0 |
LDO2 |
1 |
mg/L |
2 |
Temperature Error |
3 |
Sample Warning |
4 |
? CAL Warning |
5 |
? Meas. Warning |
6 |
Safety Warning |
7 |
Reagent Warning |
8 |
Maint. Required Warning |
9 |
Meas. CAL Error |
10 |
Electronic Adjust Error |
11 |
Cleaning Error |
12 |
Measuring Module Error |
13 |
Re-initialization Error |
14 |
Hardware Error |
15 |
Internal COMM Error |
16 |
Humidity Error |
17 |
RED AMPL LOW |
18 |
The sensor may be defective or the sensor cap is not installed properly. |
19 |
BLUE AMPL LOW |
20 |
CLEAN SENSOR |
21 |
REPLACE SENSOR |
22 |
CAL IN PROGRESS |
23 |
DEF CAP LOT |
24 |
The cap code is corrupt. Complete setup cap procedure. |
25 |
EE SETUP ERR |
26 |
The EEPROM is corrupt. Contact the manufacturer. |
27 |
EE RSRVD ERR` |
28 |
TEMP < 0C |
29 |
The process temperature is below 0 C (32 F). |
30 |
TEMP > 50C |
31 |
The process temperature is above 50 C (120 F). |
32 |
Value falls below 0.03. |
33 |
RED AMPL HIGH |
34 |
Value is greater than 0.35. |
35 |
BLUE AMPL High |
36 |
14.4f |
37 |
DO mg/L |
38 |
DO ppm |
39 |
% |
40 |
Temp F |
41 |
Temp C |
42 |
Temp C |
43 |
Temp Units |
44 |
Temperature |
45 |
Temp USL |
46 |
Temp LSL |
47 |
Temp Min Span |
48 |
Temp Damping |
49 |
Temp Class |
50 |
Temp Data Quality |
51 |
Temp Limit Status |
52 |
Temp Family Status |
53 |
Temp Sensor S/N |
54 |
Temp F |
55 |
Temp Min. Span |
56 |
Dissolved Oxygen |
57 |
DO Units |
58 |
DO Value |
59 |
DO USL |
60 |
DO LSL |
61 |
DO Min. Span |
62 |
DO Damping |
63 |
DO Class |
64 |
DO Data Quality |
65 |
DO Limit Status |
66 |
DO Family Status |
67 |
DO Sensor S/N |
68 |
Percent Saturation |
69 |
Saturation Units |
70 |
Saturation Value |
71 |
Saturation USL |
72 |
Saturation LSL |
73 |
Saturation Min. Span |
74 |
Saturation Damping |
75 |
Temp C |
76 |
Temperature Units C. |
77 |
Temp F |
78 |
Temperature Units F. |
79 |
The measured value of the process solution temperature in Celsius units. |
80 |
5.1f |
81 |
The measured value of the process solution temperature in Fahrenheit units. |
82 |
8d |
83 |
Dissolved Oxygen Class |
84 |
Dissolved Oxygen Data Quality |
85 |
Dissolved Oxygen Limit Status |
86 |
Dissolved Oxygen Family Status |
87 |
Dissolved Oxygen Sensor S/N |
88 |
Dissolved Oxygen Unit ppm. |
89 |
The measured value of the process solution Dissolved Oxygen in ppm. |
90 |
Dissolved Oxygen USL |
91 |
Dissolved Oxygen LSL |
92 |
DO Min Span |
93 |
Dissolved Oxygen Damping |
94 |
Dissolved Oxygen Unit mg/L. |
95 |
The measured value of the process solution Dissolved Oxygen in mg/L. |
96 |
Dissolved Oxygen Min Span |
97 |
Dissolved Oxygen Percent Saturation Unit %. |
98 |
The measured value of the process solution Dissolved Oxygen as Percent Saturation in %. |
99 |
Percent Sat. USL |
100 |
Percent Sat. LSL |
101 |
Percent Sat. Min Span |
102 |
Percent Sat. Damping |
103 |
Location String |
104 |
Location or Name of Sensor Module. |
105 |
Appl. Code Ver. |
106 |
Application Code Version times 100. |
107 |
Boot Code Ver. |
108 |
Boot Code Version times 100. |
109 |
Serial Number |
110 |
Sensor module serial number. |
111 |
Pressure |
112 |
Sets the value of the atmospheric pressure for the geographic location where the sensor is used (default: 760 mmHg). This value is necessary for calibration when the Air Cal or Saturation Cal procedure is used. |
113 |
7.0f |
114 |
Altitude |
115 |
Sets the value of the altitude for the geographic location where the sensor is used. This value is necessary for calibration when the Air Cal or Saturation Cal procedure is used. |
116 |
Sets the value of the altitude for the geographic location where the sensor is used. This value is necessary for calibration when the Air Cal or Saturation Cal procedure is used. |
117 |
Sets the value of the atmospheric pressure for the geographic location where the sensor is used. This value is necessary for calibration when the Air Cal or Saturation Cal procedure is used. |
118 |
Salinity |
119 |
Sets the salinity value of the water used for calibration. This value is necessary when the Saturation Cal procedure is used. It has the units of parts per thousand. |
120 |
6.2f |
121 |
0/00 |
122 |
Offset Correction |
123 |
Sets the calibration offset. Range: -3.0 to 3.0. |
124 |
5.2f |
125 |
Sets the calibration offset. Range: -50.0 to 50.0. |
126 |
Gain Correction |
127 |
Sets the calibration offset. Range: 0.000 to 3.0. |
128 |
Total Phase Shift |
129 |
For diagnostic information only. |
130 |
Blue Phase Shift |
131 |
Red Phase Shift |
132 |
Blue Amplitude |
133 |
Red Amplitude |
134 |
Temp Units C/F |
135 |
Temperature Units C/F. |
136 |
Signal Avg. |
137 |
Signal average applied to the measured value. |
138 |
3d |
139 |
Pressure/Altitude Units |
140 |
Units of pressure/altitude selection. |
141 |
Measurement Units |
142 |
Units of Dissolved Oxygen selection. |
143 |
ppm |
144 |
Log Interval |
145 |
The interval at which data will be written to the sensor data log. |
146 |
30 sec. |
147 |
1 Min. |
148 |
2 Min. |
149 |
5 Min. |
150 |
10 Min. |
151 |
15 Min. |
152 |
30 Min. |
153 |
60 Min. |
154 |
Cleaning Interval |
155 |
The interval for the sensor cleaning in days. |
156 |
2d |
157 |
Sensor Life |
158 |
The number of days until the next scheduled sensor cap replacement. |
159 |
4d |
160 |
Days to Clean |
161 |
The number of days until the next scheduled cleaning. |
162 |
Cap Lot Code |
163 |
The Lot Code for the sensor cap. |
164 |
6d |
165 |
Driver Version |
166 |
The sensor Device Driver Version. |
167 |
Process Events |
168 |
Process Event Flags. |
169 |
Normal Operation |
170 |
Hold Outputs |
171 |
Active Outputs |
172 |
Transfer Value |
173 |
Classified Status 1 |
174 |
Status Flags for Measurements 0-3. |
175 |
CAL in Progress |
176 |
The device has been placed in a calibration mode. The measurements may not be valid. |
177 |
Cleaning In Progress |
178 |
The device has been placed in a cleaning mode. The measurements may not be valid. |
179 |
Service/Maint. Menu |
180 |
The device has been placed in a service or maintenance mode in which the measurements may not be valid. |
181 |
Common Error |
182 |
Meas. 0 Quality Bad |
183 |
Meas. 0 Low Limit |
184 |
Meas. 0 High Limit |
185 |
Meas. 1 Quality Bad |
186 |
Meas. 1 Low Limit |
187 |
Meas. 1 High Limit |
188 |
Meas. 2 Quality Bad |
189 |
Meas. 2 Low Limit |
190 |
Meas. 2 High Limit |
191 |
Meas. 3 Quality Bad |
192 |
Meas. 3 Low Limit |
193 |
Meas. 3 High Limit |
194 |
Classified Status 2 |
195 |
Status Flags for Measurements 4-7. |
196 |
Reserved for Future Use |
197 |
Meas. 4 Quality Bad |
198 |
Meas. 4 Low Limit |
199 |
Meas. 4 High Limit |
200 |
Meas. 5 Quality Bad |
201 |
Meas. 5 Low Limit |
202 |
Meas. 5 High Limit |
203 |
Meas. 6 Quality Bad |
204 |
Meas. 6 Low Limit |
205 |
Meas. 6 High Limit |
206 |
Meas. 7 Quality Bad |
207 |
Meas. 7 Low Limit |
208 |
Meas. 7 High Limit |
209 |
Classified Status 3 |
210 |
Status Flags for Measurements 8-11. |
211 |
Meas. 8 Quality Bad |
212 |
Meas. 8 Low Limit |
213 |
Meas. 8 High Limit |
214 |
Meas. 9 Quality Bad |
215 |
Meas. 9 Low Limit |
216 |
Meas. 9 High Limit |
217 |
Meas. 10 Quality Bad |
218 |
Meas. 10 Low Limit |
219 |
Meas. 10 High Limit |
220 |
Meas. 11 Quality Bad |
221 |
Meas. 11 Low Limit |
222 |
Meas. 11 High Limit |
223 |
Classified Status 4 |
224 |
