trim Required |
51 |
Default values for Loop DAC trim parameters are being used; Loop output may be inaccurate. Perform trim loop current operation. |
52 |
Initializing |
53 |
Device is initializing. Initialization period persists until measurement compromising faults, if any, are cleared. |
54 |
May Be Flooded |
55 |
Possible flooded condition has been inferred from the loss of level signal from a low dielectric medium rising near the top of a twin rod probe. |
56 |
Dry Probe |
57 |
Level Signal from end of probe. Reported level has been calculated by setting distance equal to the probe length. Actual level may be lower than the end of the probe. |
58 |
Weak Signal |
59 |
CFD output is low indicating a weak signal reflection. Adjust Dielectric Range or Sensitivity. |
60 |
System Warning |
61 |
Unexpected but non-fatal software event has occurred. If warning recurs frequently, consult the factory. |
62 |
Warning 1 |
63 |
General Warning 1 |
64 |
Warning 2 |
65 |
General Warning 2 |
66 |
No Steam Target |
67 |
No Steam Target detected on Model 7xS probe. Steam compensation may not be operational. Consult factory for assistance. |
68 |
Warning 4 |
69 |
General Warning 4 |
70 |
QV is |
71 |
Trim Loop Current |
72 |
Loop Test |
73 |
in |
74 |
cm |
75 |
m |
76 |
ft |
77 |
Level Units |
78 |
Select units of measurement for level output. |
79 |
g |
80 |
l |
81 |
Volume Units |
82 |
Select units of measurement for volume output. |
83 |
.1f |
84 |
.2f |
85 |
.3f |
86 |
Level |
87 |
Level of material in vessel. When interface is present, this value corresponds to the level of the upper surface. |
88 |
Volume |
89 |
Total volume of material in vessel. |
90 |
IfcLvl |
91 |
Level of liquid-liquid interface. When no interface is present, this value corresponds to the level of the material in the vessel. |
92 |
IfcVol |
93 |
Volume of lower liquid in vessel. When no interface is present, this value corresponds to the total volume of material in the vessel. |
94 |
Level USL |
95 |
Level LSL |
96 |
Level Min Span |
97 |
Volume USL |
98 |
Volume LSL |
99 |
Volume Min Span |
100 |
4.1f |
101 |
Damping |
102 |
5.2f |
103 |
Loop |
104 |
22mA |
105 |
3.6mA |
106 |
Hold |
107 |
System Fault State |
108 |
System Fault State defines how the Analog Output will respond when the transmitter detects a system fault. |
109 |
5.1f |
110 |
% |
111 |
%Range |
112 |
Lvl 4mA Set Point |
113 |
Lvl 20mA Set Point |
114 |
Vol 4mA Set Point |
115 |
Vol 20mA Set Point |
116 |
7xA_x |
117 |
7xA: Coaxial Probe (dielectric range >= 1.4) |
118 |
7xB_x |
119 |
7xB: Rigid Twin Rod Probe (dielectric range >= 1.9) |
120 |
7xD_x |
121 |
7xD: High Temperature-High Pressure Coaxial Probe (dielectric range >= 1.4) |
122 |
7xE_x |
123 |
7xE: High Temperature-High Pressure Twin Rod Probe |
124 |
7xF_E |
125 |
7xF-E: Sanitary Single Rod Probe |
126 |
7xF_F |
127 |
7xF-F: Insulated Single Rod Probe with Faced Flange |
128 |
7xF_P |
129 |
7xF-P: Single Rod Paint Probe |
130 |
7xF_4 |
131 |
7xF-4: Insulated Single Rod Probe |
132 |
7xF_x |
133 |
7xF: Rigid Single Rod Probe |
134 |
7xG_x |
135 |
7xG: Overfill Single Rod Probe for Chambers |
136 |
7xJ_x |
137 |
7xJ: High Temperature-High Pressure Single Rod Probe |
138 |
7xK_x |
139 |
7xK: Top In/Bottom Out Coaxial Probe |
140 |
7xL_x |
141 |
7xL: High Temperature-High Pressure Coaxial Probe with Flushing Connection |
142 |
7xM_x |
143 |
7xM: Coaxial Probe for overfill applications with Flushing Connection |
144 |
7xN_x |
145 |
7xN: Coaxial Probe for interface applications with Flushing Connection |
