0 |
SVi1000 H/7 |
1 |
Fault Code 0 |
2 |
Reports Device Errors |
3 |
Reset |
4 |
Device reset indicator |
5 |
Low Power |
6 |
Input current less than 3.2 mA |
7 |
Actuator Error |
8 |
Unable to position the valve normally. |
9 |
Low Air Supply Warning |
10 |
Supply Pressure below low limit |
11 |
Position Error |
12 |
If T1 error is configured AND the position error outside the error band for time T1 |
13 |
Internal Error |
14 |
A critical internal fault |
15 |
Keypad Fault |
16 |
Possible moisture inside or connector or mechanical failure |
17 |
Marginal Power |
18 |
|
19 |
Fault Code 1 |
20 |
Calibration Failed |
21 |
Calibration of AO or pressure failed |
22 |
Find Stops Failed |
23 |
Find Stops process failed |
24 |
Autotune Failed |
25 |
Self-tuning failed to converge on acceptable parameters |
26 |
Operating System Fault |
27 |
An internal error from which the device recovered automatically |
28 |
Supply Pressure above upper limit |
29 |
Fault Code 2 |
30 |
Bias Out Of Range |
31 |
An important internal parameter is out of range |
32 |
I/P Out Of Range |
33 |
I/P current feedback is out of range |
34 |
Tempertaure outside operating range |
35 |
Low supply pressure alarm limit is configured too low |
36 |
Auxiliary memory fault |
37 |
Fault Code 3 |
38 |
NVM Checksum Error |
39 |
The device failed to read critical data from non-volatile memory |
40 |
RAM Checksum Error |
41 |
Flash Checksum Error |
42 |
Stack Error |
43 |
Factory Write Indicator |
44 |
Indicator of writes to non-volatile memory allowed only for firmware upgrade |
45 |
NVM Test Error |
46 |
An error in non-volatile memory could not be repaired |
47 |
ESD Trip occurred and requires user intervention to be released |
48 |
Factory Mode Indicator |
49 |
Device performs factory-authorized operations |
50 |
Fault Code 4 |
51 |
Ref Voltage Fault |
52 |
May indicate faulty A/D subsystem |
53 |
Position Sensor Fault |
54 |
Current Sensor Fault |
55 |
Temperature Sensor Fault |
56 |
Output Pressure Sensor Fault (Single acting) or Output 2 Pressure Sensor Fault (double acting) |
57 |
Supply Pressure Sensor Fault (Single acting) or Output 1 Pressure Sensor Fault (double acting) |
58 |
Supply Pressure Sensor Fault (double acting only) |
59 |
Fault Code 5 |
60 |
I/P Pressure Sensor Fault |
61 |
Atmospheric Pressure Sensor Fault |
62 |
NVM Write Fault |
63 |
A write to non-volatile memory failed; data may be lost on reset |
64 |
IRQ Fault |
65 |
The test didn't complete in time |
66 |
MCU Internal Malfunction |
67 |
Software Error |
68 |
special |
69 |
smart filter |
70 |
Position Quality |
71 |
Setpoint Quality |
72 |
Temperature Quality |
73 |
Pilot Pressure Quality |
74 |
Signal |
75 |
Signal Unit |
76 |
Unit_ValveStrokes |
77 |
Unit ValveStrokes |
78 |
Unit_RawCounts |
79 |
Unit RawCounts |
80 |
Unit_DirectionChanges |
81 |
Unit DirectionChanges |
82 |
Unit_SignalCharacter |
83 |
Unit SignalCharacter |
84 |
Signal Quality |
85 |
DO1 |
86 |
DO1 Unit |
87 |
DO1 Quality |
88 |
DO2 |
89 |
DO2 Unit |
90 |
DO2 Quality |
91 |
VoltsInput |
92 |
VoltsInput Unit |
93 |
VoltsInput Quality |
94 |
RawPosition |
95 |
RawPosition Unit |
96 |
RawPosition Quality |
97 |
NumStrokes |
98 |
NumStrokes Unit |
99 |
NumStrokes Quality |
100 |
NumCycles |
101 |
NumCycles Unit |
102 |
NumCycles Quality |
103 |
PosRetransmit |
104 |
PosRetransmit Unit |
105 |
PosRetransmit Quality |
106 |
IPCurrent |
107 |
IPCurrent Unit |
108 |
IPCurrent Quality |
109 |
Friction |
110 |
Friction Unit |
111 |
Friction Quality |
112 |
PosErrBand |
113 |
PosErrBand Unit |
114 |
PosErrBand Quality |
115 |
PosLimitHigh |
116 |
PosLimitHigh Unit |
117 |
PosLimitHigh Quality |
118 |
PosLimitLow |
119 |
PosLimitLow Unit |
120 |
PosLimitLow Quality |
121 |
Characterization |
122 |
Characterization Unit |
123 |
Characterization Quality |
124 |
OpenStopAdjust |
125 |
OpenStopAdjust Unit |
126 |
OpenStopAdjust Quality |
127 |
PerctRange |
128 |
PerctRange Unit |
129 |
PerctRange Quality |
130 |
Position |
131 |
Setpoint |
132 |
Temperature |
133 |
Pilot Pressure |
134 |
Volts Input |
135 |
Raw Position |
136 |
Number Strokes |
137 |
Number Cycles |
138 |
Position Retransmit |
139 |
IP Current |
140 |
Valve Position |
141 |
Valve Position- Indicates the actual valve position in % with closed = 0 % and opened = 100 % |
142 |
% |
143 |
.1f |
144 |
Temperature- Indicates the actual valve Temperature in its engineering unit. |
145 |
.2f |
146 |
Input Signal |
147 |
Input Percent Range- Input signal (current loop) in percent |
148 |
7d |
149 |
Unused |
150 |
1 |
151 |
2 |
152 |
3 |
153 |
Maintenance |
154 |
Offline |
155 |
Online (SVi1000) |
156 |
Device Setup |
157 |
Process Variables |
158 |
Status/Diagnostics |
159 |
Calibration |
160 |
Commissioning Services |
161 |
Basic Info. |
162 |
Valve Travel |
163 |
Manual Position Setpoint |
164 |
Manual Setup |
165 |
Sensors |
166 |
Status/Faults |
167 |
Fault List |
168 |
display current and historic faults |
169 |
Sensor Measurements |
170 |
Continuous Diagnostics Data |
171 |
I/P |
172 |
Raw Sensor Data |
173 |
Select the slot |
174 |
To load the tuning parameters stored in the specific slot |
175 |
Display PID in Slot |
176 |
Display the PID's parameters in the other slot. |
177 |
Valve Tuning Data |
178 |
Alarms Setup |
179 |
Switches |
180 |
General |
181 |
Device Info |
182 |
Hart Info |
183 |
Information for HART |
184 |
Configuration |
185 |
Positioning |
186 |
Limits/Shutoff |
187 |
HART |
188 |
fill |
189 |
Firmware Version |
190 |
Display the firmware version. |
191 |
Change Mode |
192 |
Change mode: Changes the mode to Normal Mode, Manual Mode, or Setup Mode. Normal Mode: The position setpoint follows the input signal. Always return to Normal Mode to return to Control by the Control System. Manual Mode: The position setpoint remains at the last position prior to entering Manual Mode. Position setpoint may adjusted with the local pushbuttons or by HART commands. Setup Mode: The position setpoint remains at the last position prior to entering Setup Mode. Use this mode to change configuration and calibration information. |
193 |
Find Stops |
194 |
Run automatic find stops will establish the calibration position of the valve at actuator pressure vented and at full supply pressure. |
195 |
Autotune |
196 |
Run PID Autotune will automatically find best dynamic parameters for responsive valve positioning. Always isolate the valve from the process. The valve will move full open and full closed. The valve will continuously change position for about 5 minutes. To SVi1000 device, this operation need the device should be set PID selector to slot 0/ATO or slot 8/ATC |
197 |
Setup Wizard |
198 |
The setup wizard will guide you through basic steps to setup the positioner on an actuator. You may skip this wizard if the valve and positioner have been factory configured and calibrated. Always isolate the valve from the process. The wizard allows setting Air Action. The valve move full open and full closed. You will be prompted if mode must be changed. The steps are in the following order:
1. Run Find Stops, to set 0% to the valve closed position and to set 100% to full opened position.
2. Run Auto Tune to determine best dynamic parameters for position control.
3. Reset Basic Configuration and Calibration to Factory Defaults. |
199 |
Reset Basic Config and Cal To Defaults |
200 |
The Reset to Factory Default will set the following list of configuration and calibration parameters to defaults:
The PID tunning parameters in slot 0 are set to defaults;
Input current range is set to 4 to 20 mA;
Open stop adjustment is set to 100%;
Position error parameters are reset and disabled;
Bumpless transfer is disabled;
Position limits are disabled;
Position rate limits are disabled;
Tight shutoff is set to 0 and enabled;
Characterization is set to linear;
|
201 |
Current Faults |
202 |
Read the device's current faults. |
203 |
Read/Clear Faults |
204 |
Read the device's current faults. Optionally, the fault status may also be cleared. |
205 |
Clear Current Faults |
206 |
Clear the current device status/faults |
207 |
Historical Faults |
208 |
Read the device's historical faults. |
209 |
Clear All Faults |
210 |
Clear the historic and current device faults/status. |
211 |
Device Variables |
212 |
Read the current device variables. |
213 |
Set Position |
214 |
Set manual valve setpoint in % or set the position by equivalent input current to simulate non-linear characterization and limits. |
215 |
Full Open |
216 |
Fully Open valve |
217 |
Full Closed |
218 |
Fully Close Valve |
219 |
Configure Wizard |
220 |
Change configuration of Tight shutoff. |
221 |
Custom / Standard Characterization |
222 |
Control valves are often characterized to give a specific relationship between flow capacity (Cv) and percent opening of the valve. The valve can be characterized with special purpose trim or with the SVi1000 positioner. Selecting Custom Configuration displays an additional screen where the user can enter a custom characterization curve to fit his/her needs. The curve can have up to 9 points and points in between are linearly interpolated. |
223 |
Manual Stops |
224 |
This method is used to set manual stops. |
225 |
Open Stop Adjustment |
226 |
Open stop adjustment; Allows to adjust travel to less than mechanical limits. Must be between 60% and 100% of mechanical travel set with Find Stops. |
227 |
Manual Tuning |
228 |
The user can set tuning PID parameters for slot 0, even though the current selector was selected for the other slot. To SVi1000 device, The only slot 0 PID parameters can be updated. |
229 |
abort restore |
230 |
Set I/P Output |
231 |
Set IP Value. A value of 0 resets the device to allow the control task to set the I/P (changing modes or a reset also switches back to normal control modes). |
232 |
Input crnt snsr. |
233 |
To calibrate or to check the 4-20mA current sensor. |
234 |
Reset Cal Factory Defaults |
235 |
Reset Calibration Data: This will restore all sensor calibration data to factory defaults. If calibrations have been performed they will be lost! |
236 |
Reset Historian |
237 |
Reset valve historian data (stroke count, cycle count, time open, time closed and time near closed ) to 0. The previous data is lost and must be recorded to be used later. |
238 |
Reset SVi1000 |
239 |
This function performs a master reset to the device, causing it to go through its startup routine and re-initializing all of its operating parameters from non volatile memory. |
240 |
Range error warning |
241 |
Stop Procedure |
242 |
Set Analog Output |
243 |
User can set a fixed analog output of the position retransmitter.This is used only for checking loop wiring. Select Finished to return to position measurement function. |
244 |
Re-transmitter Range |
245 |
User can enter the upper and lower range values of valve position corresponding to output current 4 and 20 mA. The values typically are 0 and 100 %. |
246 |
Read Device S/N |
247 |
Read device and terminal board serial number. |
248 |
Position Unit |
249 |
Position Unit- Unit for valve position |
250 |
Setpoint Unit |
251 |
Setpoint Unit- Unit for valve Setpoint |
252 |
Temperature Unit |
253 |
Temperature Unit- Unit for valve Temperature |
254 |
Pressure Units |
255 |
Pressure Unit- The actual pressure reported by the SVI II ESD can be in psi, bar or kPa. |
256 |
DO Switch 1 Action |
257 |
Define Switch 1 action. |
258 |
Always Normal Pos |
259 |
Failsafe |
260 |
Tight Shutoff Active |
261 |
Alarm Low |
262 |
Alarm High |
263 |
Manual Mode |
264 |
DO Switch 1 Normally |
265 |
Define Switch 1 is normally closed or open. |
266 |
Closed |
267 |
Open |
268 |
DO Switch 1 Value |
269 |
The value is set for alarm low/high.( Switch 1 Value must be between -99.99 and 199.99. )
Attention: This is valid under the condition of switch 1 type is alarm low/high), but to some host, like SDC-625, to activate this field maybe need to close / open this page again. |
270 |
DO Switch 2 Action |
271 |
Define Switch 2 action. |
272 |
DO Switch 2 Normally |
273 |
Define Switch 2 is normally closed or open. |
274 |
DO Switch 2 Value |
275 |
The value is set for alarm low/high.( Switch 2 Value must be between -99.99 and 199.99. )
Attention: This is valid under the condition of switch 2 type is alarm low/high, but to some host, like SDC-625, to activate this field maybe need to close / open this page again. |
276 |
DI Switch Status |
277 |
Digital Input Switch is enabled or disabled. |
278 |
Disabled |
279 |
Enabled |
280 |
device busy |
281 |
command not implemented |
282 |
Mode(local) |
283 |
Setup |
284 |
Normal |
285 |
Manual |
286 |
cannot change mode |
287 |
device malfunction |
288 |
Fault Code 6 |
289 |
Reports Device |
290 |
TM ID error |
291 |
Terminal Module identification error |
292 |
TM Data Integrity error |
293 |
Retransmit Module Data Integrity Error |
294 |
Fault Code 7 |
295 |
Fault Code 0 History |
296 |
Reports Device History Errors |
297 |
Fault Code 1 History |
298 |
Fault Code 2 History |
299 |
Fault Code 3 History |
300 |
Fault Code 4 History |
301 |
Fault Code 5 History |
302 |
Fault Code 6 History |
303 |
Fault Code 7 History |
304 |
Clear Faults Type |
305 |
Historic Faults |
306 |
Total Travel |
307 |
Total number of valve strokes. |
308 |
Strokes |
309 |
10d |
310 |
Time Closed |
311 |
Time in hours for valve closed. |
312 |
h |
313 |
Time Open |
314 |
Time in hours for valve open. |
315 |
Time Near Closed |
316 |
Time in hours for valve near closed. |
317 |
Number of Cycles |
318 |
Number of cycles (direction changes). |
319 |
cycles |
320 |
Raw Signal |
321 |
A/D value of the primary signal. |
322 |
cts |
323 |
5d |
324 |