Status Flags for Measurements 12-15. |
225 |
Meas. 12 Quality Bad |
226 |
Meas. 12 Low Limit |
227 |
Meas. 12 high Limit |
228 |
Meas. 13 Quality Bad |
229 |
Meas. 13 Low Limit |
230 |
Meas. 13 High Limit |
231 |
Meas. 14 Quality Bad |
232 |
Meas. 14 Low Limit |
233 |
Meas. 14 High Limit |
234 |
Meas. 15 Quality Bad |
235 |
Meas. 15 Low Limit |
236 |
Meas. 15 High Limit |
237 |
DO Units Select |
238 |
This method selects the Dissolved Oxygen measurement units. |
239 |
DO Trim Adjust |
240 |
This trim selects units for altitude/pressure and edits values. |
241 |
Salinity/Cond. Trim |
242 |
This trim adjusts the Salinity value by means of an entered Conductivity value in mS/cm. |
243 |
DO Salinity Adjust |
244 |
This trim edits the salinity value. |
245 |
Hach-Lange Logo |
246 |
Sensor Image |
247 |
Dissolved Oxygen mg/L |
248 |
Dissolved Oxygen ppm |
249 |
Percent Saturation % |
250 |
Measured Values |
251 |
% Saturation |
252 |
LDO2 sc |
253 |
Device Variables |
254 |
Identification |
255 |
Sensor Description |
256 |
LDO2 - Luminescent DO |
257 |
Primary Variable |
258 |
Measured Variables |
259 |
HART Tags |
260 |
Device Info. |
261 |
Revisions |
262 |
Sensor Information |
263 |
Status |
264 |
Diagnostics |
265 |
Classified Error |
266 |
Device Errors |
267 |
Device Warnings |
268 |
Sensor Diagnostics |
269 |
LDO2 Dissolved Oxygen Probe |
270 |
Maintenance |
271 |
Sensor Maintenance |
272 |
Sensor Setup |
273 |
Temperature Setup |
274 |
HART Setup |
275 |
DO Setup |
276 |
Pressure / Altitude Setup |
277 |
Salinity Setup |
278 |
Upload Variables |
279 |
Sensor Identification |
280 |
LDO2 Setup |
281 |
Loop Current A |
282 |
Alarm Type Loop A |
283 |
Channel A Flags |
284 |
Loop Current B |
285 |
Alarm Type Loop B |
286 |
Channel B Flags |
287 |
Loop Current C |
288 |
Loop Current 1 |
289 |
Loop Current 2 |
290 |
Loop 1 Current |
291 |
mA |
292 |
.3f |
293 |
Loop 2 Current |
294 |
Loop A Current |
295 |
Alarm A Select |
296 |
Loop A Flags |
297 |
Loop B Current |
298 |
Alarm B Select |
299 |
Loop B Flags |
300 |
Loop C Current |
301 |
Alarm C Select |
302 |
Loop C Flags |
303 |
LRV Loop A |
304 |
Lower Range Value Loop A. |
305 |
LRV Loop B |
306 |
Lower Range Value Loop B. |
307 |
LRV Loop C |
308 |
Lower Range Value Loop C. |
309 |
URV Loop A |
310 |
Upper Range Value Loop A. |
311 |
URV Loop B |
312 |
Upper Range Value Loop B. |
313 |
URV Loop C |
314 |
Upper Range Value Loop C. |
315 |
Transfer Val Loop A |
316 |
Sets the transfer value. Range 3.0 to 23.0 mA (default 4.000). |
317 |
Transfer Val Loop B |
318 |
Transfer Val Loop C |
319 |
Test Value Loop A. |
320 |
Test Value Loop B. |
321 |
Test Value Loop C. |
322 |
LSL Loop A |
323 |
Lower Sensor Limit Loop A. |
324 |
LSL Loop B |
325 |
Lower Sensor Limit Loop B. |
326 |
LSL Loop C |
327 |
Lower Sensor Limit Loop C. |
328 |
USL Loop A |
329 |
Upper Sensor Limit Loop A. |
330 |
USL Loop B |
331 |
Upper Sensor Limit Loop B. |
332 |
USL Loop C |
333 |
Upper Sensor Limit Loop C. |
334 |
Error Hold Mode |
335 |
Holds Outputs at last known value or switches to Transfer mode when sensor loses communication with controller. |
336 |
Transfer to defined Value |
337 |
Meas. Select Loop A |
338 |
Selection list of sensor measurements. |
339 |
DO % |
340 |
Meas. Select Loop B |
341 |
Meas. Select Loop C |
342 |
Filter Loop A |
343 |
Loop Filter Time Constant. |
344 |
5d |
345 |
Filter Loop B |
346 |
Filter Loop C |
347 |
4mA Cal Count A |
348 |
Output Loop 4mA calibration value in D/A counts. |
349 |
Counts |
350 |
4mA Cal Count B |
351 |
4mA Cal Count C |
352 |
20mA Cal Count A |
353 |
Output Loop 20mA calibration value in D/A counts. |
354 |
20mA Cal Count B |
355 |
20mA Cal Count C |
356 |
Test Enable Loop A |
357 |
Disabled |
358 |
Enabled |
359 |
Test Enable Loop B |
360 |
Test Enable Loop C |
361 |
User Hold Loop A |
362 |
Selection List of user Hold Modes. |
363 |
User Hold Loop B |
364 |
User Hold Loop C |
365 |
Units Loop A |
366 |
Engineering Units Loop A. |
367 |
Units Loop B |
368 |
Engineering Units Loop B. |
369 |
Units Loop C |
370 |
Engineering Units Loop C. |
371 |
HART Loop A Mode |
372 |
HART Loop Modes. Fixed Current = 4 mA. Active Current = 4-20 mA. |
373 |
Fixed Current |
374 |
Active Current |
375 |
HART Loop B Mode |
376 |
HART Loop A Address |
377 |
HART Loop Address (0-63) 0->Single Loop. |
378 |
HART Loop B Address |
379 |
Mapping Relay A |
380 |
Status of Relay A mapping to HART sensor. |
381 |
Inactive |
382 |
Active |
383 |
Mapping Relay B |
384 |
Status of Relay B mapping to HART sensor. |
385 |
Mapping Relay C |
386 |
Status of Relay C mapping to HART sensor. |
387 |
Mapping Relay D |
388 |
Status of Relay D mapping to HART sensor. |
389 |
Mapping Output 1 |
390 |
Status of sc200 Output 1 mapping to HART sensor. |
391 |
Mapping Output 2 |
392 |
Status of sc200 Output 2 mapping to HART sensor. |
393 |
Mapping Output A |
394 |
Status of HART Output A mapping to HART sensor. |
395 |
Mapping Output B |
396 |
Status of HART Output B mapping to HART sensor. |
397 |
Mapping Output C |
398 |
Status of HART Output C mapping to HART sensor. |
399 |
HART Channel A |
400 |
Status of HART Channel A mapping to HART sensor. |
401 |
HART Channel B |
402 |
Status of HART Channel B mapping to HART sensor. |
403 |
Status sc200 |
404 |
Status of Device Mapping sc200. |
405 |
Location or Name of HART Module. |
406 |
Application Code Version. |
407 |
Boot Code Version. |
408 |
HART Module serial number. |
409 |
Date |
410 |
Sets the controller date. |
411 |
Time |
412 |
Sets the controller time. |
413 |
Fix analog output A |
414 |
Fix analog output B |
415 |
Fix analog output C |
416 |
Fix analog output 1 |
417 |
Fix analog output 2 |
418 |
Trim analog output A |
419 |
Trim analog output B |
420 |
Trim analog output C |
421 |
Trim analog output 1 |
422 |
Trim analog output 2 |
423 |
Loop A Gauge |
424 |
Loop B Gauge |
425 |
Loop C Gauge |
426 |
Start |
427 |
Offline |
428 |
Diagnostics / Maintenance |
429 |
HART Network Module |
430 |
Loop Currents |
431 |
HART Module Identification |
432 |
Loop A |
433 |
Loop B |
434 |
Loop C |
435 |
Loop Maintenance |
436 |
Device Mapping |
437 |
Status of Relays and Outputs Mapped to the HART sensor. |
438 |
Loop A Maintenance |
439 |
Loop B Maintenance |
440 |
Loop C Maintenance |
441 |
HART Network Module Settings |
442 |
Controller Date and Time Settings |
443 |
Output A Setup |
444 |
Output B Setup |
445 |
Output C Setup |
446 |
Hold Modes |
447 |
Location |
448 |
sc200 name or location. |
449 |
sc200 controller serial number. |
450 |
sc200 controller device warning flags. |
451 |
Overfeed Timer |
452 |
The Overfeed Timer is active. |
453 |
Restart Software |
454 |
Restart to Accept sc200 Software. |
455 |
Network COMM Loss |
456 |
Network Communications Loss. |
457 |
Input 1 triggered |
458 |
Discrete Input 1 Triggered. |
459 |
Input 2 triggered |
460 |
Discrete Input 2 Triggered. |
461 |
Input 3 triggered |
462 |
Discrete Input 3 Triggered. |
463 |
sc200 controller device error flags. |
464 |
Flash Failure |
465 |
Transfer Output 1 |
466 |
Sets the transfer value of Output 1. |
467 |
Transfer Output 2 |
468 |
Sets the transfer value of Output 2. |
469 |
LRV Output 1 |
470 |
Sets the minimum output value. |
471 |
LRV Output 2 |
472 |
URV Output 1 |
473 |