146 |
7xP_x |
147 |
7xP: High Pressure Coaxial Probe (dielectric range >= 1.4) |
148 |
7xR_x |
149 |
7xR: Coaxial Probe for flooding applications (dielectric range >= 1.4) |
150 |
7xS_x |
151 |
7xS: High Temperature-High Pressure Coaxial Probe for steam (dielectric range >= 10.0) |
152 |
7xT_x |
153 |
7xT: Coaxial Probe for interface applications |
154 |
7x1_x |
155 |
7x1: Flexible Single Rod Probe |
156 |
7x2_x |
157 |
7x2: Single Rod Probe for Bulk Solids |
158 |
7x5_x |
159 |
7x5: Twin Rod Probe for Bulk Solids |
160 |
7x7_x |
161 |
7x7: Flexible Twin Rod Probe (dielectric range >= 1.9) |
162 |
Type 24 |
163 |
Type 25 |
164 |
Probe Model |
165 |
Type of probe connected to transmitter as shown on the probe nameplate. |
166 |
Probe Length |
167 |
Distance from face of process seal to end of probe. |
168 |
Level Offset |
169 |
Desired level reading when liquid surface is at end of probe. |
170 |
1.4 - 1.7 |
171 |
Dielectric constant of material lies between 1.4 and 1.7 |
172 |
1.7 - 3 |
173 |
Dielectric constant of material lies between 1.7 and 3 |
174 |
3 - 10 |
175 |
Dielectric constant of material lies between 3 and 10 |
176 |
10 - 100 |
177 |
Dielectric constant of material lies between 10 and 100 |
178 |
Dielectric Range |
179 |
Selects the range bounding the dielectric constant of the medium in vessel. For interface measurement mode, it selects the range bounding the dielectric constant of the lower liquid medium. Some ranges may not be selectable depending on the probe model. |
180 |
NPT |
181 |
US National tapered Pipe Thread |
182 |
BSP |
183 |
British Straight Pipe Thread |
184 |
Flange |
185 |
Welded ANSI, DIN, or sanitary Flange |
186 |
Type 4 |
187 |
Type 5 |
188 |
Type 6 |
189 |
Probe Mount |
190 |
Mounting style of probe fitting. |
191 |
Trim Level |
192 |
Trim Level is an offset value to be used to force the transmitter to output the exact Level. This should only be used after all parameters have been entered correctly, and it has been confirmed that the transmitter is tracking the correct level. |
193 |
Level Only |
194 |
Level and Volume |
195 |
Interface |
196 |
Ifc and Volume |
197 |
Measurement Type |
198 |
Selection of Measurement Type may be constrained by the probe model. |
199 |
New User Password |
200 |
Password required to change user parameter settings from keypad. |
201 |
3u |
202 |
Enter Password |
203 |
Certain parameters are password protected to prevent access by the user. Parameters that should never be changed by the user are protected with the Factory password. Some parameters intended for field use that may be changed in special, controlled situations require the Superuser password. |
204 |
Negative |
205 |
Positive |
206 |
Fiducial Type |
207 |
Fiducial Type can be used to override the default fiducial detection mode for some probe models. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
208 |
Fixed |
209 |
CFD |
210 |
Threshold |
211 |
Selects the negative signal threshold control for measuring the upper level pulse. Use 'CFD' except in situation with low dielectric (< 3) material over higher dielectric (> 10) material, e.g., water bottom or interface measurement, in which case use 'Fixed'. |
212 |
Window |
213 |
Factory parameter, displayed for diagnostic purposes only. |
214 |
Fiducial Gain |
215 |
Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
216 |
Sensitivity |
217 |
Sensitivity of instrument to surface of material in vessel. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
218 |