A/D value of the position sensor. |
325 |
Raw Pressure |
326 |
A/D value of the pressure sensor. |
327 |
Lower Stop |
328 |
Mechanical Stop without pressure in the actuator. |
329 |
Upper Stop |
330 |
Mechanical Stop with full pressure in the actuator. |
331 |
Board Temperature |
332 |
Internal SVi1000 temperature in °C. |
333 |
Min Temperature |
334 |
The lowest internal temperature recorded in °C. |
335 |
Max Temperature |
336 |
The upper internal temperature recorded in °C. |
337 |
I/P Current |
338 |
The actual I/P current, corrected for temperature |
339 |
4.2f |
340 |
3.1f |
341 |
PWM output |
342 |
PWM output. |
343 |
DO Switch 1 State |
344 |
DO Switch physical state (0 - closed; 1 - open) |
345 |
DO Switch 2 State |
346 |
Manual Setpoint |
347 |
Manual Setpoint- Manual valve setpoint in %. Value must be between -5.0% and +160.0%. |
348 |
6.2f |
349 |
Set Valve Pos In mA |
350 |
Manual Signal- Manual valve setpoint in mA. |
351 |
5.2f |
352 |
Setpoint Type |
353 |
Setpoint Type- Selects how setpoint or signal is set. |
354 |
Manual Signal (mA) value |
355 |
Manual Position (%) value |
356 |
Set Fully Closed |
357 |
Set Fully Opened |
358 |
Set To FailSafe |
359 |
I/P Value |
360 |
IP Value. A value of 0 resets the device to allow the control task to set the I/P (changing modes ora reset also switches back to normal control modes). |
361 |
d |
362 |
Local UI |
363 |
LocalUI locking status. |
364 |
Unlocked |
365 |
Locked |
366 |
Allow Override Limit |
367 |
Allow diag. and self-tune to override limits |
368 |
Tight Shutoff |
369 |
Tight Shutoff Limit- Threshold for tight shutoff: when selected, if the input signal would position the valve below the tight shutoff value, full supply or atmosphere pressure is supplied to the actuator to close the valve with full seating force. Set the TS value between -1% and 20%, then enable the TS function. |
370 |
Tight Shutoff Value |
371 |
Tight Shutoff Limit- Threshold for tight shutoff: when selected, if the input signal would position the valve below the tight shutoff value, full supply or atmosphere pressure is supplied to the actuator to close the valve with full seating force. Value between -1% and 20% and not to include. |
372 |
8.2f |
373 |
Pos Limit Low |
374 |
To enable or disable position lower limit |
375 |
Pos Limit Upper |
376 |
To enable or disable position upper limit |
377 |
Limit Low |
378 |
Lower Position Limit- The SVi1000 allows the user to establish software limit stops. The control functions will NOT allow the valve position to be lower than the lower position limit. However this option DOES NOT PROVIDE MECHANICAL STOPS for the valve. During an electrical or an air failure the valve will go to the failsafe position without regard to the software limit stops. The manual position full closed or full open will similarly ignore limit stops. The value should be set between -10 and 50.0 included |
379 |
Limit Upper |
380 |
Upper Position Limit- The SVi1000 allows the user to establish software limit stops. The control functions will NOT allow the valve position to be above than the upper position limit.However this option DOES NOT PROVIDE MECHANICAL STOPS for the valve. During an electrical or an air failure the valve will go to the failsafe position without regard to the software limit stops. The manual position full closed or full open will similarly ignore limit stops. the range should be limited from 10.0 to 110.0. |
381 |
Set position limits |
382 |
Set low/high limists for position. |
383 |
save current parameters |
384 |
Set Position Limits |
385 |
Set the valve position low/high limits |
386 |
Check signal range |
387 |
Check signal range. |
388 |
Rate Limit Down |
389 |
Position Rate Limit Down Enable |
390 |
Rate Limit Up |
391 |
Postion Rate Limit Up Enable |
392 |
Pos. Rate Limits |
393 |
Position Rate Limits. The range shoud be between 0.8 and 100.10. |
394 |
%/s |
395 |
Bumpless Transfer |
396 |
Bumpless Transfer- enables or disables the bumpless transfer mode. |
397 |
Transfer Time |
398 |
Transfer Time is approximately the number of seconds required to move the valve 100% toward the signal position. It is between 1 and 255 sec/100%. |
399 |
sec/100% |
400 |
3d |
401 |
Cam Configuration Flags- cam used - defines relationship between flow capacity (Cv) and percentage opening of the valve. Several cams are available: linear; =%50; =%30; quick opening coefficient of 50; custom; CamFlex=%. |
402 |
Linear |
403 |
Equal%(30) |
404 |
Equal%(50) |
405 |
Quick Open |
406 |
Custom |
407 |
Camflex% |
408 |
Supply Pressure |
409 |
.3f |
410 |
Agressiveness |
411 |
1i |
412 |
cTuneFlags |
413 |
The selections for autotune algorithm. |
414 |
skips step 3 of tune algorithm |
415 |
runs step 3 of tune algorithm |
416 |
runs by default setting and skip step 3 |
417 |
runs by default setting with step 3 |
418 |
cTuneCompletionCode |
419 |
cTuneCompletionCode(TBD) |
420 |
Air Action |
421 |
Air Action- configures the actuator to operate as Air to Close, or Air to Open. |
422 |
Air to Close |
423 |
Air to Open |
424 |
Slot |
425 |
Select the slot number(0-7), and load the specific tune data. Note: When the selector on the device indicates to 8 to F, means the device is configurated to ATC and the number of slots for ATC is still from 0 to 7. |
426 |
slot 0 |
427 |
slot 1 |
428 |
slot 2 |
429 |
slot 3 |
430 |
slot 4 |
431 |
slot 5 |
432 |
slot 6 |
433 |
slot 7 |
434 |
P |
435 |
Proportional gain in %. Common values are 50 for small valves up to 4000 for large valves. Must be between 0 and 5000. |
436 |
4d |
437 |
I |
438 |
Integral time or reset time in 1/10 s. Higher values of I cause less integral action, however a value of 0 gives no integral action. Common values are 10 to 200. Values between 0 and 1000. |
439 |
(1/10s) |
440 |
D |
441 |
Derivative time or rate time in milliSeconds. Common values are 10 to 100. Values between 0 and 200. |
442 |
mS |
443 |
Padj |
444 |
Coefficient added to P when the valve is exhausting and compensates the variation in filling and exhausting speed. Can be positive or negative value range between -3000 to +3000. |
445 |
Beta |
446 |
Nonlinear gain factor. When beta is 0, the proportional gain is linear, otherwise the gain P is a function of error. The larger the beta, the smaller the gain for small error. Typical beta value is 7 or 8. Beta is ranging from -9 to +9. |
447 |
Position Comp. |
448 |
The response of the valve is different when the valve is nearly closed than when nearly open. This coefficient allows the user to make adjustments to try to equalize the valve response. Normal value is 6 and this value is ranging from 2 to 20. |