Sets the maximum output value. |
474 |
URV Output 2 |
475 |
Knee Value Out 1 |
476 |
Bilinear knee value. |
477 |
Knee Value Out 2 |
478 |
Knee Current Out 1 |
479 |
Bilinear knee current in percent. |
480 |
Knee Current Out 2 |
481 |
50% Log Output 1 |
482 |
Log 50% setting. |
483 |
50% Log Output 2 |
484 |
PID Man Set Out 1 |
485 |
Sets the PID manual output value (0.0 to 100.0 %). |
486 |
PID Man Set Out 2 |
487 |
PID Setpt Out 1 |
488 |
Sets the PID control setpoint for Output 1. |
489 |
PID Setpt Out 2 |
490 |
Sets the PID control setpoint for Output 2. |
491 |
PID Prop Band Out 1 |
492 |
Sets the PID proportional band for Output 1. |
493 |
PID Prop Band Out 2 |
494 |
Sets the PID proportional band for Output 2. |
495 |
Transfer Outputs to Predefined Value |
496 |
Meas. Select Output 1 |
497 |
Meas. Select Output 2 |
498 |
Type Output 1 |
499 |
Selection list of output modes. |
500 |
Linear |
501 |
PID |
502 |
Logarithmic |
503 |
Bilinear |
504 |
Type Output 2 |
505 |
Filter Output 1 |
506 |
Time constant Output 1. |
507 |
Filter Output 2 |
508 |
Time constant Output 2. |
509 |
0/4 mA Sel Out 1 |
510 |
Minimum current selection Output 1. |
511 |
0 mA |
512 |
4 mA |
513 |
0/4 mA Sel Out 2 |
514 |
Minimum current selection Output 2. |
515 |
PID Mode Output 1 |
516 |
Auto/Manual mode selection for Output 1. |
517 |
Auto |
518 |
Manual |
519 |
PID Mode Output 2 |
520 |
Auto/Manual mode selection for Output 2. |
521 |
Phase Output 1 |
522 |
Direct/Reverse phase selection Output 1. |
523 |
Direct |
524 |
Reverse |
525 |
Phase Output 2 |
526 |
Direct/Reverse phase selection Output 2. |
527 |
PID Integral Out 1 |
528 |
Integral time in seconds Output 1. |
529 |
PID Integral Out 2 |
530 |
Integral time in seconds Output 2. |
531 |
PID Derivative Out 1 |
532 |
Derivative time in seconds Output 1. |
533 |
PID Derivative Out 2 |
534 |
Derivative time in seconds Output 2. |
535 |
PID Transit Out 1 |
536 |
Stops all PID control for a selected period of time. |
537 |
PID Transit Out 2 |
538 |
Range Output 1 |
539 |
Selection List of sensor ranges. |
540 |
Range 1 |
541 |
Range Output 2 |
542 |
High Alarm A |
543 |
Sets the High Alarm setpoint for Alarm A. |
544 |
High Alarm B |
545 |
Sets the High Alarm setpoint for Alarm B. |
546 |
High Alarm C |
547 |
Sets the High Alarm setpoint for Alarm C. |
548 |
High Alarm D |
549 |
Sets the High Alarm setpoint for Alarm D. |
550 |
Low Alarm A |
551 |
Sets the Low Alarm setpoint for Alarm A. |
552 |
Low Alarm B |
553 |
Sets the Low Alarm setpoint for Alarm B. |
554 |
Low Alarm C |
555 |
Sets the Low Alarm setpoint for Alarm C. |
556 |
Low Alarm D |
557 |
Sets the Low Alarm setpoint for Alarm D. |
558 |
High Deadband A |
559 |
Sets the High Alarm deadband for Alarm A. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
560 |
High Deadband B |
561 |
Sets the High Alarm deadband for Alarm B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
562 |
High Deadband C |
563 |
Sets the High Alarm deadband for Alarm C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
564 |
High Deadband D |
565 |
Sets the High Alarm deadband for Alarm D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
566 |
Low Deadband A |
567 |
Sets the Low Alarm deadband for Alarm A. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
568 |
Low Deadband B |
569 |
Sets the Low Alarm deadband for Alarm B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
570 |
Low Deadband C |
571 |
Sets the Low Alarm deadband for Alarm C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
572 |
Low Deadband D |
573 |
Sets the Low Alarm deadband for Alarm D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint |
574 |
Setpoint Control A |
575 |
Sets the Controller Setpoint for Relay A. |
576 |
Setpoint Control B |
577 |
Sets the Controller Setpoint for Relay B. |
578 |
Setpoint Control C |
579 |
Sets the Controller Setpoint for Relay C. |
580 |
Setpoint Control D |
581 |
Sets the Controller Setpoint for Relay D. |
582 |
Deadband Control A |
583 |
Proportional Band A |
584 |
Sets the Controller Deadband/Proportional Band for Relay A. |
585 |
Deadband Control B |
586 |
Proportional Band B |
587 |
Sets the Controller Deadband/Proportional Band for Relay B. |
588 |
Deadband Control C |
589 |
Proportional Band C |
590 |
Sets the Controller Deadband/Proportional Band for Relay C. |
591 |
Deadband Control D |
592 |
Proportional Band D |
593 |
Sets the Controller Deadband/Proportional Band for Relay D. |
594 |
Setpoint Event A |
595 |
Sets the Event Controller Setpoint for Relay A. |
596 |
Setpoint Event B |
597 |
Sets the Event Controller Setpoint for Relay B. |
598 |
Setpoint Event C |
599 |
Sets the Event Controller Setpoint for Relay C. |
600 |
Setpoint Event D |
601 |
Sets the Event Controller Setpoint for Relay D. |
602 |
Deadband A |
603 |
Dead Zone A |
604 |
Sets the Controller Deadband/Dead Zone for Relay B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal. |
605 |
Deadband B |
606 |
Dead Zone B |
607 |
Sets the Controller Deadband/Dead Zone for Relay B. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal |
608 |
Deadband C |
609 |
Dead Zone C |
610 |
Sets the Controller Deadband/Dead Zone for Relay C. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal |
611 |
Deadband D |
612 |
Dead Zone D |
613 |
Sets the Controller Deadband/Dead Zone for Relay D. Deadband is a hysteresis so the relay will not swing unregulated when the process value converges to the setpoint. Dead Zone is a band around the setpoint. In this band the controller does not change the output signal |
614 |
Manual Value A |
615 |
Sets the Manual control output level (%) Relay A. |
616 |
Manual Value B |
617 |
Sets the Manual control output level (%) Relay B. |
618 |
Manual Value C |
619 |
Sets the Manual control output level (%) Relay C. |
620 |
Manual Value D |
621 |
Sets the Manual control output level (%) Relay D. |
622 |
Pulse Width A |
623 |
Period A |
624 |
Sets the period/pulse width for pwm/freq control Relay A. |
625 |
4.1f |
626 |
Pulse Width B |
627 |
Period B |
628 |
Sets the period/pulse width for pwm/freq control Relay B. |
629 |
Pulse Width C |
630 |
Period C |
631 |
Sets the period/pulse width for pwm/freq control Relay C. |
632 |
Pulse Width D |
633 |
Period D |
634 |
Sets the period/pulse width for pwm/freq control Relay 4. |
635 |
Min Pulse Width A |
636 |
Min Pulses A |
637 |
Sets the minimum pulse width for PWM in seconds (0.1 to Max Pulse Width) or sets the minimum number of pulses for Frequency control (0.001 to Max Number of Pulses) for Relay A. |
638 |
6.3f |
639 |
Min Pulse Width B |
640 |
Min Pulses B |
641 |
Sets the minimum pulse width for PWM in seconds (0.1 to Max Pulse Width) or sets the minimum number of pulses for Frequency control (0.001 to Max Number of Pulses) for Relay B. |
642 |
Min Pulse Width C |
643 |
Min Pulses C |
644 |
Sets the minimum pulse width for PWM in seconds (0.1 to Max Pulse Width) or sets the minimum number of pulses for Frequency control (0.001 to Max Number of Pulses) for Relay C. |
645 |
Min Pulse Width D |
646 |
Min Pulses D |
647 |
Sets the minimum pulse width for PWM in seconds (0.1 to Max Pulse Width) or sets the minimum number of pulses for Frequency control (0.001 to Max Number of Pulses) for Relay D. |
648 |