4u |
219 |
4mA Trim Value |
220 |
Displayed for diagnostic purposes only. |
221 |
20mA Trim Value |
222 |
Magnetrol S/N |
223 |
Magnetrol S/N - Serial number of this unit. |
224 |
Date/Time/Initials |
225 |
When and by whom calibration was performed. |
226 |
Firmware Version |
227 |
5i |
228 |
Conversion Factor |
229 |
4i |
230 |
Scale Offset |
231 |
Fiducial |
232 |
Waveform Selection |
233 |
Blocking Distance |
234 |
Distance below the reference point within which level is not sensed. Operation when the liquid level is within the blocking distance is undefined. |
235 |
Neg Threshold Ampl |
236 |
Threshold for detecting level. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
237 |
Pos Threshold Ampl |
238 |
Threshold for detecting end of probe. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
239 |
Ifc Threshold Ampl |
240 |
Threshold for detecting liquid-liquid interface. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
241 |
Intrface Threshold |
242 |
Selects signal threshold control for measuring liquid-liquid interface level. Use 'CFD' for sensing transition from lower to higher dielectric liquids. |
243 |
4.2f |
244 |
Upper Dielectric |
245 |
For manual compensation, the value entered is used as the dielectric constant of the vapor phase above a single liquid medium. For interface mode, the value entered is used as the dielectric constant of the upper liquid medium. |
246 |
None |
247 |
Manual |
248 |
Auto |
249 |
Compensation Mode |
250 |
Compensation is limited to None in interface mode. Manual uses the upper dielectric to compensate for a vapor phase above the liquid. Automatic is available only with a targeted probe. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
251 |
7xK Correction |
252 |
For 7xK probes, enter distance correction in mm. Value entered at factory is suitable for most installations. Refer to I & O manual before adjusting. |
253 |
5u |
254 |
Ticks |
255 |
Fiducial Ticks |
256 |
Level Ticks |
257 |
Interface Ticks |
258 |
Unknown |
259 |
Unrecognized waveform |
260 |
Oil Only |
261 |
Waveform indicates probe wetted with upper liquid only |
262 |
Thin Oil |
263 |
Probe senses a thin layer (< 15 cm) of upper liquid over lower liquid |
264 |
Thick Oil |
265 |
Probe senses a thick layer (> 15 cm) of upper liquid over lower liquid |
266 |
Waveform indicates no liquid on probe |
267 |
Interface Medium |
268 |
Characteristic of liquid(s) sensed by probe in interface measurement mode. |
269 |
Signal Strength |
270 |
°C |
271 |
Elec Temperature |
272 |
Present temperature measured in the electronics compartment. |
273 |
Max Temperature |
274 |
Maximum temperature measured in the electronics compartment. |
275 |
Min Temperature |
276 |
Minimum temperature measured in the electronics compartment. |
277 |
Latch 3.6 |
278 |
Latch 22 |
279 |
SZ Fault State |
280 |
Safety Zone Fault State defines how the Analog Output will respond when the transmitter detects a level within the Safety Zone. Enable fault to ensure safe, reliable high-level readings in critical applications. If Latch 3.6 or Latch 22 is chosen, the unit will remain in alarm until it is manually cleared with the SZ Alarm Reset. |
281 |
SZ Height |
282 |
Safety Zone Height - Enter a distance that defines a safety zone just below the Blocking Distance. If Safety Zone Fault is enabled, the transmitter will report a fault if the level rises into this zone. |
283 |
SZ Hysteresis |
284 |
Level must fall this distance below Safety Zone to clear a Safety Zone Fault. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