449 |
2d |
450 |
Dead Zone |
451 |
When the valve position is within the setpoint +/- the dead zone and, no additional position control is performed. This value is normally 0%. Used to avoid limit cycling for high friction valves. Range is 0 to 5%. |
452 |
Non Linear |
453 |
Active PID slot |
454 |
Current active PID parameters |
455 |
PID selector from hardware setting |
456 |
ATO/ATC from hardware setting |
457 |
Wired HART Enabled |
458 |
Enabled on Wired HART device connection only |
459 |
Wireless HART Enabled |
460 |
Enabled on WirelessHART device connection only |
461 |
Both Wired and WirelessHart Enabled |
462 |
Enabled on both Wired and WirelessHART connections |
463 |
wrong mode |
464 |
Open Stop Adj |
465 |
Open Stop Adjustment- This value must be between 60%-100%.For some valves the open position stop exceeds the nominal travel of the valve.By entering a value, the position calibration can be adjusted to nominal travel. |
466 |
5.1f |
467 |
Boost |
468 |
Boost- Compensates for the pnuematic dead band. Value ranges between 0 and 20. |
469 |
Reserved 1 |
470 |
Auto Tune Fault |
471 |
Reason for Auto Tune failure |
472 |
Pos Error Band |
473 |
Position Error Band- Value corresponding to the maximun error (in % opening) between the setpoint and the valve position allowed, before setting a status flag regards the selection of associated variable Time. Value must be between 0.5% and 200.0% . |
474 |
Pos Error Enable |
475 |
Enable Position Error- enable position error testing |
476 |
Pos Error Time |
477 |
Position Error Time- time before to set the warning when position error exceeds the configured position error limit. This time must be between 0 s and 327 s. |
478 |
s |
479 |
Position Fail |
480 |
Enable Position Fail- enable position fail testing |
481 |
Pos Fail Time |
482 |
Position Fail Time- Time before to put the valve in failsafe mode. This time must be >= 0 s and below 328 s and in any case 2s greater than Time 1 However to inhibit failsave mode set -1 s. |
483 |
Near Closed Value |
484 |
Near Closed Position- Value of valve position below which the user considers the valve as being near closed position. Used in continuous diagnostic calculations. Must be between 0 and 20%. |
485 |
Stop Adjust Flag |
486 |
Low |
487 |
High |
488 |
find stops failed |
489 |
Low Signal |
490 |
Low Input Signal- Lower range value of input signal for valve closed (direct acting) or lower range value of input signal for valve open (reverse acting). Must be between 3.8 mA and 14 mA. |
491 |
mA |
492 |
High Signal |
493 |
High Input Signal- high input signal for valve open (direct acting) or high input signal for valve closed (reverse acting). Must be between 8 mA and 20.2 mA. The difference between Lower Range Value and Upper Range Value (span) must be at least 5 mA. |
494 |
Set signal range |
495 |
Set Signal range value. |
496 |
Set the signal range |
497 |
Set the low/high input signal |
498 |
Select Storage |
499 |
Selects which curve the recently created valve signature is stored to |
500 |
Not Saved |
501 |
Saved |
502 |
Baseline |
503 |
Calibration Sensor |
504 |
Pressure |
505 |
Calibration Type |
506 |
Calibration Value |
507 |
The current value of the signal being calibrated. |
508 |
calibration failed |
509 |
Diag. Test Type |
510 |
Step Response Test |
511 |
Std. Actuator Sig. |
512 |
Speed Level |
513 |
Valve speed motion for recording data. Min is 1 and max is 10. |
514 |
Start Pos |
515 |
Start Position |
516 |
End Pos |
517 |
End Position |
518 |
Sample Rate |
519 |
Diag Param 2 |
520 |
Diag Param 3 |
521 |
diagnostics not allowed |
522 |
Option Config 0 |
523 |
Pressure Sensor 1 |
524 |
Pressure Sensor 2 |
525 |
Pressure Supply |
526 |
Input/Output Switches |
527 |
Retransmit Output |
528 |
Model |
529 |
System Version-Positioner function level. |
530 |
Undefined |
531 |
ES - EZ Smart |
532 |
SD - Standard Diag |
533 |
AD - Advanced Diag |
534 |
DP - Diagnostics Plus |
535 |
ESD |
536 |
Option Config 1 |
537 |
Remote Position Sensor |
538 |
Single Acting |
539 |
Process Variable Input |
540 |
Reserved |
541 |
Active Position Sensor |
542 |
Position Sensor- built-in or remote position sensor. |
543 |
Hall Sensor |
544 |
Remote Sensor |
545 |
I/P Type |
546 |
I/P Type- defined within Option Config 2 |
547 |
Relay Type |
548 |
Relay Type- defined within Option Config 2 |
549 |
Option Config X |
550 |
I/P Pressure |
551 |
20mA signal at a travel of |
552 |
Position in % at 20mA for position retransmitter |
553 |
7.2f |
554 |
4mA signal at a travel of |
555 |
Position in % at 4mA for position retransmitter |
556 |
X1 |
557 |
X defines signal input in % and Y defines position value in % The min and max values are respectively 4.0% and 100.0% The custom curve must have at least one pair. A coordinate (100.0%,100.0%) indicates the end of the curve. |
558 |
Y1 |
559 |
X defines signal input in % and Y defines position value in % The min and max values are respectively 0.0% and 100.0% The custom curve must have at least one pair. A coordinate (100.0%,100.0%) indicates the end of the curve. |
560 |
X2 |
561 |
Y2 |
562 |
X3 |
563 |
Y3 |
564 |
X4 |
565 |
Y4 |
566 |
X5 |
567 |
Y5 |
568 |
X6 |
569 |
Y6 |
570 |
X7 |
571 |
Y7 |
572 |
X8 |
573 |
Y8 |
574 |
X9 |
575 |
Y9 |
576 |
X10 |
577 |
Y10 |
578 |
Signal A/D |
579 |
Value of Signal A/D |
580 |
2nd Signal A/D |
581 |
Temperature A/D |
582 |
Temperature A/D value |
583 |
Atm Press A/D |
584 |
Atm Pressure AD value |
585 |
Rem Pos A/D |
586 |
Remote Position A/D value |
587 |
Supply Press A/D |
588 |
Supply Pressure A/D value |
589 |
Hall Sensor A/D |
590 |
Hall Sensor A/D value |
591 |
Act Press 1 A/D |
592 |
Acutuator Pressure 1 A/D value |
593 |
Act Press 2 A/D |
594 |
Actuator Pressure 2 A/D value. |
595 |
Pilot Press A/D |
596 |
Pilot Pressure A/D value. |
597 |
I/P Current A/D |
598 |
I/P current A/D value. |
599 |
Ref Voltage A/D |
600 |
Reference voltage A/D value. |
601 |
Signal value after factory temperature correction. |
602 |
the 2nd Signal value after factory temperature correction. |
603 |
Temperature A/D value after factory temperature correction. |
604 |
Atm pressure value after factory temperature correction. |
605 |
Remote Position value after temperature correction |
606 |
Supply pressure value |
607 |
Hall sensor value |
608 |
Actuator pressure 1 value |
609 |
Actuator Pressure 2 value |
610 |
Pilot pressure value |
611 |
I/P current value |
612 |
Reference voltage value |
613 |
SubMode |
614 |
SubMode - SVi1000 out-of-service modes can be: setup, low power, failsafe. SubMode is only valid when Mode is Out-Of-Service. |