Max Pulse Width A |
649 |
Max Pulses A |
650 |
Sets the maximum pulse width for PWM in seconds (Min Pulse Width to Period) or sets the maximum number of pulses for Frequency control (Min Number of Pulses to 60 ) for Relay A. |
651 |
Max Pulse Width B |
652 |
Max Pulses B |
653 |
Sets the maximum pulse width for PWM in seconds (Min Pulse Width to Period) or sets the maximum number of pulses for Frequency control (Min Number of Pulses to 60 ) for Relay B. |
654 |
Max Pulse Width C |
655 |
Max Pulses C |
656 |
Sets the maximum pulse width for PWM in seconds (Min Pulse Width to Period) or sets the maximum number of pulses for Frequency control (Min Number of Pulses to 60 ) for Relay C. |
657 |
Max Pulse Width D |
658 |
Max Pulses D |
659 |
Sets the maximum pulse width for PWM in seconds (Min Pulse Width to Period) or sets the maximum number of pulses for Frequency control (Min Number of Pulses to 60 ) for Relay D. |
660 |
Calculated Value |
661 |
sc200 Calculated Value |
662 |
Language |
663 |
sc200 language selection list. |
664 |
English |
665 |
German |
666 |
Spanish |
667 |
French |
668 |
Italian |
669 |
Netherland |
670 |
Danish |
671 |
Swedish |
672 |
Chinese |
673 |
Polish |
674 |
Japanese |
675 |
Korean |
676 |
Portuguese |
677 |
Russian |
678 |
Hungarian |
679 |
Romanian |
680 |
Czech |
681 |
Greek |
682 |
Slovenian |
683 |
Polish UTF8 |
684 |
Date Format |
685 |
sc200 Date Format. |
686 |
YYYY-MM-DD 24h |
687 |
YYYY-MM-DD 12h |
688 |
DD/MM/YYYY 24h |
689 |
DD/MM/YYYY 12h |
690 |
MM/DD/YYYY 24h |
691 |
MM/DD/YYYY 12h |
692 |
Display Contrast |
693 |
sc200 Display Contrast, 1 to 9. |
694 |
Calc Log Mode |
695 |
sc200 Calculated Value logging mode |
696 |
Snap Shot |
697 |
Average |
698 |
Maximum |
699 |
Minimum |
700 |
Calc Log Interval |
701 |
sc200 Calculated Value logging interval |
702 |
5 sec. |
703 |
Calc Math Formula |
704 |
The formula selection for the Calculated Value. |
705 |
None |
706 |
X/Y |
707 |
X+Y |
708 |
X-Y |
709 |
(X/Y)% |
710 |
(X/Y)/2 |
711 |
X*Y |
712 |
(X-Y)%/X |
713 |
Discrete Input 1 |
714 |
Input State 1 |
715 |
Discrete Input 2 |
716 |
Input State 2 |
717 |
Discrete Input 3 |
718 |
Input State 3 |
719 |
Modbus Stats Clear |
720 |
Clears statistics for Modbus communication. |
721 |
Normal |
722 |
Clear |
723 |
Measurement A |
724 |
Measurement selection list for Relay A. |
725 |
Measurement B |
726 |
Measurement selection list for Relay B. |
727 |
Measurement C |
728 |
Measurement selection list for Relay C. |
729 |
Measurement D |
730 |
Measurement selection list for Relay D. |
731 |
Type A |
732 |
Type selection list for Relay A. |
733 |
Alarm |
734 |
Feeder Control |
735 |
Warning |
736 |
Event Control |
737 |
PWM Control |
738 |
Freq Control |
739 |
Type B |
740 |
Type selection list for Relay B. |
741 |
Type C |
742 |
Type selection list for Relay C. |
743 |
Type D |
744 |
Type selection list for Relay D. |
745 |
Transfer Set A |
746 |
State selection for Relay A. |
747 |
Transfer Set B |
748 |
State selection for Relay B. |
749 |
Transfer Set C |
750 |
State selection for Relay C. |
751 |
Transfer Set D |
752 |
State selection for Relay D. |
753 |
Alarm On Delay A |
754 |
Turn on time delay for Relay A. |
755 |
Alarm On Delay B |
756 |
Turn on time delay for Relay B. |
757 |
Alarm On Delay C |
758 |
Turn on time delay for Relay C. |
759 |
Alarm On Delay D |
760 |
Turn on time delay for Relay D. |
761 |
Alarm Off Delay A |
762 |
Turn off time delay for Relay A. |
763 |
Alarm Off Delay B |
764 |
Turn off time delay for Relay B. |
765 |
Alarm Off Delay C |
766 |
Turn off time delay for Relay C. |
767 |
Alarm Off Delay D |
768 |
Turn off time delay for Relay D. |
769 |
Phase Control A |
770 |
Select direct/reverse for Relay A. |
771 |
Low |
772 |
High |
773 |
Phase Control B |
774 |
Select direct/reverse for Relay B. |
775 |
Phase Control C |
776 |
Select direct/reverse for Relay C. |
777 |
Phase Control D |
778 |
Select direct/reverse for Relay D. |
779 |
Overfeed Timer A |
780 |
Sets a time period for de-activating an active relay if the process setpoint cannot be reached. Once an overfeed alarm is present, it must be manually reset. |
781 |
Overfeed Timer B |
782 |
Overfeed Timer C |
783 |
Overfeed Timer D |
784 |
On Delay A |
785 |
Sets a delay time for the relay to turn on. |
786 |
On Delay B |
787 |
On Delay C |
788 |
On Delay D |
789 |
Off Delay A |
790 |
Sets a delay time for the relay to turn off. |
791 |
Off Delay B |
792 |
Off Delay C |
793 |
Off Delay D |
794 |
Reset Overfeed Timer A |
795 |
Resets the Overfeed Timer for Relay A |
796 |
Undefined |
797 |
Non-Zero Resets |
798 |
Reset Overfeed Timer B |
799 |
Resets the Overfeed Timer for Relay B |
800 |
Reset Overfeed Timer C |
801 |
Resets the Overfeed Timer for Relay C |
802 |
Reset Overfeed Timer D |
803 |
Resets the Overfeed Timer for Relay D |
804 |
Max On Time A |
805 |
Sets the maximum time the relay can remain on. |
806 |
Max On Time B |
807 |
Max On Time C |
808 |
Max On Time D |
809 |
Min On Time A |
810 |
Sets the time the relay will remain on independent from the measured value. |
811 |
Min On Time B |
812 |
Min On Time C |
813 |
Min On Time D |
814 |
Max Off Time A |
815 |
Sets the maximum time the relay can remain off. |
816 |
Max Off Time B |
817 |
Max Off Time C |
818 |
Max Off Time D |
819 |
Min Off Time A |
820 |
Sets the time the relay will remain off independent from the measured value. |
821 |
Min Off Time B |
822 |
Min Off Time C |
823 |
Min Off Time D |
824 |
Warning Level A |
825 |
Sets the level for device warning activation. |
826 |
Warning Level B |
827 |
Warning Level C |
828 |
Warning Level D |
829 |
PWM Mode A |
830 |
Freq Mode A |
831 |
Auto, the relay output works as a PID controller. Manual, the relay output has an On/Off ratio as set in the PI Manual setting. |
832 |
PWM Mode B |
833 |
Freq Mode B |
834 |
PWM Mode C |
835 |
Freq Mode C |
836 |
PWM Mode D |
837 |
Freq Mode D |
838 |
PWM Integral A |
839 |
Freq Integral A |
840 |
Sets the integral part of the PID controller. |
841 |
PWM Integral B |
842 |
Freq Integral B |
843 |
PWM Integral C |
844 |
Freq Integral C |
845 |
PWM Integral D |
846 |
Freq Integral D |
847 |
Range A |
848 |
Range B |
849 |
Range C |
850 |
Range D |
851 |
Fail Safe Mode A |
852 |
Fail Safe Mode 'ON' implies the relay is energized in the non-alarm state. |
853 |
NO |
854 |
YES |
855 |
Fail Safe Mode B |
856 |
Fail Safe Mode C |
857 |
Fail Safe Mode D |
858 |
Net Port % Good Msg |
859 |
Percentage of good messages on the Network Card Port. |
860 |
Svc Port % Good Msg |
861 |
Percentage of good messages on the Service Port. |
862 |
Port 1 % Good Msg |
863 |
Percentage of good messages on Port 1. |
864 |
Port 2 % Good Msg |
865 |
Percentage of good messages on Port 2. |
866 |
Port 3 % Good Msg |
867 |
Percentage of good messages on Port 3. |
868 |
Port 4 % Good Msg |
869 |
Percentage of good messages on Port 4. |
870 |
Current Temperature |
871 |
Internal temperature of the sc200. |
872 |
12V Supply Voltage |
873 |
Measured value of the sc200 12 volt supply. |
874 |
3.3V Supply Current |
875 |
Total 3.3V Supply current (A). |
876 |
5.3f |
877 |
Amps |
878 |
12V Supply Current |
879 |
Total 12V Supply current (A). |
880 |
Port 3 Supply Current |
881 |
Smart Sensor 1 - 12V Supply current (A). |
882 |