285 |
Derating Factor |
286 |
Run |
287 |
Cal |
288 |
Off |
289 |
Target Oper Mode |
290 |
Run mode enables automatic compensation. Cal mode allows calibration of target through air; Off mode disables automatic compensation. Default setting is suitable for most installations. Refer to I & O manual before adjusting. |
291 |
6i |
292 |
Targ Calib Value |
293 |
Target Ticks |
294 |
Target Amplitude |
295 |
HF Cable |
296 |
Remote HF Cable distance. Consult factory for assistance. |
297 |
Factory Param 2 |
298 |
OK |
299 |
Hi Volume Alarm |
300 |
Safety Zone Alarm |
301 |
Default Parameters |
302 |
Unknown Status |
303 |
Status |
304 |
Describes a fault or warning event. |
305 |
Run Time |
306 |
Elapsed time since power on. Also, run time is reset to zero when the Status History is cleared. |
307 |
Occurred |
308 |
The elapsed time when fault or warning occurred. |
309 |
Duration |
310 |
When a fault or warning is cleared, the difference between the present elapsed time and the time recorded when the event occurred is recorded as the duration. |
311 |
s |
312 |
h |
313 |
Strap Table Length |
314 |
Number of valid level-volume pairs in strapping table. A level-volume pair is valid if both values are greater than all preceding level-volume pairs. |
315 |
Pt01Lvl |
316 |
Strapping table level entry corresponding to VolPt01. |
317 |
Pt02Lvl |
318 |
Strapping table level entry corresponding to VolPt02. |
319 |
Pt03Lvl |
320 |
Strapping table level entry corresponding to VolPt03. |
321 |
Pt04Lvl |
322 |
Strapping table level entry corresponding to VolPt04. |
323 |
Pt05Lvl |
324 |
Strapping table level entry corresponding to VolPt05. |
325 |
Pt06Lvl |
326 |
Strapping table level entry corresponding to VolPt06. |
327 |
Pt07Lvl |
328 |
Strapping table level entry corresponding to VolPt07. |
329 |
Pt08Lvl |
330 |
Strapping table level entry corresponding to VolPt08. |
331 |
Pt09Lvl |
332 |
Strapping table level entry corresponding to VolPt09. |
333 |
Pt10Lvl |
334 |
Strapping table level entry corresponding to VolPt10. |
335 |
Pt11Lvl |
336 |
Strapping table level entry corresponding to VolPt11. |
337 |
Pt12Lvl |
338 |
Strapping table level entry corresponding to VolPt12. |
339 |
Pt13Lvl |
340 |
Strapping table level entry corresponding to VolPt13. |
341 |
Pt14Lvl |
342 |
Strapping table level entry corresponding to VolPt14. |
343 |
Pt15Lvl |
344 |
Strapping table level entry corresponding to VolPt15. |
345 |
Pt16Lvl |
346 |
Strapping table level entry corresponding to VolPt16. |
347 |
Pt17Lvl |
348 |
Strapping table level entry corresponding to VolPt17. |
349 |
Pt18Lvl |
350 |
Strapping table level entry corresponding to VolPt18. |
351 |
Pt19Lvl |
352 |
Strapping table level entry corresponding to VolPt19. |
353 |
Pt20Lvl |
354 |
Strapping table level entry corresponding to VolPt20. |
355 |
Pt01Vol |
356 |
Strapping table volume entry corresponding to LvlPt01. |
357 |
Pt02Vol |
358 |
Strapping table volume entry corresponding to LvlPt02. |
359 |
Pt03Vol |
360 |
Strapping table volume entry corresponding to LvlPt03. |
361 |
Pt04Vol |
362 |
Strapping table volume entry corresponding to LvlPt04. |
363 |
Pt05Vol |
364 |
Strapping table volume entry corresponding to LvlPt05. |
365 |
Pt06Vol |
366 |
Strapping table volume entry corresponding to LvlPt06. |
367 |
Pt07Vol |
368 |
Strapping table volume entry corresponding to LvlPt07. |
369 |
Pt08Vol |
370 |
Strapping table volume entry corresponding to LvlPt08. |
371 |
Pt09Vol |
372 |