615 |
This mode allows the user to describe the basic valve setup,to perform calibration and diagnostic. |
616 |
Fail Safe |
617 |
It is the positioner security mode when critical faults occurs or in case of position error with time 2 activated. |
618 |
Mode |
619 |
Mode - SVi1000 modes can be: out-of-service (setup, low power, fail safe, fail safe LP), normal, or manual. |
620 |
This mode allows the user to describe the basic valve setup, to perform calibration and diagnostic. |
621 |
Fail Safe with Low Power |
622 |
Fail safe with Low Power |
623 |
Out-Of-Service |
624 |
Out of service |
625 |
In the normal operating mode,the positioner responds to the 4-20 mA input signal and positions the valve appropriately. |
626 |
In the manual mode, the valve no longer responds to the 4-20mA and is locked in place.The position may be changed manually from the buttons or by send a HART command. |
627 |
Process Busy |
628 |
Percent Complete |
629 |
Analog Output |
630 |
Analog Output - fixed value to set for position retransmitter. Set to 0 will put the retrasmitter out of the fixed output mode. |
631 |
Unused(int3) |
632 |
LED State |
633 |
State of LED |
634 |
Rotary Switch State |
635 |
Push Button Edges |
636 |
the Pushbutton became pressed |
637 |
the Pushbutton became released |
638 |
Push Button State |
639 |
Pressed |
640 |
Not pressed |
641 |
Quad Encoder Step |
642 |
Terminal Board |
643 |
Presence of the DO card. |
644 |
No Card |
645 |
DO Card is installed |
646 |
Retransmit Card is installed |
647 |
DO 1 Hardware State |
648 |
Close |
649 |
DO 2 Hardware State |
650 |
Terminal Module Serial Number |
651 |
The Serial Number of Terminal Board. |
652 |
BoardID Signature |
653 |
Family |
654 |
SVi1000 Module |
655 |
Group |
656 |
Module |
657 |
Retransmit |
658 |
Version |
659 |
Firmware CRC |
660 |
Firmware Build Date |
661 |
nDummy1 |
662 |
cExtDevType |
663 |
cPreambles |
664 |
CmdRev |
665 |
cDevRev |
666 |
cSWRev |
667 |
cHWRevSignaling |
668 |
cFlags |
669 |
cDeviceID |
670 |
cnPreambles |
671 |
cnVariables |
672 |
cCfgChangeCount |
673 |
cExtStatusc |
674 |
cManufacturerID |
675 |
cPrivateLabelDistributor |
676 |
cDeviceProfile |
677 |
Read Custom Characterization Data |
678 |
Select characterization. If Custom is selected then continue to set custom points. |
679 |
Edit |
680 |
Current |
681 |
History |
682 |
Instrumentation |
683 |
Actuator |
684 |
Critical |
685 |
Pneumatics |
686 |
Electronics |
687 |
Device Faults |
688 |
Device Configuration |
689 |
Device Diagnostics |
690 |
SVi1000 |
691 |
characterization |
692 |
SVi1000 Retransmit |
693 |
SVi1000 Switches |
694 |
Sig. |
695 |
Input signal in current: 4.0 - 20.0 mA |
696 |
Input signal in percent: 0 - 100% |
697 |
Pos. |
698 |
Valve position in percent: 0 - 100% |
699 |
Y-Axis: 0 - 100% |
700 |
Input signal 4-20mA |
701 |
Input signal 0-100% |
702 |
Manual SetPoint |
703 |
Valve Position3 |
704 |
Trend |
705 |
STRIP0 |
706 |
STRIP1 |
707 |
STRIP2 |
708 |
The chart for valve position and manual set position |
709 |
Position Bar |
710 |
Signal Bar |
711 |
Monitor |
712 |
SVi1000 HART |
713 |
Setup Wizards |
714 |
Valve Travel Tuning |
715 |
Auto tune |
716 |
Custom Characterization |
717 |
Input Signal(%) |
718 |
Position Value(%) |
719 |
Charc Curve |
720 |
Characterization Curve.
Please use method Custom / Standard Characterization for editing custom characterization. |
721 |
Characterization Curve |
722 |
Graph of Characterization Curve, The X-axis is for input signal(%) and Y-axis is for valve position(%). |
723 |
Device Status |
724 |
Clear Faults |
725 |
Continuous Diagnostics |
726 |
Raw data |
727 |
Main signal |
728 |
What is Current Mode? |
729 |
What is Current Mode? - To get the current mode of this device |
730 |
Read Factory Values |
731 |
Values |
732 |
General Information |
733 |
Common Tables |
734 |
Interoperable * |
735 |
Universal |
736 |
Common Practice |
737 |
Pressure Universal |
738 |
Pressure Common |
739 |
Menu |
740 |
Rosemount Magflow |
741 |
Rosemount Pressure |
742 |
Moore Products |
743 |
Siemens Pressure |
744 |
P2 Common |
745 |
PV |
746 |
Wireless |
747 |
Munge |
748 |
Analytical Oxygen/Combustion/Methane/CO |
749 |
Flow Totalizer - Rate Indicator |
750 |
Flow Variable Area |
751 |
Flow Coriolis Mass |
752 |
Flow Ultrasonic - Gas |
753 |
Flow Ultrasonic - Liquid |
754 |
Level Ultrasonic |
755 |
Level Radar - Microwave |
756 |
Pressure Absolute |
757 |
Level Capacitance |
758 |
Pressure Differential |
759 |
Level Hydrostatic |
760 |
Level |
761 |
Pressure Absolute/Relative |
762 |
Level Microwave |
763 |
Level Radar |
764 |
Level Density - Nuclear |
765 |
Flow Electromagnetic |
766 |
Flow Vortex |
767 |
Flow Magmeter - Rack Version |
768 |
Flow High Performance Magmeter |
769 |
Flow Mass |
770 |
Flow Ultrasonic |
771 |
Flow Magnetic |
772 |
Flow Magnetic - Custody Transfer |
773 |
Flow 2-wire Magnetic |
774 |
Flow Coriolis Mass - Custody Transfer |
775 |
Flow 2-wire Vortex |
776 |
Flow Thermal |
777 |
Temperature Head Mounted |
778 |
Temperature Compact |
779 |
Temperature DIN Rail Mounted |
780 |
Analytical pH |
781 |
Analytical Conductivity - Conductive |
782 |
Analytical Conductivity - Inductive |
783 |
Temperature Sensor |
784 |
Level Servo |
785 |
Temperature DIN Rail |
786 |
Analytical ORP |
787 |
Analytical pION |
788 |
Analytical IConc |
789 |
Analytical Electrode Conductivity |
790 |
Analytical Toroidal Conductivity |
791 |
Valve Positioner |
792 |
Valve Controller |
793 |
Pressure Multivariable |
794 |
Analytical pH/ORP |
795 |
Flow Coriolis |
796 |
Analytical Conductivity |
797 |
Valve Positioner/Pneumatic |
798 |
Valve Actuator - Electrical |
799 |
Temperature Dual Input |
800 |
Valve Actuator |
801 |
Analytical Corrosion |
802 |
Analytical Density - Nuclear |
803 |
Level Continuous Gamma |
804 |
Level Laser Ranging |
805 |
Pressure Safety |
806 |
Flow Vortex/Swirl |
807 |
Flow |
808 |
Analytical 4 Electrode Conductivity |
809 |
Analytical 2 Electrode Conductivity |
810 |
Analytical Troidal Conductivity |
811 |
Analytical pH/ORP/pIon/Ion Conc |
812 |
Analytical Oxygen |
813 |
Pressure P, dP + Multivariable |
814 |
Temperature 2-wire |
815 |
Pressure w/ PID |
816 |
Pressure Saturated Steam |
817 |
Analytical Density - Gamma Ray |
818 |
Level Gamma Ray |
819 |
Analytical Weight - Gamma Ray |
820 |
Level w/ Density - Gamma Ray |
821 |
Analytical Motorized Density |
822 |
Flow Multivariable Mass |
823 |
Temperature Multivariable |
824 |
Specialty HART-to-Analog Signal Conv |
825 |
Flow dP |
826 |
Pressure Low Power |
827 |
Pressure Certified SIL2 |
828 |
Interoperable |
829 |