Port 4 Supply Current |
883 |
Smart Sensor 2 - 12V Supply current (A). |
884 |
Port 1 Supply Current |
885 |
Analog Sensor 1 - 12V Supply current (A). |
886 |
Port 2 Supply Current |
887 |
Analog Sensor 2 - 12V Supply current (A). |
888 |
Max Temperature |
889 |
The daily maximum sc200 internal temperature in C. |
890 |
Min Temperature |
891 |
The daily minimum sc200 internal temperature in C. |
892 |
12V Ground Potential |
893 |
The measured voltage between A/D common and the 12 volt supply ground. |
894 |
Output 1 Value |
895 |
Display or Set level on Output 1. |
896 |
Output 2 Value |
897 |
Display or Set level on Output 2. |
898 |
Off |
899 |
Sensor 1 |
900 |
Sensor 2 |
901 |
Network Module |
902 |
sc200 |
903 |
Network Errors |
904 |
Sensor communication error flags. |
905 |
Sensor 1 Com Err |
906 |
Sensor 2 Com Err |
907 |
Network Status |
908 |
Sensor communication and relay status. |
909 |
Sensor 1 Connected |
910 |
Sensor 2 Connected |
911 |
Relay A Active |
912 |
Relay B Active |
913 |
Relay C Active |
914 |
Relay D Active |
915 |
Set Defaults |
916 |
Test Enable Out 1 |
917 |
Enables test of Output 1. |
918 |
Test Enable Out 2 |
919 |
Enables test of Output 2. |
920 |
Cal 4 mA Out 1 |
921 |
Calibration counts for 4 mA Output 1. |
922 |
Cal 4 mA Out 2 |
923 |
Calibration counts for 4 mA Output 2. |
924 |
Cal 20 mA Out 1 |
925 |
Calibration counts for 20 mA Output 1. |
926 |
Cal 20 mA Out 2 |
927 |
Calibration counts for 20 mA Output 2. |
928 |
Test Relay A Enable |
929 |
Enables test of Relay A. |
930 |
Test Relay B Enable |
931 |
Enables test of Relay B. |
932 |
Test Relay C Enable |
933 |
Enables test of Relay C. |
934 |
Test Relay D Enable |
935 |
Enables test of Relay D. |
936 |
Test Relay A Value |
937 |
Sets the test state for Relay A. |
938 |
Test Relay B Value |
939 |
Sets the test state for Relay B. |
940 |
Test Relay C Value |
941 |
Sets the test state for Relay C. |
942 |
Test Relay D Value |
943 |
Sets the test state for Relay D. |
944 |
Discrete Input 1 Mode |
945 |
The Modes ON/High and OFF/Low are set by jumpers per the Instruction Manual. |
946 |
Active ON/High |
947 |
Active OFF/Low |
948 |
Discrete Input 2 Mode |
949 |
Discrete Input 3 Mode |
950 |
DI 1 Warning Enable |
951 |
When Enabled, a sc200 Warning will be generated when the Input is Active. |
952 |
DI 2 Warning Enable |
953 |
DI 3 Warning Enable |
954 |
DI 1 Hold Mode |
955 |
Selection list for the Discrete Input's Output Hold Mode. |
956 |
Hold |
957 |
Transfer |
958 |
DI 2 Hold Mode |
959 |
DI 3 Hold Mode |
960 |
DI 1 On Delay (0-999) |
961 |
The delay in seconds before the Input goes Active. |
962 |
DI 2 On Delay (0-999) |
963 |
DI 3 On Delay (0-999) |
964 |
DI 1 Off Delay (0-999) |
965 |
The delay in seconds before the Input goes Inactive. |
966 |
DI 2 Off Delay (0-999) |
967 |
DI 3 Off Delay (0-999) |
968 |
Net Mod Good Msg |
969 |
Number of good messages on the network module. |
970 |
10d |
971 |
Net Mod Bad Msg |
972 |
Number of bad messages on the Network Card Port. |
973 |
Svc Port Good Msg |
974 |
Number of good messages on the Service Port. |
975 |
Svc Port Bad Msg |
976 |
Number of bad messages on the Service Port. |
977 |
Port 1 Good Msg |
978 |
Number of good messages on Port 1. |
979 |
Port 1 Bad Msg |
980 |
Number of bad messages on Port 1. |
981 |
Port 2 Good Msg |
982 |
Number of good messages on Port 2. |
983 |
Port 2 Bad Msg |
984 |
Number of bad messages on Port 2. |
985 |
Port 3 Good Msg |
986 |
Number of good messages on Port 3. |
987 |
Port 3 Bad Msg |
988 |
Number of bad messages on Port 3. |
989 |
Port 4 Good Msg |
990 |
Number of good messages on Port 4. |
991 |
Port 4 Bad Msg |
992 |
Number of bad messages on Port 4. |
993 |
Math Units |
994 |
Six character measurement unit for the sc200 Calculated Value. |
995 |
Math Parameters |
996 |
Six character measurement parameter for the sc200 Calculated Value. |
997 |
Reset Modbus Stats |
998 |
Clears all Modbus stats to zero. |
999 |
Test Relay A |
1000 |
Method to test Relay A. |
1001 |
Test Relay B |
1002 |
Method to test Relay B. |
1003 |
Test Relay C |
1004 |
Method to test Relay C. |
1005 |
Test Relay D |
1006 |
Method to test Relay D. |
1007 |
Clear Event Logs |
1008 |
This method is used to clear an event log for sensor1, sensor2, the network module or the sc200. |
1009 |
Clear Data Logs |
1010 |
This method is used to clear a data log for sensor1, sensor2, the network module or the sc200. |
1011 |
Resets controller configuration settings to default values. |
1012 |
This method will reset the over feed timer for relay A. |
1013 |
This method will reset the over feed timer for relay B |
1014 |
This method will reset the over feed timer for relay C. |
1015 |
This method will reset the over feed timer for relay D. |
1016 |
Output 1 Setup |
1017 |
This method is used to setup or configure Output 1. |
1018 |
Output 2 Setup |
1019 |
This method is used to setup or configure output 2. |
1020 |
Relay A Setup |
1021 |
This method is used to setup or configure relay A. |
1022 |
Relay B Setup |
1023 |
This method is used to setup or configure relay B. |
1024 |
Relay C Setup |
1025 |
This method is used to setup or configure relay C. |
1026 |
Relay D Setup |
1027 |
This method is used to setup or configure relay D. |
1028 |
Return Relay to Automatic Control |
1029 |
Sensor 1 Status |
1030 |
Sensor 1 Installation Status. |
1031 |
Installed |
1032 |
Sensor 2 Status |
1033 |
Sensor 2 Installation Status |
1034 |
Loop 1 Gauge |
1035 |
Loop 2 Gauge |
1036 |
sc200 Identification |
1037 |
Loop 1 |
1038 |
Loop 2 |
1039 |
Tests and Resets |
1040 |
sc200 Warnings / Errors |
1041 |
Internal Modbus Message Stats |
1042 |
Loop Output Maintenance |
1043 |
Output 1 Maintenance |
1044 |
Output 2 Maintenance |
1045 |
Overfeed Timer A Reset |
1046 |
Overfeed Timer B Reset |
1047 |
Overfeed Timer C Reset |
1048 |
Overfeed Timer D Reset |
1049 |
12V Supply |
1050 |
Supply Currents |
1051 |
sc200 Internal Temperatures |
1052 |
Relay Test |
1053 |
Log Clears |
1054 |
Service Port Status |
1055 |
Ports Status |
1056 |
Sensor Network Status |
1057 |
sc200 Relay Status |
1058 |
Discrete Inputs |
1059 |
User Setup Menu |
1060 |
Output Information |
1061 |
An Output must be mapped to the HARTFieldDevice (sensor)to be displayed. |
1062 |
See Output Info Help. |
1063 |
Output Setup |
1064 |
Relay Information |
1065 |
A Relay must be mapped to the HART Field Device (sensor)to be displayed. |
1066 |
See Relay Info Help. |
1067 |
Relays Setup |
1068 |
Discrete Input Setup |
1069 |
General Setup |
1070 |
sc200 Output 1 |
1071 |
sc200 Output 2 |
1072 |
Error Mode |
1073 |
Common Tables |
1074 |
Interoperable * |
1075 |
Universal |
1076 |
Common Practice |
1077 |
Pressure Universal |
1078 |
Pressure Common |
1079 |
Menu |
1080 |
Rosemount Magflow |
1081 |
Rosemount Pressure |
1082 |
Moore Products |
1083 |
Siemens Pressure |
1084 |
P2 Common |
1085 |
PV |
1086 |
Wireless |
1087 |
Munge |
1088 |
Analytical Oxygen/Combustion/Methane/CO |
1089 |
Flow Totalizer - Rate Indicator |
1090 |
Flow Variable Area |
1091 |
Flow Coriolis Mass |
1092 |
Flow Ultrasonic - Gas |
1093 |
Flow Ultrasonic - Liquid |
1094 |
Level Ultrasonic |
1095 |
Level Radar - Microwave |
1096 |
Pressure Absolute |
1097 |
Level Capacitance |
1098 |