Strapping table volume entry corresponding to LvlPt09. |
373 |
Pt10Vol |
374 |
Strapping table volume entry corresponding to LvlPt10. |
375 |
Pt11Vol |
376 |
Strapping table volume entry corresponding to LvlPt11. |
377 |
Pt12Vol |
378 |
Strapping table volume entry corresponding to LvlPt12. |
379 |
Pt13Vol |
380 |
Strapping table volume entry corresponding to LvlPt13. |
381 |
Pt14Vol |
382 |
Strapping table volume entry corresponding to LvlPt14. |
383 |
Pt15Vol |
384 |
Strapping table volume entry corresponding to LvlPt15. |
385 |
Pt16Vol |
386 |
Strapping table volume entry corresponding to LvlPt16. |
387 |
Pt17Vol |
388 |
Strapping table volume entry corresponding to LvlPt17. |
389 |
Pt18Vol |
390 |
Strapping table volume entry corresponding to LvlPt18. |
391 |
Pt19Vol |
392 |
Strapping table volume entry corresponding to LvlPt19. |
393 |
Pt20Vol |
394 |
Strapping table volume entry corresponding to LvlPt20. |
395 |
No Target detected |
396 |
Comp Mode not AUTO or a measurement fault exists |
397 |
In RUN mode or a measurement fault exists |
398 |
No earlier event |
399 |
No later event |
400 |
Level point value too high |
401 |
Level point value too low |
402 |
Volume point value too high |
403 |
Volume point value too low |
404 |
Lvl and Vol values out of range |
405 |
Main Menu |
406 |
Device Variables |
407 |
Device Setup |
408 |
Calibration |
409 |
Volume Parameters |
410 |
Strapping Table |
411 |
Variable Selection |
412 |
Interface Params |
413 |
Basic Setup |
414 |
Advanced Setup |
415 |
Compensation |
416 |
Target Calibration |
417 |
Factory Settings |
418 |
Diagnostics |
419 |
Present Status |
420 |
Status History |
421 |
Review |
422 |
QV |
423 |
Min Span |
424 |
20mA Set Point |
425 |
4mA Set Point |
426 |
SZ Alarm Reset |
427 |
Clear a latched Safety Zone Alarm |
428 |
Reset Temperatures |
429 |
Reset the maximum and minimum temperatures measured in the electronics compartment to the present temperature. |
430 |
Reset History |
431 |
Clears entries in Status History and resets the run time to zero. |
432 |
View History |
433 |
Displays current status and run time as well as a history of most recent faults and warnings. For each event, the run time when the event occurred and the duration of the event are shown. The run time is the time since power up or a history reset. |
434 |
Auto Target Calib |
435 |
Captures time-of-flight to target on steam probe types at room ambient. |
436 |
Targ Cal Abort |
437 |
FMU860 |
438 |
DT |
439 |
DVC |
440 |
FCX |
441 |
AS800 |
442 |
340T |
443 |
340C |
444 |
344 |
445 |
Siemens Micro K |
446 |
SITRANS L |
447 |
SIPAN PH |
448 |
SITRANS P |
449 |
75SMT |
450 |
UNIDELTA |
451 |
8320 |
452 |
SX5100 |
453 |
TSV175 |
454 |
DMU |
455 |
3780 |
456 |
BM70 |
457 |
8u |
458 |
Are You Sure? |
459 |
No;Yes |
460 |
Current Status:
%{0}
Run Time: %[.0f]{1} %{2} |
461 |
Current Status:
%{0}
Run Time: %[.1f]{1} %{2} |
462 |
No events in Status History. |
463 |
Status Event:
%{0}
Occurred: %[.0f]{3} %{4}
Code: %[.0f]{5} |
464 |
Get earlier event;Get later event |
465 |
Status Event:
%{0}
Occurred: %[.1f]{3} %{4}
Code: %[.0f]{5} |
466 |
Status Event:
%{0}
Occurred: %[.0f]{3} %{4}
Duration: %[.0f]{5} %{6} |
467 |
Status Event:
%{0}
Occurred: %[.0f]{3} %{4}
Duration: %[.1f]{5} %{6} |
468 |
Status Event:
%{0}
Occurred: %[.1f]{3} %{4}
Duration: %[.0f]{5} %{6} |
469 |
Status Event:
%{0}
Occurred: %[.1f]{3} %{4}
Duration: %[.1f]{5} %{6} |
470 |
Target Ticks = %[4i]{0} |
471 |
Calibration ok
Set Targ Air = %[4i]{0}? |
472 |
Calibration bad
TargTks unstable |