Analytical Density |
830 |
Valve Positioner/Electropneumatic |
831 |
Flow Magnetic Inductive |
832 |
Analytical |
833 |
Analytical Opacity Monitor |
834 |
Analytical Oxygen/Combustion |
835 |
Analytical 2-wire Conductivity |
836 |
Analytical 2-wire Conductivity w/ pure H20 |
837 |
Analytical 2-wire Conductivity w/ % conc |
838 |
Analytical 2-wire Oxygen/Chlorine/Ozone |
839 |
Analytical 2-wire pH/ORP |
840 |
Analytical 2-wire Contacting/Toroidal |
841 |
Analytical Oxygen - High Temperature |
842 |
Analytical Conductivity w/ % conc |
843 |
Analytical pH/ORP/ISE |
844 |
Analytical Amperometric Analyzer |
845 |
Analytical Dual Input Analyzer |
846 |
Pressure Pressure/Level |
847 |
Analytical Consistency |
848 |
Analytical Microwave-Pulp Consistency |
849 |
Analytical Microwave Consistency |
850 |
Valve Digital Positioner |
851 |
Valve Smart Valve Positioner |
852 |
Temperature Differential |
853 |
Analytical Dissolved Oxygen |
854 |
Analytical pH Controller |
855 |
Analytical Conductivity/Resistivity Controller |
856 |
Analytical Conductivity/Resistivity |
857 |
Level Displacement |
858 |
Temperature w/ PID |
859 |
Valve Universal Positioner |
860 |
Level Radar - Reflex |
861 |
Pressure Gauge |
862 |
Pressure Sealed Gauge |
863 |
Valve Digital Positioner w/ PID |
864 |
Valve Intelligent Positioner |
865 |
Analytical Gas Detector |
866 |
Analytical Infrared Gas Detector |
867 |
Analytical Electrochemical Gas Detector |
868 |
Analytical Measuring Instruments |
869 |
Level Magnostrictive |
870 |
Valve On/Off Valve Controller |
871 |
Level Radar - Pulse |
872 |
Level Continuous Capactive |
873 |
Valve Damper Actuator |
874 |
Valve Electric Damper Actuator |
875 |
Flow Electronic |
876 |
Flow Vortex - Multiparameter Mass |
877 |
Analytical UV Flame Detector |
878 |
Analytical UVIR Flame Detector |
879 |
Analytical IR Flame Detector |
880 |
Analytical Fire Detector |
881 |
Analytical Gas Detector Display Interface |
882 |
Analytical Toxic Gas Detector |
883 |
Analytical Flame Detector |
884 |
Level Nuclear |
885 |
Analytical Microwave |
886 |
Analytical Density - Radiometric |
887 |
Level Radiometric |
888 |
Analytical Gas/Steam |
889 |
Level Nucleonic |
890 |
Analytical Blade Consistency |
891 |
Analytical Rotary Consistency |
892 |
Pressure Absolute/Differential |
893 |
Positioner |
894 |
Analytical Combustible Gas Detector |
895 |
Analytical Point Infrared Gas Detector |
896 |
Analytical Multi-Detector Monitor System |
897 |
Valve Positioner/Electro-hydraulic |
898 |
Valve Monitor |
899 |
Enable on token-passing data link layer only |
900 |
Enable on TDMA data link layer only |
901 |
Enable on both TDMA and token-passing data link layer |
902 |
WirelessHART Device |
903 |
IEEE 802.15.4 2.4GHz DSSS with O-QPSK Modulation |
904 |
Primary |
905 |
Primary Analog Channel |
906 |
Secondary |
907 |
Secondary Analog Channel |
908 |
Tertiary |
909 |
Tertiary Analog Channel |
910 |
Quaternary |
911 |
Quaternary Analog Channel |
912 |
Quinary |
913 |
Quinary Analog Channel |
914 |
Critical Power Failure |
915 |
Lock all |
916 |
Trim not supported |
917 |
Configuration cannot be changed |
918 |
Locked by gateway |
919 |
Analog Input Channel |
920 |
Analog Output Channel |
921 |
Secondary Analog Channel Saturated |
922 |
Tertiary Analog Channel Saturated |
923 |
Quaternary Analog Channel Saturated |
924 |
Quinary Analog Channel Saturated |
925 |
Secondary Analog Channel Fixed |
926 |
Tertiary Analog Channel Fixed |
927 |
Quaternary Analog Channel Fixed |
928 |
Quinary Analog Channel Fixed |
929 |
Simulation active |
930 |
The device is in simulation mode and the level output is not representative of the process. |
931 |
Non-Volatile memory failure |
932 |
The Non-Volatile memory check is invalid or maybe corrupt, or the battery of a battery-backed memory has failed. |
933 |
Volatile memory error |
934 |
The RAM memory check is invalid or maybe corrupt |
935 |
Watchdog reset executed |
936 |
A watchdog reset has been performed |
937 |
Voltage conditions out of range |
938 |
A voltage condition is outside its allowable range |
939 |
Environmental conditions out of range |
940 |
An internal or environmental condition is beyond acceptable limits. |
941 |
Electronic failure |
942 |
A hardware problem not related to the sensor has been detected. |
943 |
Subdevice list changed |
944 |
Duplicate master detected |
945 |
Capacity Denied |
946 |
The device was unable to acquire the communication bandwidth required to support the Burst Messaging specified |
947 |
Bandwidth allocation pending |
948 |
Block transfer pending |
949 |
Continuous |
950 |
The Burst Message is published continuously at (worst case) the Minimum Update Period. |
951 |
Window |
952 |
The Burst Message is triggered when the source value deviates more than the specified trigger value |
953 |
Rising |
954 |
The Burst Message is triggered when source value Rises Above the specified trigger value. |
955 |
Falling |
956 |
The Burst Message is triggered when the source value Falls Below the specified trigger value. |
957 |
On-change |
958 |
The Burst Message is triggered when any value in the message changes. |
959 |
Battery life |
960 |
Percent range |
961 |
Loop current |
962 |
Primary variable |
963 |
Secondary variable |
964 |
Tertiary variable |
965 |
Quaternary variable |
966 |
Off |
967 |
Enable event notification on token-passing data link layer |
968 |
Enable event notification on TDMA data link layer |
969 |
Enable event notification on TDMA and token-passing data link layer |
970 |
Configuration changed event pending |
971 |
Device status event pending |
972 |
More status available event pending |
973 |
Disable |
974 |
Enable single data point trending |
975 |
Enable filtered trending |
976 |
The sample is filtered using a time constant equal to one-third the trend sample period |
977 |
Enable average trending |
978 |
The value obtained by dividing the sum of all samples in the period by the total number of samples |
979 |
Read receive time |
980 |
Write date and time |
981 |
Non-Volatile Clock |
982 |
When set the device contains a battery-backed clock. In this case, the clock does not need to be reset if there is a power failure. |
983 |
Clock Uninitialized |
984 |
The real-time clock has never been set with the date and time. For example, the clock is volatile and power was removed from and restored to the device. |
985 |
Line Power |
986 |
Device is powered via a hardwired connection to a plant power source. |
987 |
Battery Power |
988 |
Device operates solely of its (internal) battery. |
989 |
Recharg.B./Power Sc. |
990 |