Pressure Differential |
1099 |
Level Hydrostatic |
1100 |
Level |
1101 |
Pressure Absolute/Relative |
1102 |
Level Microwave |
1103 |
Level Radar |
1104 |
Level Density - Nuclear |
1105 |
Flow Electromagnetic |
1106 |
Flow Vortex |
1107 |
Flow Magmeter - Rack Version |
1108 |
Flow High Performance Magmeter |
1109 |
Flow Mass |
1110 |
Flow Ultrasonic |
1111 |
Flow Magnetic |
1112 |
Flow Magnetic - Custody Transfer |
1113 |
Flow 2-wire Magnetic |
1114 |
Flow Coriolis Mass - Custody Transfer |
1115 |
Flow 2-wire Vortex |
1116 |
Flow Thermal |
1117 |
Temperature Head Mounted |
1118 |
Temperature Compact |
1119 |
Temperature DIN Rail Mounted |
1120 |
Analytical pH |
1121 |
Analytical Conductivity - Conductive |
1122 |
Analytical Conductivity - Inductive |
1123 |
Temperature Sensor |
1124 |
Level Servo |
1125 |
Temperature DIN Rail |
1126 |
Analytical ORP |
1127 |
Analytical pION |
1128 |
Analytical IConc |
1129 |
Analytical Electrode Conductivity |
1130 |
Analytical Toroidal Conductivity |
1131 |
Valve Positioner |
1132 |
Valve Controller |
1133 |
Pressure Multivariable |
1134 |
Analytical pH/ORP |
1135 |
Flow Coriolis |
1136 |
Analytical Conductivity |
1137 |
Valve Positioner/Pneumatic |
1138 |
Valve Actuator - Electrical |
1139 |
Temperature Dual Input |
1140 |
Valve Actuator |
1141 |
Analytical Corrosion |
1142 |
Analytical Density - Nuclear |
1143 |
Level Continuous Gamma |
1144 |
Level Laser Ranging |
1145 |
Pressure Safety |
1146 |
Flow Vortex/Swirl |
1147 |
Flow |
1148 |
Analytical 4 Electrode Conductivity |
1149 |
Analytical 2 Electrode Conductivity |
1150 |
Analytical Troidal Conductivity |
1151 |
Analytical pH/ORP/pIon/Ion Conc |
1152 |
Analytical Oxygen |
1153 |
Pressure P, dP + Multivariable |
1154 |
Temperature 2-wire |
1155 |
Pressure w/ PID |
1156 |
Pressure Saturated Steam |
1157 |
Analytical Density - Gamma Ray |
1158 |
Level Gamma Ray |
1159 |
Analytical Weight - Gamma Ray |
1160 |
Level w/ Density - Gamma Ray |
1161 |
Analytical Motorized Density |
1162 |
Flow Multivariable Mass |
1163 |
Temperature Multivariable |
1164 |
Specialty HART-to-Analog Signal Conv |
1165 |
Flow dP |
1166 |
Pressure Low Power |
1167 |
Pressure Certified SIL2 |
1168 |
Interoperable |
1169 |
Analytical Density |
1170 |
Valve Positioner/Electropneumatic |
1171 |
Flow Magnetic Inductive |
1172 |
Analytical |
1173 |
Analytical Opacity Monitor |
1174 |
Analytical Oxygen/Combustion |
1175 |
Analytical 2-wire Conductivity |
1176 |
Analytical 2-wire Conductivity w/ pure H20 |
1177 |
Analytical 2-wire Conductivity w/ % conc |
1178 |
Analytical 2-wire Oxygen/Chlorine/Ozone |
1179 |
Analytical 2-wire pH/ORP |
1180 |
Analytical 2-wire Contacting/Toroidal |
1181 |
Analytical Oxygen - High Temperature |
1182 |
Analytical Conductivity w/ % conc |
1183 |
Analytical pH/ORP/ISE |
1184 |
Analytical Amperometric Analyzer |
1185 |
Analytical Dual Input Analyzer |
1186 |
Pressure Pressure/Level |
1187 |
Analytical Consistency |
1188 |
Analytical Microwave-Pulp Consistency |
1189 |
Analytical Microwave Consistency |
1190 |
Valve Digital Positioner |
1191 |
Valve Smart Valve Positioner |
1192 |
Temperature Differential |
1193 |
Analytical Dissolved Oxygen |
1194 |
Analytical pH Controller |
1195 |
Analytical Conductivity/Resistivity Controller |
1196 |
Analytical Conductivity/Resistivity |
1197 |
Level Displacement |
1198 |
Temperature w/ PID |
1199 |
Valve Universal Positioner |
1200 |
Level Radar - Reflex |
1201 |
Pressure Gauge |
1202 |
Pressure Sealed Gauge |
1203 |
Valve Digital Positioner w/ PID |
1204 |
Valve Intelligent Positioner |
1205 |
Analytical Gas Detector |
1206 |
Analytical Infrared Gas Detector |
1207 |
Analytical Electrochemical Gas Detector |
1208 |
Analytical Measuring Instruments |
1209 |
Level Magnostrictive |
1210 |
Valve On/Off Valve Controller |
1211 |
Level Radar - Pulse |
1212 |
Level Continuous Capactive |
1213 |
Valve Damper Actuator |
1214 |
Valve Electric Damper Actuator |
1215 |
Flow Electronic |
1216 |
Flow Vortex - Multiparameter Mass |
1217 |
Analytical UV Flame Detector |
1218 |
Analytical UVIR Flame Detector |
1219 |
Analytical IR Flame Detector |
1220 |
Analytical Fire Detector |
1221 |
Analytical Gas Detector Display Interface |
1222 |
Analytical Toxic Gas Detector |
1223 |
Analytical Flame Detector |
1224 |
Level Nuclear |
1225 |
Analytical Microwave |
1226 |
Analytical Density - Radiometric |
1227 |
Level Radiometric |
1228 |
Analytical Gas/Steam |
1229 |
Level Nucleonic |
1230 |
Analytical Blade Consistency |
1231 |
Analytical Rotary Consistency |
1232 |
Pressure Absolute/Differential |
1233 |
Positioner |
1234 |
Analytical Combustible Gas Detector |
1235 |
Analytical Point Infrared Gas Detector |
1236 |
Analytical Multi-Detector Monitor System |
1237 |
Valve Positioner/Electro-hydraulic |
1238 |
Valve Monitor |
1239 |
Enable on token-passing data link layer only |
1240 |
Enable on TDMA data link layer only |
1241 |
Enable on both TDMA and token-passing data link layer |
1242 |
WirelessHART Device |
1243 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
1244 |
Primary |
1245 |
Primary Analog Channel |
1246 |
Secondary |
1247 |
Secondary Analog Channel |
1248 |
Tertiary |
1249 |
Tertiary Analog Channel |
1250 |
Quaternary |
1251 |
Quaternary Analog Channel |
1252 |
Quinary |
1253 |
Quinary Analog Channel |
1254 |
Critical Power Failure |
1255 |
Lock all |
1256 |
Trim not supported |
1257 |
Configuration cannot be changed |
1258 |
Locked by gateway |
1259 |
Analog Input Channel |
1260 |
Analog Output Channel |
1261 |
Secondary Analog Channel Saturated |
1262 |
Tertiary Analog Channel Saturated |
1263 |
Quaternary Analog Channel Saturated |
1264 |
Quinary Analog Channel Saturated |
1265 |
Secondary Analog Channel Fixed |
1266 |
Tertiary Analog Channel Fixed |
1267 |
Quaternary Analog Channel Fixed |
1268 |
Quinary Analog Channel Fixed |
1269 |
Simulation active |
1270 |
The device is in simulation mode and the level output is not representative of the process. |
1271 |
Non-Volatile memory failure |
1272 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
1273 |
Volatile memory error |
1274 |
The RAM memory check is invalid or maybe corrupt |
1275 |
Watchdog reset executed |
1276 |
A watchdog reset has been performed |
1277 |
Voltage conditions out of range |
1278 |
A voltage condition is outside its allowable range |
1279 |
Environmental conditions out of range |
1280 |
An internal or environmental condition is beyond acceptable limits. |
1281 |
Electronic failure |
1282 |
A hardware problem not related to the sensor has been detected. |
1283 |
Subdevice list changed |
1284 |
Duplicate master detected |
1285 |
Capacity Denied |
1286 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
1287 |
Bandwidth allocation pending |
1288 |
Block transfer pending |
1289 |
Continuous |
1290 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
1291 |
Window |
1292 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
1293 |
Rising |
1294 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
1295 |
Falling |
1296 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
1297 |
On-change |
1298 |
The Burst Message is triggered when any value in the message changes. |
1299 |
Battery life |
1300 |
Percent range |
1301 |
Loop current |
1302 |
Primary variable |