Rechargeable battery power or power scavenging. The device has a short-term supply of energy (must be at least one hour) that is continuously being replenished. The device's power is being replenished by harvesting and converting energy from the environment surrounding the device (e.g. solar, vibration, heat). |
991 |
Idle |
992 |
Active search |
993 |
Looking for neighbors. |
994 |
Negotiating |
995 |
Quarantined |
996 |
Operational |
997 |
Suspended |
998 |
Deep sleep |
999 |
Ultra-low power |
1000 |
Wireless Signal Found |
1001 |
Network packets have been recognized |
1002 |
Wireless Signal Identified |
1003 |
Wireless Time Synchronized |
1004 |
WirelessHart Signal Identified |
1005 |
Network Admission Requested |
1006 |
Set on transmission of first join request |
1007 |
Join Retrying |
1008 |
Set after the first join request retry (i.e. when number of join requests > 2). Cleared when device is authenticated or when Active Search started. |
1009 |
Join Failed |
1010 |
Set on transition from Active Search mode to Deep Sleep mode. Cleared on transition from Deep Sleep to Active Search. |
1011 |
Network Security Clearance Granted |
1012 |
Network Manager Session Established |
1013 |
Network Joined |
1014 |
Network Joined. Normal superframe and links obtained. |
1015 |
Network Bandwidth Requested |
1016 |
Negotiating Network Properties. Gateway session obtained. |
1017 |
Join Complete |
1018 |
Normal operating mode. |
1019 |
None |
1020 |
No restrictions |
1021 |
SI Only |
1022 |
Restricted to SI Unit codes only |
1023 |
Open - accept all devices |
1024 |
Use Whitelist for access |
1025 |
Use Blacklist |
1026 |
Use Whitelist and Blacklist |
1027 |
Lockdown - disable join links |
1028 |
Process automation device |
1029 |
Discrete device |
1030 |
Hybrid:Process automation+discrete |
1031 |
I/O System |
1032 |
WirelessHART Process automation device |
1033 |
WirelessHART Discrete device |
1034 |
WirelessHART Process automation+discrete |
1035 |
WirelessHART Gateway |
1036 |
WirelessHART Access point |
1037 |
WirelessHART Process adapter |
1038 |
WirelessHART Discrete adapter |
1039 |
WirelessHART enabled handheld/portable maintenance tool |
1040 |
Don't attempt to join |
1041 |
Join now |
1042 |
Attempt to join immediately on powerup or reset |
1043 |
US |
1044 |
United States of America |
1045 |
JP |
1046 |
Japan |
1047 |
DE |
1048 |
Germany |
1049 |
FR |
1050 |
France |
1051 |
ES |
1052 |
Spain |
1053 |
RU |
1054 |
Russian Federation |
1055 |
CN |
1056 |
People's Republic of China |
1057 |
Nominal |
1058 |
Device Power Status is nominal |
1059 |
Device Power Status is low |
1060 |
Critically Low |
1061 |
Device Power Status is critically low |
1062 |
Recharging - Low |
1063 |
Recharging - High |
1064 |
Secondary Master |
1065 |
Primary Master |
1066 |
PV Analog Channel Saturated |
1067 |
PV Analog Channel Fixed |
1068 |
Device Profile |
1069 |
Device Diagnostic Status 0 |
1070 |
Device Diagnostic Status 1 |
1071 |
I/O and Subdevice Status |
1072 |
WirelessHART Status |
1073 |
Time Set Option |
1074 |
Real Time Clock Flags |
1075 |
Country |
1076 |
Country of intended device installation |
1077 |
SI Unit Control |
1078 |
Indicates if device should only display SI unit codes |
1079 |
Burst Trigger Units |
1080 |
Burst Trigger Class |
1081 |
Burst Msg Trigger Mode |
1082 |
Event Status |
1083 |
Event Notification Control |
1084 |
Operate mode Mask |
1085 |
Not Defined |
1086 |
Power Source |
1087 |
Power source of the device |
1088 |
Wireless Mode |
1089 |
Join Status |
1090 |
The current location in the joining sequence of operations. |
1091 |
Network Access Mode |
1092 |
Join Mode |
1093 |
Power Status |
1094 |
Power status of the device |
1095 |
Master Mode |
1096 |
This sets the I/O System's Data Link Layer operation on all channels as either primary or secondary master mode. |
1097 |
Trend Units |
1098 |
Trend Device Variable Unit Code |
1099 |
Trend Classification |
1100 |
Trend Device Variable Classification |
1101 |
Trend Control |
1102 |
Device Status Mask |
1103 |
Ext dev status Mask |
1104 |
Device Diagnostic Status 0 Mask |
1105 |
Device Diagnostic Status 1 Mask |
1106 |
Device Diagnostic Status 2 Mask |
1107 |
Device Diagnostic Status 3 Mask |
1108 |
AO saturated Mask |
1109 |
AO fixed Mask |
1110 |
Latched Device Status |
1111 |
Latched Ext dev status |
1112 |
Latched Operate Mode |
1113 |
Latched Device Diagnostic Status 0 |
1114 |
Latched Device Diagnostic Status 1 |
1115 |
Latched Device Diagnostic Status 2 |
1116 |
Latched Device Diagnostic Status 3 |
1117 |
Latched AO saturated |
1118 |
Latched AO fixed |
1119 |
Wireless Module Device Type |
1120 |
Device type of the Wireless Radio |
1121 |
Wireless Module Manufacturer ID |
1122 |
Manufacturer of the Wireless Radio contained in this device |
1123 |
Subdevice Device Type |
1124 |
Device type of the subdevice attached to this system. |
1125 |
Subdevice Manufacturer Id |
1126 |
Manufacturer of the subdevice attached to this system. |
1127 |
Time Stamp |
1128 |
Relative time value on network |
1129 |
8u |
1130 |
Dynamic Variables Returned for Device Variables |
1131 |
Command Response Truncated |
1132 |
Configuration Change Counter Mismatch |
1133 |
Status bytes mismatch |
1134 |
Device Variable |
1135 |
8.3f |
1136 |
Select the slot to read: |
1137 |
Firmware Version - %{0}.%{1}.%{2} |
1138 |
This DD is tokenized by version 8
65ef0101.ddl's Time-stamp: <2011-05-24 11:27:14 SHA2118L0003> |
1139 |
Warning: To restore from %{0} mode, please clear faults first, Otherwise it will cause unexpected result! |
1140 |
Continue;Exit |
1141 |
Set Mode Target |
1142 |
Normal;Manual;Setup |
1143 |
WARNING. Leaving Normal Mode will result in LOSS of process CONTROL and may cause sudden valve move. DO YOU WISH TO CONTINUE? |
1144 |
Procedure aborted
Try Again. |
1145 |
WARNING. This procedure requires to enter setup mode and will result in LOSS of PROCESS CONTROL. Do you wish to proceed ? |
1146 |
Changing mode to Setup |
1147 |
WARNING. This procedure will MOVE the valve. This will result in LOSS of process CONTROL. DO YOU WISH TO CONTINUE? |
1148 |
Please wait while searching for mechanical stops. This should take approximately 30 seconds. Press OK to continue. |
1149 |
Please Wait ... |
1150 |
Press the ABORT key to abort the process. |
1151 |
Find stop limits FAILED. |
1152 |
Please verify that:
- Air supply is sufficient;
- The valve is not handjacked;
- The mounting is adequate;
- The find stop is not allowed in marginal status;
- The find stop is not allowed in lower power status; |
1153 |
Please verify that (Continued):
- The Travel Sensor is set properly.