1303 |
Secondary variable |
1304 |
Tertiary variable |
1305 |
Quaternary variable |
1306 |
Enable event notification on token-passing data link layer |
1307 |
Enable event notification on TDMA data link layer |
1308 |
Enable event notification on TDMA and token-passing data link layer |
1309 |
Configuration changed event pending |
1310 |
Device status event pending |
1311 |
More status available event pending |
1312 |
Disable |
1313 |
Enable single data point trending |
1314 |
Enable filtered trending |
1315 |
The sample is filtered using a time constant equal to one-third the trend sample period |
1316 |
Enable average trending |
1317 |
The value obtained by dividing the sum of all samples in the period by the total number of samples |
1318 |
Read receive time |
1319 |
Write date and time |
1320 |
Non-Volatile Clock |
1321 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
1322 |
Clock Uninitialized |
1323 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
1324 |
Line Power |
1325 |
Device is powered via a hardwired connection to a plant power source. |
1326 |
Battery Power |
1327 |
Device operates solely of its (internal) battery. |
1328 |
Recharg.B./Power Sc. |
1329 |
Rechargeable battery power or power scavenging. The device has a short-term supply of energy (must be at least one hour) that is continuously being replenished. The device's power is being replenished by harvesting and converting energy from the environment surrounding the device (e.g. solar, vibration, heat). |
1330 |
Idle |
1331 |
Active search |
1332 |
Looking for neighbors. |
1333 |
Negotiating |
1334 |
Quarantined |
1335 |
Operational |
1336 |
Suspended |
1337 |
Deep sleep |
1338 |
Ultra-low power |
1339 |
Wireless Signal Found |
1340 |
Network packets have been recognized |
1341 |
Wireless Signal Identified |
1342 |
Wireless Time Synchronized |
1343 |
WirelessHart Signal Identified |
1344 |
Network Admission Requested |
1345 |
Set on transmission of first join request |
1346 |
Join Retrying |
1347 |
Set after the first join request retry (i.e. when number of join requests > 2). Cleared when device is authenticated or when Active Search started. |
1348 |
Join Failed |
1349 |
Set on transition from Active Search mode to Deep Sleep mode. Cleared on transition from Deep Sleep to Active Search. |
1350 |
Network Security Clearance Granted |
1351 |
Network Manager Session Established |
1352 |
Network Joined |
1353 |
Network Joined. Normal superframe and links obtained. |
1354 |
Network Bandwidth Requested |
1355 |
Negotiating Network Properties. Gateway session obtained. |
1356 |
Join Complete |
1357 |
Normal operating mode. |
1358 |
No restrictions |
1359 |
SI Only |
1360 |
Restricted to SI Unit codes only |
1361 |
Open - accept all devices |
1362 |
Use Whitelist for access |
1363 |
Use Blacklist |
1364 |
Use Whitelist and Blacklist |
1365 |
Lockdown - disable join links |
1366 |
Process automation device |
1367 |
Discrete device |
1368 |
Hybrid:Process automation+discrete |
1369 |
I/O System |
1370 |
WirelessHART Process automation device |
1371 |
WirelessHART Discrete device |
1372 |
WirelessHART Process automation+discrete |
1373 |
WirelessHART Gateway |
1374 |
WirelessHART Access point |
1375 |
WirelessHART Process adapter |
1376 |
WirelessHART Discrete adapter |
1377 |
WirelessHART enabled handheld/portable maintenance tool |
1378 |
Don't attempt to join |
1379 |
Join now |
1380 |
Attempt to join immediately on powerup or reset |
1381 |
US |
1382 |
United States of America |
1383 |
JP |
1384 |
Japan |
1385 |
DE |
1386 |
Germany |
1387 |
FR |
1388 |
France |
1389 |
ES |
1390 |
Spain |
1391 |
RU |
1392 |
Russian Federation |
1393 |
CN |
1394 |
People's Republic of China |
1395 |
Nominal |
1396 |
Device Power Status is nominal |
1397 |
Device Power Status is low |
1398 |
Critically Low |
1399 |
Device Power Status is critically low |
1400 |
Recharging - Low |
1401 |
Recharging - High |
1402 |
Secondary Master |
1403 |
Primary Master |
1404 |
PV Analog Channel Saturated |
1405 |
PV Analog Channel Fixed |
1406 |
Device Profile |
1407 |
|
1408 |
Device Diagnostic Status 0 |
1409 |
Device Diagnostic Status 1 |
1410 |
I/O and Subdevice Status |
1411 |
WirelessHART Status |
1412 |
Time Set Option |
1413 |
Real Time Clock Flags |
1414 |
Country |
1415 |
Country of intended device installation |
1416 |
SI Unit Control |
1417 |
Indicates if device should only display SI unit codes |
1418 |
Burst Trigger Units |
1419 |
Burst Trigger Class |
1420 |
Wired HART Enabled |
1421 |
Enabled on Wired HART device connection only |
1422 |
Wireless HART Enabled |
1423 |
Enabled on WirelessHART device connection only |
1424 |
Both Wired and WirelessHart Enabled |
1425 |
Enabled on both Wired and WirelessHART connections |
1426 |
Burst Msg Trigger Mode |
1427 |
Event Status |
1428 |
Event Notification Control |
1429 |
Operate mode Mask |
1430 |
Not Defined |
1431 |
Power Source |
1432 |
Power source of the device |
1433 |
Wireless Mode |
1434 |
Join Status |
1435 |
The current location in the joining sequence of operations. |
1436 |
Network Access Mode |
1437 |
Join Mode |
1438 |
Power Status |
1439 |
Power status of the device |
1440 |
Master Mode |
1441 |
This sets the I/O System's Data Link Layer operation on all channels as either primary or secondary master mode. |
1442 |
Trend Units |
1443 |
Trend Device Variable Unit Code |
1444 |
Trend Classification |
1445 |
Trend Device Variable Classification |
1446 |
Trend Control |
1447 |
Device Status Mask |
1448 |
Ext dev status Mask |
1449 |
Device Diagnostic Status 0 Mask |
1450 |
Device Diagnostic Status 1 Mask |
1451 |
Device Diagnostic Status 2 Mask |
1452 |
Device Diagnostic Status 3 Mask |
1453 |
AO saturated Mask |
1454 |
AO fixed Mask |
1455 |
Latched Device Status |
1456 |
Latched Ext dev status |
1457 |
Latched Operate Mode |
1458 |
Latched Device Diagnostic Status 0 |
1459 |
Latched Device Diagnostic Status 1 |
1460 |
Latched Device Diagnostic Status 2 |
1461 |
Latched Device Diagnostic Status 3 |
1462 |
Latched AO saturated |
1463 |
Latched AO fixed |
1464 |
Wireless Module Device Type |
1465 |
Device type of the Wireless Radio |
1466 |
Wireless Module Manufacturer ID |
1467 |
Manufacturer of the Wireless Radio contained in this device |
1468 |
Subdevice Device Type |
1469 |
Device type of the subdevice attached to this system. |
1470 |
Subdevice Manufacturer Id |
1471 |
Manufacturer of the subdevice attached to this system. |
1472 |
Time Stamp |
1473 |
Relative time value on network |
1474 |
8u |
1475 |
Dynamic Variables Returned for Device Variables |
1476 |
Command Response Truncated |
1477 |
Configuration Change Counter Mismatch |
1478 |
Status bytes mismatch |
1479 |
8.3f |
1480 |
Change units from %{eHACHMeasurementUnits}? |
1481 |
No;Yes |
1482 |
Select units |
1483 |
Please close SC LDO2 Setup Window and re-open to view new DO Values. |
1484 |
Change units from %{eHACHPressureAndAltitudeUnits}? |
1485 |
Input Altitude in meters between -1524.0 and 4572.0 |
1486 |
Input Altitude in feet between -5000.0 and 15000.0 |
1487 |
Input Pressure in mmHg between 0.0 and 2000.0 |
1488 |
Input Pressure in Torr between 0.0 and 2000.0 |
1489 |
Please close SC LDO2 Setup Window and re-open to view new Pressure/Altitude value. |
1490 |
Please wait. |
1491 |
Change Salinity value from %{rSalinity_ppt} 0/00? |
1492 |
Select the measured value of Conductivity as an integer from 5 to 54 mS/cm. 0 mS/cm = 0 Salinity |
1493 |
0;5;6;7;8;9;10; |
1494 |
11;12;13;14;15;16;17;18;19;20;21;22;23;24;25;26;27;28;29;30;31;32;33;34;35;36;37;38;39;40;42;44;46;48;50;52;54 |
1495 |
Please enter a measured value of Conductivity that is an integer from 5 to 54 mS/cm or a value of 0. |
1496 |
The Salinity value is now equal to %{rSalinity_ppt} 0/00. |
1497 |
Please close SC LDO2 Setup Window and re-open to view the new Salinity value. |
1498 |
Input Salinity value in ppt between 0.0 and 250.0 |
1499 |
Please close LDO2 Setup Window and re-open to view new Salinity value. |
1500 |
Output A is fixed at %{0}. |
1501 |
Output B is fixed at %{0}. |
1502 |
Output C is fixed at %{0}. |
1503 |
Output 1 is fixed at %{0}. |
1504 |
Output 2 is fixed at %{0}. |
1505 |
Output A output equal to reference meter? |
1506 |
Output B output equal to reference meter? |
1507 |
Output C output equal to reference meter? |
1508 |
Output 1 output equal to reference meter? |
1509 |
Output 2 output equal to reference meter? |
1510 |
Loops and Relays can be returned to automatic control. |
1511 |
Do you want to reset Modbus statistics? |
1512 |
NO;YES |
1513 |
Modbus statistics not reset, aborting. |
1514 |
This will reset Modbus Statistics to zero. Proceed? |
1515 |
Unexpected error, Modbus statistics not reset. Aborting |
1516 |
Modbus statistics have been reset. |
1517 |
Relay A should be removed from automatic control. |
1518 |
Do you want to test Relay A? |
1519 |
Aborting test. |
1520 |
Setting Relay A active. |
1521 |
Is Relay A active? |
1522 |
Relay A Error, Fix and retest. Aborting test. |
1523 |
Setting Relay A inactive. |
1524 |
Is Relay A inactive? |
1525 |
Relay A test successful. |
1526 |
Relay B should be removed from automatic control. |
1527 |
Do you want to test Relay B? |
1528 |
Setting Relay B active. |
1529 |
Is Relay B active? |
1530 |
Setting Relay B inactive. |
1531 |
Is Relay B inactive? |
1532 |
Relay B Error, Fix and retest. Aborting test. |
1533 |
Relay B test successful. |
1534 |
Relay C should be removed from automatic control. |
1535 |
Do you want to test Relay C? |
1536 |
Setting Relay C active. |
1537 |
Is Relay C active? |
1538 |
Relay C Error, Fix and retest. Aborting test. |
1539 |
Setting Relay C inactive. |
1540 |
Is Relay C inactive? |
1541 |
Relay C test successful. |
1542 |
Relay D should be removed from automatic control. |
1543 |
Do you want to test Relay D? |
1544 |
Setting Relay D active. |
1545 |
Is Relay D active? |
1546 |
Relay D Error, Fix and retest. Aborting test. |
1547 |
Setting Relay D inactive. |
1548 |
Is Relay D inactive? |
1549 |
Relay D test successful. |
1550 |
Which event log do you want to clear? |
1551 |
None;Sensor1;Sensor2;Network Module;sc200 |
1552 |
No event logs cleared, aborting. |
1553 |
Unexpected error, event log not cleared. Aborting. |
1554 |
Sensor 1 event log cleared. |
1555 |
Sensor 2 event log cleared. |
1556 |
Network card event log cleared. |
1557 |
sc200 event log cleared. |
1558 |
Which data log do you want to clear? |
1559 |
No data logs cleared, aborting. |
1560 |
Unexpected error, data log not cleared. Aborting. |
1561 |
Sensor 1 data log cleared. |
1562 |
Sensor 2 data log cleared. |
1563 |
Network data event log cleared. |
1564 |
sc200 data log cleared. |
1565 |
WARN-Loops and Relays should be removed from automatic control. |
1566 |
Do you want to set default values? |
1567 |
Default values not set, aborting. |
1568 |
This will set default values to system. Proceed? |
1569 |
Unexpected error, default values not set. Aborting |
1570 |
Default values have been set. |
1571 |
Do you want to reset Overfeed Timer Relay A? |
1572 |
Overfeed Timer Relay A will not be reset, aborting. |
1573 |
This will reset Overfeed Timer A. Proceed? |
1574 |
Overfeed Timer Relay A not reset, aborting. |
1575 |
Unexpected error, Overfeed Timer Relay A not reset. Aborting |
1576 |
Overfeed Timer Relay A reset. |
1577 |
Do you want to reset Overfeed Timer Relay B? |
1578 |
Overfeed Timer Relay B will not be reset, aborting. |
1579 |
This will reset Overfeed Timer B. Proceed? |
1580 |
Overfeed Timer Relay B not reset, aborting. |
1581 |
Unexpected error, Overfeed Timer Relay B not reset. Aborting |
1582 |
Overfeed Timer Relay B reset. |
1583 |
Do you want to reset Overfeed Timer Relay C? |
1584 |
Overfeed Timer Relay C will not be reset, aborting. |
1585 |
This will reset Overfeed Timer C. Proceed? |
1586 |
Overfeed Timer Relay C not reset, aborting. |
1587 |
Unexpected error, Overfeed Timer Relay C not reset. Aborting |
1588 |
Overfeed Timer Relay C reset. |
1589 |
Do you want to reset Overfeed Timer Relay D? |
1590 |
Overfeed Timer Relay D will not be reset, aborting. |
1591 |
This will reset Overfeed Timer D. Proceed? |
1592 |
Overfeed Timer Relay D not reset, aborting. |
1593 |
Unexpected error, Overfeed Timer Relay D not reset. Aborting |
1594 |
Overfeed Timer Relay D reset. |
1595 |
Select output function |
1596 |
Linear;PID;Logarithmic;Bilinear |
1597 |
Input lower range value. |
1598 |
Please wait |
1599 |
Input upper range value. |
1600 |
Select Auto or Manual Mode. |
1601 |
Select Direct or Reverse Action. |
1602 |
Input Setpoint value. |
1603 |
Input Manual Value in percent (0-100). |
1604 |
Input Proportional Band. |
1605 |
Input Integral Time in seconds (0-9999). |
1606 |
Input Derivative Time in seconds (0-9999). |
1607 |
Input Transit Time in seconds (0-9999). |
1608 |
Input 50% value. |
1609 |
Input Knee point value. |
1610 |
Input Knee point current value in percent (0-100). |
1611 |
Please close Application Window and re-open to view new Values. |
1612 |
Select output function. |
1613 |
Input Knee point current value. |
1614 |
Input transfer setting. |
1615 |
Input Fail Safe Mode. |
1616 |
Select relay function. |
1617 |
Alarm;Feeder Control;Warning;Event Control;PWM Control;Freq Control |
1618 |
Input Low Alarm Value. |
1619 |
Input High Alarm Value. |
1620 |
Input Low Deadband Value. |
1621 |
Input High Deadband Value. |
1622 |
Input Alarm Off Delay Time in seconds (0-999). |
1623 |
Input Alarm On Delay Time in seconds (0-999). |
1624 |
Input Feeder Control Phase. |
1625 |
Input Setpoint Value. |
1626 |
Input Deadband Value. |
1627 |
Input Overfeed Timer Value in minutes (0-999). |
1628 |
Input Off Delay Time in seconds (0-999). |
1629 |
Input On Delay Time in seconds (0-999). |
1630 |
Input Warning Level. |
1631 |
Input On Maximum Time in minutes (0-9999). |
1632 |
Input Off Maximum Time in minutes (0-9999). |
1633 |
Input On Minimum Time in minutes (0-9999). |
1634 |
Input Off Minimum Time in minutes (0-9999). |
1635 |
Input Pulse Width Mode. |
1636 |
Input Control Phase. |
1637 |
Input Dead Zone Value. |
1638 |
Input Period Value in seconds (4-60). |
1639 |
Input Minimum Pulse Width in seconds (0.1-Maximum Pulse Width). |
1640 |
Input Maximum Pulse Width in seconds (Minimum Pulse Width - Period). |
1641 |
Input Integral Time in minutes (0-999). |
1642 |
Input Pulse Width (0.1-0.7). |
1643 |
Input Minimum Pulses (0.001-Maximum Pulses). |
1644 |
Input Maximum Pulses (Minimum Pulses-60). |
1645 |
Relay may be returned to automatic control. |