Then re-run find stops again. |
1154 |
Find stop limits completed. |
1155 |
The device was configured as ATO valve, so current PID selector must be at 0 position to do autotune! |
1156 |
The device was configured as ATC valve, so current PID selector must be at 8 position to do autotune! |
1157 |
Remove software limit stops to be able to perform autotune. |
1158 |
Agressiveness: |
1159 |
0;9;8;7;6;5;4;3;2;1;-1;-2;-3;-4;-5;-6;-7;-8;-9 |
1160 |
Enter the Supply pressure in psi |
1161 |
Error: Supply pressure is out of range. Enter again! |
1162 |
Please wait while searching for optimal valve tuning parameters. This should take approximately 3 to 6 minutes. |
1163 |
Auto Tune FAILED! |
1164 |
Auto Tune completed. |
1165 |
This mode is not appropriate for the Setup Wizard. Click OK to change mode or Abort to exit Setup Wizard. |
1166 |
Find Valve Stops: |
1167 |
Skip This Task;Run Find Stops |
1168 |
Auto Tune: |
1169 |
Skip This Task;Run Auto Tune |
1170 |
Enter the Supply pressure |
1171 |
Auto Tune FAILED. |
1172 |
Reset to Factory Default: |
1173 |
Skip This Task;Set All Values to Default |
1174 |
Are you sure you want to reset all data to factory defaults? |
1175 |
Reset to Factory Defaults completed. |
1176 |
No faults recorded. |
1177 |
Do you want to clear current faults ? |
1178 |
No;Yes |
1179 |
Device has malfunctioned |
1180 |
Faults have been cleared. |
1181 |
No Historic Faults. |
1182 |
Do you want to clear all faults ? |
1183 |
Please select the device variable list to display: |
1184 |
device variable 0 – 7;device variable 8 – 15;device variable 16 – 19;Exit |
1185 |
0. Position: %[.3f]{0} %{1} (%{2})
1. Setpoint: %[.3f]{3} %{4} (%{5})
2. Temperature: %[.3f]{6} %{7} (%{8})
3. Pilot Pressure: %[.3f]{9} %{10} (%{11})
4. Signal: %[.3f]{12} %{13} (%{14})
5. DO1: %[.3f]{15} %{16} (%{17})
6. DO2: %[.3f]{18} %{19} (%{20})
7. Volt Input: %[.3f]{21} %{22} (%{23})
|
1186 |
8. Raw Position: %[.3f]{0} %{1} (%{2})
9. Num. Strokes: %[.3f]{3} %{4} (%{5})
10. Number Cycles: %[.3f]{6} %{7} (%{8})
11. Pos. Retransmit: %[.3f]{9} %{10} (%{11})
12. IP Current: %[.3f]{12} %{13} (%{14})
13. Friction: %[.3f]{15} %{16} (%{17})
14. Pos Error Band: %[.3f]{18} %{19} (%{20})
15. Pos. Limit High: %[.3f]{21} %{22} (%{23})
|
1187 |
16. Pos. Limit Low: %[.3f]{0} %{1} (%{2})
17. Characterization: %[.3f]{3} %{4} (%{5})
18. Open Stop Adjustment: %[.3f]{6} %{7} (%{8})
19. Percent of Range: %[.3f]{9} %{10} (%{11})
|
1188 |
The device must not be in NORMAL mode to perform this operation. Please change the device mode. |
1189 |
WARNING! This command strokes the valve and should not be performed with process running. |
1190 |
Manual Setpoint: |
1191 |
Set By Position;Set By Signal |
1192 |
Valve Position (-5% ~ 160%): |
1193 |
Position Corresponding to Signal in mA: |
1194 |
Select Tight Shutoff |
1195 |
Apply new changes to the device? Press:
OK for YES
ABORT for NO |
1196 |
Select Cam
|
1197 |
Enter/edit cam characterization ? |
1198 |
Input your 9 XY pairs of points between (0.0%, 0.0%) and (100.0%,100.0%). |
1199 |
X[%{i}]
|
1200 |
Y[%{i}]
|
1201 |
Invalid segment: X%{a} = %{0}, X%{i} = %{1} - Same X. |
1202 |
Invalid segment: X0=0, X%{i}=%{0} - Slope too low. |
1203 |
Invalid segment: X%{a}=%{0}, X%{i}=%{1} - Slope too low. |
1204 |
Invalid segment: X0=0, X%{i}=%{0} - Slope too high. |
1205 |
Invalid segment: X%{a}=%{0}, X%{i}=%{1} - Slope too high. |
1206 |
Invalid segment: X%{a}=%{0}, X%{i}=%{1} - Slope changes too much. |
1207 |
Invalid segment: X%{a}=%{0}, X%{i}=%{1} - Slope changes too small. |
1208 |
Found bad segment(s), please choose: |
1209 |
Enter data again to correct error(s); Abort the method; Proceed to save without correction. |
1210 |
Set Manual Stop Position: |
1211 |
Low;High;Exit |
1212 |
Wait until the valve is stable, then press OK. |
1213 |
Procedure failed. The stroke could be too small. |
1214 |
Open Stop Adjustment set properly. |
1215 |
The current active PID slot is not at 0, following process only set the PID parameters for slot 0.Press OK to continue! |
1216 |
slot = 0
P = %{0} %
I = %{1} (1/10s)
D = %{2} ms
Padj = %{3} %
|
1217 |
Beta = %{4}
Compens. pos.= %{6}
Dead zone = %{7} %
Boost = %{8}
|
1218 |
P (%) |
1219 |
I (1/10s) |
1220 |
D (ms) |
1221 |
Padj (%) |
1222 |
Compensation |
1223 |
Dead Zone (%) |
1224 |
Set the I/P Value(0~65535): |
1225 |
WARNING. This procedure requires to enter setup mode. Please change to setup mode first. |
1226 |
Enter the low signal value between 3.5 and 8.0 mA: |
1227 |
The value must be between 3.5 and 8.0 mA |
1228 |
Set the input current to %[.2f]{0} mA and press OK |
1229 |
The current is not set correctly! Please check and set again. |
1230 |
Enter the high signal value between 16.0 and 21.0 mA: |
1231 |
The value must be between 16.0 and 21.0 mA |
1232 |
WARNING. To continue this procedure will reset calibration data to factory defaults. DO YOU WISH TO CONTINUE? |
1233 |
Calibraton data have been reset to factory defaults. |
1234 |
After resetting, the current values will be lost. Press ABORT if you do not wish to proceed. |
1235 |
Valve Historian data was reset. |
1236 |
A reset may cause the valve to move !
Perform a RESET? |
1237 |
Reset on SVi1000 did NOT correctly execute. |
1238 |
Reset on SVi1000 correctly executed. |
1239 |
Error: Lower(%[.3f]{0}) and upper(%[.3f]{1}) ranges must differ by at least 5%. |
1240 |
Limit Upper (%[.3f]{1} %) must be greater than Limit Low (%[.3f]{0} %)
Please enter the value again!
|
1241 |
High Signal (%[.2f]{1} mA) must be at least 5mA larger than Low Signal (%[.2f]{0} mA)
Please enter the value again!
|
1242 |
The lower value(%{0} %) must lower than upper value(%{1} %)
Please enter the value again!
|
1243 |
Interrupting the current process. |
1244 |
The process was interrupted. |
1245 |
The position retransmitter has restored to the normal control mode of following the position. |
1246 |
There is no position retrasmitter for this device. |
1247 |
Setting analog output is disabled when device is in the %{0} mode |
1248 |
Set current output of the position retransmitter: |
1249 |
Set Output in mA;Finished |
1250 |
Set Output in 4 - 20 mA: |
1251 |
Change output range is disabled when device is in the %{0} mode |
1252 |
Lower Range in % |
1253 |
Upper Range in % |
1254 |
Error: Lower and upper ranges must differ by at least 5%. |
1255 |
Set re-transmitter range is done. |
1256 |
The main board serial number is: %[d]{nSerialNumber}
The Position Retransmit board serial number is: %[d]{nTerminalNumber} |
1257 |
The main board serial number is: %[d]{nSerialNumber} |
1258 |
The both of position limits are disabled, No position limits can be set. |
1259 |
Enter the lower limit value(-10.0%, 50.0%): |
1260 |
Enter the upper limit value(10.0%, 110.0%): |
1261 |
The new lower limit(%{0}%) is greater more than upper limit(%{1}%) |
1262 |
No accept, quit;Re-enter the limits |
1263 |
The new limits:%{0}%, %{1}% |
1264 |
Accept new limits;Re-enter the limits;No accept, quit |
1265 |
Enter the low input current(3.8mA, 14.0mA): |
1266 |
Enter the high input current(8.0mA, 20.2mA): |
1267 |
High Signal (%[.2f]{1} mA) must be at least 5mA larger than Low Signal (%[.2f]{0} mA)
|
1268 |
No accept, quit;Re-enter the range |
1269 |
The new signal range:%{0}mA, %{1}mA |
1270 |
Accept new range;Re-enter the range;No accept, quit |
1271 |
Device is currently in NORMAL mode. |
1272 |
Device is currently in MANUAL mode. |
1273 |
Device is currently in SETUP mode. |
1274 |
Device is currently in LOW POWER mode. |
1275 |
Device is currently in FAILSAFE mode. |
1276 |
Device is currently in OOS mode. |
1277 |
Device Mode Unknown. |